Jade PTA Balloon Dilatation Catheter 597004022Recalls & reported problems
OrbusNeich Medical (ShenZhen) Co., Ltd. · Percutaneous Transluminal Angioplasty (PTA) Catheter · Class 2
All models of Jade PTA Balloon Dilatation Catheter (15)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 06934955903991), received between Mar 10, 2026 and Mar 11, 2026 — classified as 2 injuries.
The most frequently reported specific device problem is “Material Separation” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation2
Patient problems
- Foreign Body In Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMar 11, 2026Material Separation
IT WAS REPORTED THAT DURING PROCEDURE, A JADE BALLOON RUPTURED AND GOT TRAPPED IN-VIVO RESULTING IN SURGICAL CUTDOWN FOR REMOVAL OF THE BALLOON. ADDITIONAL INFORMATION WAS RECEIVED ON 2026/02/13: 1. DID THE BALLOON BECOME FULLY DETACHED OR WAS IT DAMAGED INTACT? - PART OF THE BALLOON BECAME DETACHED AS WELL AS THE BALLOON MARKER. THE OTHER HALF OF THE BALLOON WAS REMOVED FROM THE BODY. 2. WAS THERE ANY BALLOON FRAGMENTS LEFT IN-VIVO OR WAS IT COMPLETELY REMOVED? - IT WAS COMPLETELY REMOVED WITH SURGERY. 3. ANATOMICAL INFORMATION? (VESSEL / LOCATION / TORTUOSITY / CALCIFICATION / STENOSIS % / ETC.) - PROXIMAL SFA, HEAVY CALCIFICATION, ~90-95% STENOSIS. 4. WAS THERE A DELAY IN THE PROCEDURE THAT RESULTED IN PATIENT HARM? IF YES, WHAT WAS THE HARM? - NO. 5. WILL THE FACILITY RELEASE ANY IMAGES OR FILMS OF THE INCIDENT FOR REVIEW BY THE ENGINEERS? - NO. 6. WHAT IS THE STATUS OF THE PATIENT? - STABLE UPON COMPLETION OF THE PROCEDURE.
Patient problems: Foreign Body In Patient
MAUDE report 24573476 ↗InjuryMar 10, 2026Material Separation
IT WAS REPORTED THAT DURING PROCEDURE, A JADE BALLOON RUPTURED AND GOT TRAPPED IN-VIVO RESULTING IN SURGICAL CUTDOWN FOR REMOVAL OF THE BALLOON. ADDITIONAL INFORMATION WAS RECEIVED ON 2026/2/13: 1. DID THE BALLOON BECOME FULLY DETACHED OR WAS IT DAMAGED INTACT? PART OF THE BALLOON BECAME DETACHED AS WELL AS THE BALLOON MARKER. THE OTHER HALF OF THE BALLOON WAS REMOVED FROM THE BODY 2. WAS THERE ANY BALLOON FRAGMENTS LEFT IN-VIVO OR WAS IT COMPLETELY REMOVED? IT WAS COMPLETELY REMOVED WITH SURGERY. 3. ANATOMICAL INFORMATION? (VESSEL / LOCATION / TORTUOSITY / CALCIFICATION / STENOSIS % / ETC.) PROXIMAL SFA, HEAVY CALCIFICATION, 90-95% STENOSIS. 4. WAS THERE A DELAY IN THE PROCEDURE THAT RESULTED IN PATIENT HARM? IF YES, WHAT WAS THE HARM? NO 5. WILL THE FACILITY RELEASE ANY IMAGES OR FILMS OF THE INCIDENT FOR REVIEW BY THE ENGINEERS? NO 6. WHAT IS THE STATUS OF THE PATIENT? STABLE UPON COMPLETION OF THE PROCEDURE.
Patient problems: Foreign Body In Patient
MAUDE report 24565781 ↗
Other Catheter, Angioplasty, Peripheral, Transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page