The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 06934955901621), received between Jul 25, 2024 and Aug 5, 2024 — classified as 2 injuries.
The most frequently reported specific device problem is “Material Separation” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation2
Patient problems
- Foreign Body In Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryAug 5, 2024Material Separation; Material Separation
DURING TREATMENT OF AN 80%-90% MIXED CALCIFIED PLAQUE/LESION WHICH WAS APPROXIMATELY 2.0 MM- 2.5 MM IN LENGTH, IN THE DISTAL POSTERIOR TIBIAL ARTERY, THE JADE BALLOON SEPARATED FROM THE CATHETER ON EXITING THE SHEATH DURING THE PROCEDURE. THE BALLOON WAS INITIALLY INFLATED AT LESION SITE. THERE WAS DIFFICULTY TO REMOVE THE ORBITAL ATHERECTOMY DEVICE (OAD) AFTER BALLOON DEFLATION. HOWEVER, THE BALLOON WAS PULLED AND EVENTUALLY REMOVED. AFTER PHYSICAL INSPECTION AND RESULTS FROM ANGIOGRAPHY SHOWED A PART OF BALLOON AT THE SITE OF INFLATION AND THIS PART OF THE BALLOON WAS LEFT IN THE LEG. BACKGROUND: RECEIVED THE FOLLOWING ADDITIONAL INFORMATION ON JULY 25, 2024: 1. WHEN DID THE PROBLEM OCCUR? (DURING SET-UP, PROCEDURE, OR PERFORMANCE TESTING) - DURING PROCEDURE. 2. COULD YOU PLEASE PROVIDE VESSEL AND LESION DETAILS? (VESSEL, LESION LOCATION, % STENOSIS, LEVEL OF CALCIFICATION, DIAMETER, ETC.) - DISTAL PT. 80%-90% MIXED CALCIFIED/PLAQUE LESION. APPROX. 2.0MM-2.5MM DIAMETER. 3. WAS THE BALLOON INFLATED DURING THE PROCEDURE? - BALLOON WAS INFLATED AT LESION SITE. 4. DID THE BALLOON REMAIN INTACT AND COMPLETELY REMOVED FROM THE BODY? - AFTER BALLOON DEFLATION, DEVICE WAS DIFFICULT TO RE
Patient problems: Foreign Body In Patient
MAUDE report 19903963 ↗InjuryJul 25, 2024Material Separation; Material Separation
DURING TREATMENT OF AN 80%-90% MIXED CALCIFIED PLAQUE/LESION WHICH WAS APPROXIMATELY 2.0 MM- 2.5 MM IN LENGTH, IN THE DISTAL POSTERIOR TIBIAL ARTERY, THE JADE BALLOON SEPARATED FROM THE CATHETER ON EXITING THE SHEATH DURING THE PROCEDURE. THE BALLOON WAS INITIALLY INFLATED AT LESION SITE. THERE WAS DIFFICULTY TO REMOVE THE ORBITAL ATHERECTOMY DEVICE (OAD) AFTER BALLOON DEFLATION. HOWEVER, THE BALLOON WAS PULLED AND EVENTUALLY REMOVED. AFTER PHYSICAL INSPECTION AND RESULTS FROM ANGIOGRAPHY SHOWED A PART OF BALLOON AT THE SITE OF INFLATION AND THIS PART OF THE BALLOON WAS LEFT IN THE LEG.
Patient problems: Foreign Body In Patient
MAUDE report 19829743 ↗
Other Catheter, Angioplasty, Peripheral, Transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page