SIUI Apogee 2300Recalls & reported problems
Shantou Institute of Ultrasonic Instruments Co., Ltd. · Digital Color Doppler Ultrasound Imaging System · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This SIUI Apogee 2300 ultrasound system is used with a probe to guide prostate procedures. Most of the 70 reports describe harm to the rectum during those procedures — cuts, perforations of the bowel wall, and infections, with some patients needing surgery or a colostomy; 2 deaths and 67 injuries were reported. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 71
- Deaths reported
- 2
- Injuries reported
- 68
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 71 adverse event reports that list this exact device (DI 06938396423001), received between Aug 4, 2021 and Sep 11, 2026 — classified as 2 deaths, 68 injuries.
The most frequently reported specific device problem is “Use of Device Problem” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem62
- Use of Device Problem8
- Insufficient Device Problem Information1
Patient problems
- Perforation58
- Laceration(s)8
- Sepsis4
- Bowel Perforation2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
Recent adverse event reports
The 25 most recent of 71 reports, in the reporter's own words as published by the FDA.
InjurySep 11, 2026Adverse Event Without Identified Device or Use Problem
On 29-Jul-2026, a male patient underwent Aquablation therapy for symptomatic benign prostatic hyperplasia (BPH). PROCEPT became aware that, post-Aquablation therapy, the patient presented with a rectal perforation. According to the information received, the night following the procedure, the patient complained of pain. A computed tomography (CT) scan was subsequently performed, which identified a rectal perforation. The patient required additional surgery to repair the perforation and currently has a temporary stoma in place as part of the postoperative management. Additional information indicated no resistance during TRUS probe insertion and no blood on the TRUS probe upon removal. According to the treating surgeon, the confirmed cause of the perforation remains unknown.No malfunction of the Apogee 2300 Digital Color Doppler Ultrasound Imaging System and associated component ECBP-1 TRUS Probe were reported during this event.
Patient problems: Perforation
MAUDE report 50725270 ↗InjuryMay 21, 2026Adverse Event Without Identified Device or Use Problem
A MALE PATIENT UNDERWENT AQUABLATION THERAPY FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS (PROCEPT) BECAME AWARE THAT, DURING AQUABLATION THERAPY, A SMALL LACERATION ON THE OUTSIDE OF THE PATIENT'S RECTUM WAS INADVERTENTLY MADE BY THE RESIDENT DURING TRUS INSERTION. THE RESIDENT HAD DIFFICULTY LOCATING THE PATIENT'S RECTUM, AND THE TREATING SURGEON HAD TO STEP IN. THE TREATING SURGEON WAS MADE AWARE OF THE PRESENCE OF BLOOD, AND A SMALL LACERATION WAS OBSERVED ON THE OUTSIDE OF THE PATIENT'S RECTUM. THE LACERATION WAS CLOSED WITH STITCHES AFTER COMPLETION OF AQUABLATION THERAPY, AND THE PATIENT HAS SINCE BEEN DISCHARGED. NO MALFUNCTION OF THE AQUABEAM ROBOTIC SYSTEM WAS REPORTED.
Patient problems: Laceration(s)
MAUDE report 25247694 ↗InjuryApr 11, 2026Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2026, A MALE PATIENT UNDERWENT AQUABLATION THERAPY FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BECAME AWARE THAT, POST-AQUABLATION THERAPY, THE PATIENT EXPERIENCED RECTAL PERFORATION. DURING FOCAL BLADDER NECK CAUTERY (FBNC), THE TREATING SURGEON ADVANCED THE RESECTION BEYOND THE INTENDED TREATMENT AREA INTO THE MID-PROSTATE. THE POSITIONING AND ANGULATION OF THE TRUS PROBE MAY HAVE CONTRIBUTED TO THIS EVENT. A SMALL RECTAL TEAR WAS IDENTIFIED DURING THE PROCEDURE, AND THE TRUS PROBE WAS VISUALIZED ON CYSTOSCOPY. THE SURGEON NOTED THAT THE PATIENT'S ANATOMY WAS INHERENTLY WEAK AND HAD THIN PROSTATIC TISSUE IN THE MID-PROSTATE REGION, WHICH MAY HAVE INCREASED THE RISK OF THIS OCCURRENCE. NO ADDITIONAL SURGERIES OR INTERVENTIONS WERE REQUIRED FOLLOWING THE AQUABLATION PROCEDURE. THE DECISION WAS MADE TO MANAGE THE PERFORATION CONSERVATIVELY, WITH PROLONGED CATHETERIZATION AND AN EXTENDED INPATIENT STAY FOR MONITORING ON THE WARD. THE PATIENT IS REPORTED TO BE DOING WELL. THE SURGEON DOES NOT ATTRIBUTE THE RECTAL PERFORATION TO THE AQUABLATION THERAPY OR DEVICE ITSELF. NO DEVICE MALFUNCTION WAS REPORTED DURING THIS EVENT.
