Jade PTA Balloon Dilatation Catheter 596024032Recalls & reported problems

OrbusNeich Medical (ShenZhen) Co., Ltd. · Percutaneous Transluminal Angioplasty (PTA) Catheter · Class 2

All models of Jade PTA Balloon Dilatation Catheter (15)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 06934955904936), received between Apr 25, 2024 and May 13, 2024 — classified as 2 injuries.

The most frequently reported specific device problem is “Material Separation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation1
  • Material Split, Cut or Torn1

Patient problems

  • Foreign Body In Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 13, 2024Material Separation; Material Separation

    THE JADE OTW .018" 200CM (PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY) PTA BALLOON WAS USED FOR TREATMENT OF 99% STENOSED, HEAVILY CALCIFIED, IN-STENT RESTENOSIS (ISR) LESION IN SUPERFICIAL FEMORAL ARTERY (SFA) VIA RADIAL ACCESS 6FR GUIDING SHEATH WAS IN USE. THE NON-CSI/NON-ABBOTT ATHERECTOMY CATHETER HAD DIFFICULTY TRACKING ON THE VIPERWIRE ADVANCE GUIDE WIRE WITH FLEX TIP BUT ATHERECTOMY WAS ABLE TO BE PERFORMED ON THE ISR LESION. THERE WAS NO ALLEGATION OF MALFUNCTION AGAINST THE VIPERWIRE. THE VIPERWIRE AND NON-CSI/NON-ABBOTT ATHERECTOMY CATHETER WERE REMOVED. THE JADE OTW BALLOON WAS ADVANCED OVER THE NON-CSI/NON- ABBOTT WORKHORSE FRONTLINE GUIDE WIRE INTO THE ISR AND INFLATED AT SIXTEEN ATMOSPHERES (ATM). THE NOMINAL PRESSURE IS FOURTEEN ATM AND THE BURSTING PRESSURE IS EIGHTEEN ATM. DURING INFLATION, THE PATIENT BENT THEIR LEG THAT WAS BEING TREATED. THE BALLOON BURST. THE BURST BALLOON HAD THREE COMPONENTS. THE MAJOR PART WAS ATTACHED TO THE CATHETER AND WAS REMOVED ALONG WITH THE SECOND PART. CINE ANGIOGRAPHIC IMAGING AND FLUOROSCOPY REVEALED THE THIRD COMPONENT REMAINED IN THE RIGHT BRACHIAL ARTERY. MECHANICAL ASPIRATION WAS ATTEMPTED VIA RADIAL ACCESS TO SNARE THE FRACTURED C

    Patient problems: Foreign Body In Patient

    MAUDE report 19299306 ↗
  • InjuryApr 25, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    THE JADE OTW .018" 200CM PTA BALLOON WAS USED FOR TREATMENT OF 99% STENOSED, HEAVILY CALCIFIED, IN-STENT RESTENOSIS (ISR) LESION IN SUPERFICIAL FEMORAL ARTERY (SFA) VIA RADIAL ACCESS. 6FR TERUMO GUIDING SHEATH WAS IN USE. THE NON-CSI/NON-ABBOTT ATHERECTOMY CATHETER (AURYON) HAD DIFFICULTY TRACKING ON THE VIPERWIRE ADVANCE GUIDE WIRE WITH FLEX TIP BUT ATHERECTOMY WAS ABLE TO BE PERFORMED ON THE ISR LESION. THERE WAS NO ALLEGATION OF MALFUNCTION AGAINST THE VIPERWIRE. THE VIPERWIRE AND NON-CSI/NON-ABBOTT ATHERECTOMY CATHETER WERE REMOVED. THE JADE OTW BALLOON WAS ADVANCED OVER THE ASAHI MINAMO WORKHORSE FRONTLINE GUIDE WIRE INTO THE ISR AND INFLATED AT SIXTEEN ATMOSPHERES (ATM). THE NOMINAL PRESSURE IS FOURTEEN ATM AND THE BURSTING PRESSURE IS EIGHTEEN ATM. DURING INFLATION, THE PATIENT BENT HIS LEG THAT WAS BEING TREATED. THE BALLOON BURST. THE BURST BALLOON HAD THREE COMPONENTS. THE MAJOR PART WAS ATTACHED TO THE CATHETER AND WAS REMOVED ALONG WITH THE SECOND PART. CINE ANGIOGRAPHIC IMAGING AND FLUOROSCOPY REVEALED THE THIRD COMPONENT REMAINED IN THE RIGHT BRACHIAL ARTERY. MECHANICAL ASPIRATION WAS ATTEMPTED VIA RADIAL ACCESS TO SNARE THE FRACTURED COMPONENT BUT FAILED. MECHANICAL

    Patient problems: Foreign Body In Patient

    MAUDE report 19178419 ↗

Other Catheter, Angioplasty, Peripheral, Transluminal devices

All Catheter, Angioplasty, Peripheral, Transluminal devices →