The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 06934955901348), received between Aug 7, 2025 and Aug 15, 2025 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 3 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Split, Cut or Torn2
  • Material Separation1

Patient problems

  • Foreign Body In Patient3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 15, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE JADE BALLOON WAS USED WITH A NON-ABBOTT SHEATH VIA PEDAL ACCESS TO TREAT A LESION IN LEFT SUPERFICIAL FEMORAL ARTERY (SFA). THE BALLOON WAS INFLATED TO 18 ATMOSPHERES (ATM) FOR 2 MINUTES AND THEN DEFLATED. AN INDEFLATOR REMAINED HOOKED TO BALLOON AT NEGATIVE PRESSURE WHILE THE BALLOON WAS REMOVED FROM THE WIRE. THERE WAS RESISTANCE AND THE BALLOON SHEARED OFF AT MIDDLE POINT OF BALLOON, LEAVING HALF OF THE BALLOON ABOUT 50 MILLIMETERS (MM) IN-VIVO AS IT WAS TAKEN OUT OF THE BODY. NEW BALLOONS ACROSS FRAGMENTED BALLOON WERE PLACED TO SNARE THE FRAGMENT AND A SNARE WAS USED TO ATTEMPT FOR REMOVAL OF THE FRAGMENT. A STENT WAS PLACED TO ADHERE BALLOON AGAINST THE ARTERY WALL. THE PATIENT'S STATUS WAS UNKNOWN.

    Patient problems: Foreign Body In Patient

    MAUDE report 22813616 ↗
  • InjuryAug 15, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE JADE BALLOON WAS USED WITH A NON-ABBOTT SHEATH VIA PEDAL ACCESS TO TREAT A LESION IN LEFT SUPERFICIAL FEMORAL ARTERY (SFA). THE BALLOON WAS INFLATED TO 18 ATMOSPHERES (ATM) FOR 2 MINUTES AND THEN DEFLATED. AN INDEFLATOR REMAINED HOOKED TO BALLOON AT NEGATIVE PRESSURE WHILE THE BALLOON WAS REMOVED FROM THE WIRE. THERE WAS RESISTANCE AND THE BALLOON SHEARED OFF AT MIDDLE POINT OF BALLOON, LEAVING HALF OF THE BALLOON ABOUT 50 MILLIMETERS (MM) IN-VIVO AS IT WAS TAKEN OUT OF THE BODY. NEW BALLOONS ACROSS FRAGMENTED BALLOON WERE PLACED TO SNARE THE FRAGMENT AND A SNARE WAS USED TO ATTEMPT FOR REMOVAL OF THE FRAGMENT. A STENT WAS PLACED TO ADHERE BALLOON AGAINST THE ARTERY WALL. THE PATIENT'S STATUS WAS UNKNOWN.

    Patient problems: Foreign Body In Patient

    MAUDE report 22813602 ↗
  • InjuryAug 7, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT THE JADE BALLOON WAS USED WITH A NON-ABBOTT SHEATH VIA PEDAL ACCESS TO TREAT A LESION IN LEFT SUPERFICIAL FEMORAL ARTERY (SFA). THE BALLOON WAS INFLATED TO 18 ATM FOR 2 MINUTES AND THEN DEFLATED. INDEFLATOR REMAINED HOOKED TO BALLOON AT NEGATIVE PRESSURE WHILE BALLOON WAS WALKED OF WIRE. THERE WAS RESISTANCE AND THE BALLOON SHEARED OFF AT MIDDLE POINT OF BALLOON, LEAVING HALF OF THE BALLOON ABOUT 50MM IN-VIVO AS IT WAS TAKEN OUT OF THE BODY. NEW BALLOONS ACROSS FRAGMENTED BALLOON WAS PLACED TO SNARE THE FRAGMENT AND A SNARE WAS USED TO ATTEMPT FOR REMOVAL OF THE FRAGMENT. A STENT WAS PLACED TO ADHERE BALLOON AGAINST THE ARTERY WALL. ADDITIONAL INFORMATION WAS RECEIVED AND THE PATIENT WAS DISCHARGED.

    Patient problems: Foreign Body In Patient