Patient problems: Perforation
MAUDE report 24852492 ↗InjuryApr 4, 2026Adverse Event Without Identified Device or Use Problem
A MALE PATIENT UNDERWENT AQUABLATION THERAPY FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS (PROCEPT) BECAME AWARE THAT, POST-AQUABLATION THERAPY, BLOOD WAS OBSERVED ON THE TIP OF THE TRUS PROBE UPON REMOVAL. IT WAS REPORTED THAT THE TREATING SURGEON INSERTED THE TRUS PROBE SLIGHTLY BELOW THE RECTUM, CAUSING A CUT IN THE PATIENT'S RECTUM. A STITCH WAS PLACED, AND THE PATIENT WAS ADMITTED OVERNIGHT, DISCHARGED THE NEXT DAY. NO MALFUNCTION OF THE APOGEE 2300 DIGITAL COLOR DOPPLER ULTRASOUND IMAGING SYSTEM AND ASSOCIATED COMPONENT ECBP-1 TRUS PROBE WERE REPORTED DURING THIS EVENT.
Patient problems: Laceration(s)
MAUDE report 24784757 ↗InjuryJan 22, 2026Adverse Event Without Identified Device or Use Problem
A MALE PATIENT UNDERWENT AQUABLATION THERAPY FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION (PROCEPT) BECAME AWARE THAT THE PATIENT HAD A RECTAL PERFORATION. THE PATIENT WAS TRANSFERRED TO A DIFFERENT FACILITY AND RECEIVED A COLOSTOMY. THE PATIENT IS REPORTED TO BE DOING FINE. NO MALFUNCTION OF THE APOGEE 2300 DIGITAL COLOR DOPPLER ULTRASOUND IMAGING SYSTEM AND ASSOCIATED COMPONENT ECBP-1 TRUS PROBE WERE REPORTED DURING THIS EVENT.
Patient problems: Perforation
MAUDE report 24142542 ↗InjuryJan 22, 2026Adverse Event Without Identified Device or Use Problem
A MALE PATIENT UNDERWENT AQUABLATION THERAPY FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BECAME AWARE THAT POST-AQUABLATION THERAPY, THE PATIENT HAD A RECTAL PERFORATION. THE PATIENT WAS THEN TRANSFERRED TO A DIFFERENT FACILITY AND STAYED IN THE INTENSIVE CARE UNIT (ICU). CURRENTLY, THE PATIENT IS IN THE GENERAL SURGERY WARD TO SEE IF THE PERFORATION HEALS ITSELF OR IF SURGICAL REPAIR IS NEEDED. NO MALFUNCTION OF THE APOGEE 2300 DIGITAL COLOR DOPPLER ULTRASOUND IMAGING SYSTEM AND ASSOCIATED COMPONENT ECBP-1 TRUS PROBE WERE REPORTED DURING THIS EVENT.
Patient problems: Perforation
MAUDE report 24142532 ↗InjuryJan 22, 2026Adverse Event Without Identified Device or Use Problem
A MALE PATIENT UNDERWENT AQUABLATION THERAPY FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION (PROCEPT) BECAME AWARE THAT, POST AQUABLATION THERAPY, THE PATIENT WENT SEPTIC. THE PATIENT WAS TREATED WITH ANTIBIOTICS FOR SEPSIS. THE PATIENT RECEIVED AN ABDOMINAL CT (COMPUTED TOMOGRAPHY) SCAN DUE TO THE INFECTION, AND A RECTAL INJURY WAS DETECTED. THE PATIENT THEN UNDERWENT SURGERY, AND AN ARTIFICIAL COLOSTOMY WAS APPLIED. THE PATIENT IS REPORTED TO BE DOING FINE. NO MALFUNCTION OF THE APOGEE 2300 DIGITAL COLOR DOPPLER ULTRASOUND IMAGING SYSTEM AND ASSOCIATED COMPONENT ECBP-1 TRUS PROBE WERE REPORTED DURING THIS EVENT.
