Skip to content

Jade PTA Balloon Dilatation Catheter 596024022Recalls & reported problems

OrbusNeich Medical (ShenZhen) Co., Ltd. · Percutaneous Transluminal Angioplasty (PTA) Catheter · Class 2

All models of Jade PTA Balloon Dilatation Catheter (15)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 06934955903953), received between Sep 2, 2025 and Sep 4, 2025 — classified as 3 injuries.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 3 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Split, Cut or Torn2
  • Material Rupture1

Patient problems

  • Foreign Body In Patient3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjurySep 4, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING PROCEDURE; THE JADE BALLOON WAS INFLATED IN MID SUPERFICIAL FEMORAL ARTERY (SFA) TO 8ATM AND IT RUPTURED. THE PHYSICIAN ATTEMPTED TO PULL THE BALLOON OUT OF THE SFA OVER THE WIRE AND IT COULD NOT MOVE, IT JUST ACCORDIONED. THE PEDAL ACCESS WAS OBTAINED IN AN ATTEMPT TO CAPTURE BALLOON WITH A SNARE. SOME OF THE BALLOON WAS SNARED OUT AND A STENT WAS PLACED TO ADHERE THE LEFTOVER BALLOON TO THE ARTERY WALL.

    Patient problems: Foreign Body In Patient

    MAUDE report 22968736 ↗
  • InjurySep 2, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING PROCEDURE; THE JADE BALLOON WAS INFLATED IN MID SUPERFICIAL FEMORAL ARTERY (SFA) TO 8 ATM AND IT RUPTURED. THE PHYSICIAN ATTEMPTED TO PULL THE BALLOON OUT OF THE SFA OVER THE WIRE AND IT COULD NOT MOVE, IT JUST ACCORDION. THE PEDAL ACCESS WAS OBTAINED IN AN ATTEMPT TO CAPTURE BALLOON WITH A SNARE. SOME OF THE BALLOON WAS SNARED OUT AND A STENT WAS PLACED TO ADHERE THE LEFTOVER BALLOON TO THE ARTERY WALL.

    Patient problems: Foreign Body In Patient

    MAUDE report 22948898 ↗
  • InjurySep 2, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING PROCEDURE; THE JADE BALLOON WAS INFLATED IN MID SUPERFICIAL FEMORAL ARTERY (SFA) TO 8 ATM AND IT RUPTURED. THE PHYSICIAN ATTEMPTED TO PULL THE BALLOON OUT OF THE SFA OVER THE WIRE AND IT COULD NOT MOVE, IT JUST ACCORDION. THE PEDAL ACCESS WAS OBTAINED IN AN ATTEMPT TO CAPTURE BALLOON WITH A SNARE. SOME OF THE BALLOON WAS SNARED OUT AND A STENT WAS PLACED TO ADHERE THE LEFTOVER BALLOON TO THE ARTERY WALL.

    Patient problems: Foreign Body In Patient

    MAUDE report 22948897 ↗

Other Catheter, Angioplasty, Peripheral, Transluminal devices

All Catheter, Angioplasty, Peripheral, Transluminal devices →

Sources and updates