Patient problems: Bowel Perforation; Sepsis
MAUDE report 24142508 ↗InjuryJan 22, 2026Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2025, A MALE PATIENT UNDERWENT AQUABLATION THERAPY FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BECAME AWARE THAT POST-AQUABLATION THERAPY, THE PATIENT PRESENTED WITH A FEVER IN POST-OP CARE AND WAS EXPERIENCING TACHYCARDIA. IT WAS DISCOVERED THAT THE PATIENT HAD AIR IN HIS ABDOMEN, AND THE PATIENT WAS THEN SENT FOR A COMPUTED TOMOGRAPHY (CT) IMAGING, REVEALING A RECTAL PERFORATION. ON (B)(6) 2025, A BOWEL DIVERSION PROCEDURE WAS PERFORMED. THE PATIENT WAS DISCHARGED FROM THE HOSPITAL ON (B)(6) 2025. IT WAS REPORTED THAT NO ISSUES WERE EXPERIENCED DURING THE AQUABLATION THERAPY. STANDARD PROTOCOLS WERE FOLLOWED, AND ALL RECTAL INJURY PREVENTION PRECAUTIONS (PATIENT IN LITHOTOMY POSITION, BUTTOCKS 2 INCHES OFF THE END OF THE BED, DIGITAL RECTAL EXAM, 60CCS OF LUBRICANT DISPENSED IN THE RECTUM, 30CCS OF LUBRICANT PLACED ON TH RECTAL ULTRASOUND) WERE PERFORMED. FOLLOWING THE AQUABLATION THERAPY, THE TRUS PROBE WAS INSPECTED, AND NO BLOOD WAS VISUALIZED. FOLLOWING THE FOCAL BLADDER NECK CAUTERY PROTOCOLS, THE PATIENT'S RECTUM WAS INSPECTED BY THE PHYSICIAN FOR ANY INJURY. NO INJURY WAS FOUND UPON INSPECTION. NO MALFUNCTION OF THE APOGEE 2300 DIGITAL COLOR DOPPL
Patient problems: Perforation
MAUDE report 24142489 ↗InjuryJan 22, 2026Adverse Event Without Identified Device or Use Problem
A MALE PATIENT UNDERWENT AQUABLATION THERAPY FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BECAME AWARE THAT THE PATIENT SUSTAINED A RECTAL PERFORATION DURING HEMOSTASIS. IT WAS NOTED THAT THE ECBP-1 TRUS PROBE REMAINED IN PLACE DURING HEMOSTASIS AND THE TREATING SURGEON RE-POSITIONED THE ECBP-1 TRUS PROBE APPROXIMATELY THREE (3) TIMES TO GIVE MORE SPACE FOR THE RESECTOSCOPE. DURING LOOP RESECTION AT THE POSTERIOR MID-FOSSA, THE TIP OF THE ECBP-1 TRUS PROBE WAS OBSERVED TO BE PROTRUDING THROUGH THE RECTAL WALL, INDICATING A RECTAL INJURY. EVALUATION FROM A DIFFERENT PHYSICIAN REPORTED THERE WAS A ~1CM OPENING. THE TREATING SURGEON CONFIRMED THAT THE PATIENT RECEIVED A ROBOTIC REPAIR THE SAME DAY AND WAS DISCHARGED THE FOLLOWING DAY. NO MALFUNCTION OF THE APOGEE 2300 DIGITAL COLOR DOPPLER ULTRASOUND IMAGING SYSTEM AND ASSOCIATED COMPONENT ECBP-1 TRUS PROBE WERE REPORTED DURING THIS EVENT.
Patient problems: Perforation
MAUDE report 24142439 ↗InjuryJan 22, 2026Adverse Event Without Identified Device or Use Problem
A MALE PATIENT UNDERWENT AQUABLATION THERAPY FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION (PROCEPT) BECAME AWARE OF A RECTAL MUCOSAL TEAR WITH THE APOGEE 2300 DIGITAL COLOR DOPPLER ULTRASOUND IMAGING SYSTEM AND ITS ASSOCIATED ECBP-1 TRUS PROBE. THE TREATING SURGEON REPORTED THAT BLEEDING WAS NOTED AFTER ECBP-1 TRUS PROBE INSERTION. THE TREATING SURGEON DID A RECTAL EXAM POST ECBP-1 TRUS PROBE REMOVAL AND NOTED SMALL EXTERNAL MUCOSAL TEAR NEAR ANUS. THE TREATING SURGEON NOTED NO ISSUES WHEN DOING AN INTERNAL RECTAL FINGER SWEEP. NO ADDITIONAL INTERVENTIONS OR CARE WERE NEEDED. THE TREATING SURGEON IS UNSURE WHY THERE WAS A SMALL MUCOSAL TEAR, AS NO RESISTANCE WAS ENCOUNTERED WHEN ENTERING THE PATIENT (SUSPECTED IT COULD BE DUE TO THE PATIENT BEING OLDER AND HAVING FRAGILE SKIN). THE TREATING SURGEON WILL BE SEEING THE PATIENT IN 2 WEEKS FOR A FOLLOW-UP. NO MALFUNCTION OF THE APOGEE 2300 DIGITAL COLOR DOPPLER ULTRASOUND IMAGING SYSTEM AND ASSOCIATED COMPONENT ECBP-1 TRUS PROBE WERE REPORTED DURING THIS EVENT.
Patient problems: Laceration(s)
MAUDE report 24142168 ↗InjuryDec 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A MALE PATIENT UNDERWENT AQUABLATION THERAPY FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION (PROCEPT) BECAME AWARE THAT THE PATIENT HAD A RECTAL PERFORATION. THE PATIENT WAS TRANSFERRED TO A DIFFERENT FACILITY AND RECEIVED A COLOSTOMY. THE PATIENT IS REPORTED TO BE DOING FINE. NO MALFUNCTION OF THE APOGEE 2300 DIGITAL COLOR DOPPLER ULTRASOUND IMAGING SYSTEM AND ASSOCIATED COMPONENT ECBP-1 TRUS PROBE WERE REPORTED DURING THIS EVENT.
Patient problems: Perforation
MAUDE report 23758825 ↗InjuryNov 21, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A MALE PATIENT UNDERWENT AQUABLATION THERAPY FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PRECEPT BECAME AWARE THAT POST-AQUABLATION THERAPY, THE PATIENT HAD A RECTAL PERFORATION. THE PATIENT WAS THEN TRANSFERRED TO A DIFFERENT FACILITY AND STAYED IN THE INTENSIVE CARE UNIT (ICU). CURRENTLY, THE PATIENT IS IN THE GENERAL SURGERY WARD TO SEE IF THE PERFORATION HEALS ITSELF OR IF SURGICAL REPAIR IS NEEDED. NO MALFUNCTION OF THE APOGEE 2300 DIGITAL COLOR DOPPLER ULTRASOUND IMAGING SYSTEM AND ASSOCIATED COMPONENT ECBP-1 TRUS PROBE WERE REPORTED DURING THIS EVENT.
Patient problems: Perforation
MAUDE report 23623132 ↗InjuryOct 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A MALE PATIENT UNDERWENT AQUABLATION THERAPY FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION (PROCEPT) BECAME AWARE THAT, POST AQUABLATION THERAPY, THE PATIENT WENT SEPTIC. THE PATIENT WAS TREATED WITH ANTIBIOTICS FOR SEPSIS. THE PATIENT RECEIVED AN ABDOMINAL CT (COMPUTED TOMOGRAPHY) SCAN DUE TO THE INFECTION, AND A RECTAL INJURY WAS DETECTED. THE PATIENT THEN UNDERWENT SURGERY, AND AN ARTIFICIAL COLOSTOMY WAS APPLIED. THE PATIENT IS REPORTED TO BE DOING FINE. NO MALFUNCTION OF THE APOGEE 2300 DIGITAL COLOR DOPPLER ULTRASOUND IMAGING SYSTEM AND ASSOCIATED COMPONENT ECBP-1 TRUS PROBE WERE REPORTED DURING THIS EVENT.
Patient problems: Bowel Perforation; Sepsis
MAUDE report 23375417 ↗InjuryJun 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A MALE PATIENT UNDERWENT AQUABLATION THERAPY FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION (PROCEPT) BECAME AWARE OF A RECTAL MUCOSAL TEAR WITH THE APOGEE 2300 DIGITAL COLOR DOPPLER ULTRASOUND IMAGING SYSTEM AND ITS ASSOCIATED ECBP-1 TRUS PROBE. THE TREATING SURGEON REPORTED THAT BLEEDING WAS NOTED AFTER ECBP-1 TRUS PROBE INSERTION. THE TREATING SURGEON DID A RECTAL EXAM POST ECBP-1 TRUS PROBE REMOVAL AND NOTED SMALL EXTERNAL MUCOSAL TEAR NEAR ANUS. THE TREATING SURGEON NOTED NO ISSUES WHEN DOING AN INTERNAL RECTAL FINGER SWEEP. NO ADDITIONAL INTERVENTIONS OR CARE WERE NEEDED. THE TREATING SURGEON IS UNSURE WHY THERE WAS A SMALL MUCOSAL TEAR, AS NO RESISTANCE WAS ENCOUNTERED WHEN ENTERING THE PATIENT (SUSPECTED IT COULD BE DUE TO THE PATIENT BEING OLDER AND HAVING FRAGILE SKIN). THE TREATING SURGEON WILL BE SEEING THE PATIENT IN 2 WEEKS FOR A FOLLOW-UP. NO MALFUNCTION OF THE APOGEE 2300 DIGITAL COLOR DOPPLER ULTRASOUND IMAGING SYSTEM AND ASSOCIATED COMPONENT ECBP-1 TRUS PROBE WERE REPORTED DURING THIS EVENT.
Patient problems: Laceration(s)
MAUDE report 22324791 ↗OtherJun 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A MALE PATIENT UNDERWENT AQUABLATION THERAPY FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BECAME AWARE THAT THE PATIENT SUSTAINED A RECTAL PERFORATION DURING HEMOSTASIS. IT WAS NOTED THAT THE ECBP-1 TRUS PROBE REMAINED IN PLACE DURING HEMOSTASIS AND THE TREATING SURGEON RE-POSITIONED THE ECBP-1 TRUS PROBE APPROXIMATELY THREE (3) TIMES TO GIVE MORE SPACE FOR THE RESECTOSCOPE. DURING LOOP RESECTION AT THE POSTERIOR MID-FOSSA, THE TIP OF THE ECBP-1 TRUS PROBE WAS OBSERVED TO BE PROTRUDING THROUGH THE RECTAL WALL, INDICATING A RECTAL INJURY. EVALUATION FROM A DIFFERENT PHYSICIAN REPORTED THERE WAS A ~1CM OPENING. THE TREATING SURGEON CONFIRMED THAT THE PATIENT RECEIVED A ROBOTIC REPAIR THE SAME DAY AND WAS DISCHARGED THE FOLLOWING DAY. NO MALFUNCTION OF THE APOGEE 2300 DIGITAL COLOR DOPPLER ULTRASOUND IMAGING SYSTEM AND ASSOCIATED COMPONENT ECBP-1 TRUS PROBE WERE REPORTED DURING THIS EVENT.
Patient problems: Perforation
MAUDE report 22288365 ↗InjuryJun 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A MALE PATIENT UNDERWENT AQUABLATION THERAPY FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BECAME AWARE THAT A DAY POST-AQUABLATION THERAPY, THE PATIENT EXPRESSED ABDOMINAL PAIN TO THE TREATING SURGEON. A CT SCAN WAS PERFORMED WHICH CONFIRMED THAT THE OBSERVED FLUID WAS RELATED TO AN INFECTION. THE TREATING SURGEON STATED THAT THE PATIENT HAD A RECTAL PERFORATION AND REQUIRED A COLOSTOMY DIVERSION. NO RESISTANCE DURING THE INSERTION OF THE TRUS PROBE WAS REPORTED. THE PATIENT REQUIRED EXTENDED HOSPITALIZATION. NO MALFUNCTION OF THE APOGEE 2300 DIGITAL COLOR DOPPLER ULTRASOUND IMAGING SYSTEM AND ASSOCIATED COMPONENT ECBP-1 TRUS PROBE WERE REPORTED DURING THIS EVENT.
Patient problems: Perforation
MAUDE report 22129266 ↗InjuryMay 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A MALE PATIENT UNDERWENT AQUABLATION THERAPY FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BECAME AWARE THAT A DAY POST-AQUABLATION THERAPY, THE PATIENT EXPRESSED ABDOMINAL PAIN TO THE TREATING SURGEON. A CT SCAN WAS PERFORMED WHICH CONFIRMED THAT THE OBSERVED FLUID WAS RELATED TO AN INFECTION. THE TREATING SURGEON STATED THAT THE PATIENT HAD A RECTAL PERFORATION AND REQUIRED A COLOSTOMY DIVERSION. NO RESISTANCE DURING THE INSERTION OF THE TRUS PROBE WAS REPORTED. THE PATIENT REQUIRED EXTENDED HOSPITALIZATION. NO MALFUNCTION OF THE APOGEE 2300 DIGITAL COLOR DOPPLER ULTRASOUND IMAGING SYSTEM AND ASSOCIATED COMPONENT ECBP-1 TRUS PROBE WERE REPORTED DURING THIS EVENT.
Patient problems: Perforation
MAUDE report 22054911 ↗InjuryMar 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A MALE PATIENT UNDERWENT AQUABLATION THERAPY FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION (PROCEPT) BECAME AWARE OF A RECTAL MUCOSAL TEAR WITH THE APOGEE 2300 DIGITAL COLOR DOPPLER ULTRASOUND IMAGING SYSTEM AND ITS ASSOCIATED ECBP-1 TRUS PROBE AFTER AQUABLATION THERAPY HAD BEEN COMPLETED. BLOOD WAS NOTED ON THE PROBE WHEN THE TREATING SURGEON REMOVED THE ECBP-1 TRUS PROBE. THE TREATING SURGEON PROCEEDED TO PERFORM A DIGITAL RECTAL EXAM AND OBSERVED BLOOD ON HIS GLOVE. A COLORECTAL SURGEON WAS BROUGHT IN AND IDENTIFIED A MUCOSAL TEAR IN THE RECTUM. THE MUCOSAL TEAR WAS SUTURED. THE PATIENT WAS DISCHARGED HOME THE NEXT DAY. THE TREATING SURGEON REPORTED THAT NO RESISTANCE WAS MET WHEN THE ECBP-1 TRUS PROBE WAS INSERTED IN THE PATIENT AND DID NOT MENTION THAT THERE WAS ANYTHING UNUSUAL WITH THE PATIENT'S ANATOMY. NO MALFUNCTION OF THE APOGEE 2300 DIGITAL COLOR DOPPLER ULTRASOUND IMAGING SYSTEM AND ASSOCIATED COMPONENT ECBP-1 TRUS PROBE WERE REPORTED DURING THIS EVENT.
Patient problems: Laceration(s)
MAUDE report 21671182 ↗InjuryMar 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A MALE PATIENT UNDERWENT AQUABLATION THERAPY FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION (PROCEPT) BECAME AWARE OF A RECTAL MUCOSAL TEAR WITH THE APOGEE 2300 DIGITAL COLOR DOPPLER ULTRASOUND IMAGING SYSTEM AND ITS ASSOCIATED ECBP-1 TRUS PROBE AFTER AQUABLATION THERAPY HAD BEEN COMPLETED. BLOOD WAS NOTED ON THE PROBE WHEN THE TREATING SURGEON REMOVED THE ECBP-1 TRUS PROBE. THE TREATING SURGEON PROCEEDED TO PERFORM A DIGITAL RECTAL EXAM AND OBSERVED BLOOD ON HIS GLOVE. A COLORECTAL SURGEON WAS BROUGHT IN AND IDENTIFIED A MUCOSAL TEAR IN THE RECTUM. THE MUCOSAL TEAR WAS SUTURED. THE PATIENT WAS DISCHARGED HOME THE NEXT DAY. THE TREATING SURGEON REPORTED THAT NO RESISTANCE WAS MET WHEN THE ECBP-1 TRUS PROBE WAS INSERTED IN THE PATIENT AND DID NOT MENTION THAT THERE WAS ANYTHING UNUSUAL WITH THE PATIENT'S ANATOMY. NO MALFUNCTION OF THE APOGEE 2300 DIGITAL COLOR DOPPLER ULTRASOUND IMAGING SYSTEM AND ASSOCIATED COMPONENT ECBP-1 TRUS PROBE WERE REPORTED DURING THIS EVENT.
Patient problems: Laceration(s)
MAUDE report 21659829 ↗InjuryDec 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A MALE PATIENT UNDERWENT AQUABLATION THERAPY FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BECAME AWARE THAT DURING HEMOSTASIS, THE TREATING SURGEON NOTED THAT THE PATIENT'S RECTAL WALL WAS PERFORATED. AN ILEOSTOMY PROCEDURE WAS PERFORMED TO REPAIR THE PERFORATION. THE PATIENT HAS BEEN DISCHARGED AND WILL HAVE THE OSTOMY FOR APPROXIMATELY THREE MONTHS. IT WAS REPORTED THAT THERE WAS NOTHING UNUSUAL ABOUT THE PATIENT'S ANATOMY. NO RESISTANCE DURING INSERTION OF THE TRUS PROBE WAS REPORTED. NO MALFUNCTION OF THE APOGEE 2300 DIGITAL COLOR DOPPLER ULTRASOUND IMAGING SYSTEM AND ASSOCIATED COMPONENT ECBP-1 TRUS PROBE WERE REPORTED DURING THIS EVENT.
Patient problems: Perforation
MAUDE report 21023673 ↗InjuryNov 8, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2024, A MALE PATIENT UNDERWENT AQUABLATION THERAPY FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BECAME AWARE THAT POST AQUABLATION THERAPY, THE PATIENT COMPLAINED OF PAIN. UPON INSPECTION, THE PATIENT'S ABDOMEN WAS NOTED TO BE DISTENDED AND COMPUTED TOMOGRAPHY (CT) IMAGING WITH CONTRACT WAS PERFORMED, REVEALING A RECTAL PERFORATION. ON (B)(6) 2024, AN EXPLORATORY LAPAROTOMY AND SUBSEQUENT BOWEL DIVERSION PROCEDURE WERE PERFORMED. THE PATIENT REMAINS HOSPITALIZED UNDER OBSERVATION AND A DISCHARGE PLAN HAS NOT YET BEEN DETERMINED. IT WAS REPORTED THAT NO ISSUES WERE EXPERIENCED DURING THE AQUABLATION THERAPY, AND THE PATIENT DID NOT HAVE ANYTHING UNUSUAL WITH HIS ANATOMY. NO MALFUNCTION OF THE APOGEE 2300 DIGITAL COLOR DOPPLER ULTRASOUND IMAGING SYSTEM AND ASSOCIATED COMPONENT ECBP-1 TRUS PROBE WERE REPORTED DURING THIS EVENT.
Patient problems: Perforation
MAUDE report 20637628 ↗InjuryOct 31, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2024, A MALE PATIENT UNDERWENT AQUABLATION THERAPY FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BECAME AWARE THAT POST AQUABLATION THERAPY, THE PATIENT COMPLAINED OF PAIN. UPON INSPECTION, THE PATIENT¿S ABDOMEN WAS NOTED TO BE DISTENDED AND COMPUTED TOMOGRAPHY (CT) IMAGING WITH CONTRACT WAS PERFORMED, REVEALING A RECTAL PERFORATION. ON (B)(6) 2024, AN EXPLORATORY LAPAROTOMY AND SUBSEQUENT BOWEL DIVERSION PROCEDURE WERE PERFORMED. THE PATIENT REMAINS HOSPITALIZED UNDER OBSERVATION AND A DISCHARGE PLAN HAS NOT YET BEEN DETERMINED. IT WAS REPORTED THAT NO ISSUES WERE EXPERIENCED DURING THE AQUABLATION THERAPY, AND THE PATIENT DID NOT HAVE ANYTHING UNUSUAL WITH HIS ANATOMY. NO MALFUNCTION OF THE APOGEE 2300 DIGITAL COLOR DOPPLER ULTRASOUND IMAGING SYSTEM AND ASSOCIATED COMPONENT ECBP-1 TRUS PROBE WERE REPORTED DURING THIS EVENT.
Patient problems: Perforation
MAUDE report 20578645 ↗InjuryOct 15, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2024, A MALE PATIENT WITH A HISTORY OF RECTUM DIVERTICULUM UNDERWENT AQUABLATION THERAPY FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BECAME AWARE THAT TWO DAYS AFTER POST-AQUABLATION THERAPY, THE PATIENT BEGAN TO EXPERIENCE ABDOMINAL PAIN, AND A DIAGNOSTIC PROCEDURE WAS PERFORMED. THE PATIENT WAS FOUND TO HAVE A RECTAL PERFORATION. THE PERFORATION WAS SURGICALLY REPAIRED. IT WAS REPORTED THAT THE CURRENT STATUS OF THE PATIENT IS UNKNOWN; HOWEVER, HE HAS BEEN DISCHARGED. IT WAS REPORTED THAT NO ISSUES WERE EXPERIENCED DURING THE AQUABLATION THERAPY, AND NO RESISTANCE WAS EXPERIENCED DURING ECBP-1 TRUS PROBE INSERTION. NO BLOOD WAS NOTED ON THE ECBP-1 TRUS PROBE WHEN IT WAS REMOVED FROM THE PATIENT'S RECTUM. NO MALFUNCTION OF THE APOGEE 2300 DIGITAL COLOR DOPPLER ULTRASOUND IMAGING SYSTEM AND ASSOCIATED COMPONENT ECBP-1 TRUS PROBE WERE REPORTED DURING THIS EVENT.
Patient problems: Perforation
MAUDE report 20457955 ↗InjuryOct 4, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2024, A MALE PATIENT WITH A HISTORY OF RECTUM DIVERTICULUM UNDERWENT AQUABLATION THERAPY FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BECAME AWARE THAT TWO DAYS AFTER POST-AQUABLATION THERAPY, THE PATIENT BEGAN TO EXPERIENCE ABDOMINAL PAIN, AND A DIAGNOSTIC PROCEDURE WAS PERFORMED. THE PATIENT WAS FOUND TO HAVE A RECTAL PERFORATION. THE PERFORATION WAS SURGICALLY REPAIRED. IT WAS REPORTED THAT THE CURRENT STATUS OF THE PATIENT IS UNKNOWN; HOWEVER, HE HAS BEEN DISCHARGED. IT WAS REPORTED THAT NO ISSUES WERE EXPERIENCED DURING THE AQUABLATION THERAPY, AND NO RESISTANCE WAS EXPERIENCED DURING ECBP-1 TRUS PROBE INSERTION. NO BLOOD WAS NOTED ON THE ECBP-1 TRUS PROBE WHEN IT WAS REMOVED FROM THE PATIENT¿S RECTUM. NO MALFUNCTION OF THE APOGEE 2300 DIGITAL COLOR DOPPLER ULTRASOUND IMAGING SYSTEM AND ASSOCIATED COMPONENT ECBP-1 TRUS PROBE WERE REPORTED DURING THIS EVENT.
Patient problems: Perforation
MAUDE report 20377801 ↗InjuryJun 1, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2024 A MALE PATIENT UNDERWENT AQUABLATION THERAPY FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). ON 01-MAY-2024 PROCEPT BIOROBOTICS CORPORATION (PROCEPT) BECAME AWARE OF AN UNSPECIFIED ADVERSE EVENT WHERE THE PATIENT WAS REPORTED TO HAVE UNDERGONE A COLORECTAL PROCEDURE FOLLOWING AQUABLATION THERAPY. AT THIS TIME, PROCEPT HAS BEEN UNABLE TO OBTAIN FURTHER INFORMATION, BUT IT IS SUSPECTED THAT A RECTAL INJURY OCCURRED, RESULTING IN THE COLORECTAL PROCEDURE. THE ONLY INFORMATION PROVIDED TO PROCEPT IS THAT THE PATIENT HAS BEEN DISCHARGED HOME AND THE TREATING SURGEON WAS UNABLE TO DETERMINE WHAT CAUSED THE INJURY. NO MALFUNCTION OF THE APOGEE 2300 DIGITAL COLOR DOPPLER ULTRASOUND IMAGING SYSTEM AND ASSOCIATED COMPONENT ECBP-1 TRUS PROBE WERE REPORTED DURING THIS EVENT.
Patient problems: Perforation
MAUDE report 19443106 ↗
FDA clearance and classification
Cleared through 510(k) K173000 on Dec 7, 2017 (decision: Substantially Equivalent). Submitted by Shantou Institute of Ultrasonic Instruments Co., Ltd. (Siui) as “Apogee 2100 Digital Color Doppler Ultrasound Imaging System Apogee 2300 Digital Color Doppler Ultrasound Imaging System”. The FDA review took 71 days. FDA record ↗
The FDA regulates product code IYN (system, imaging, pulsed doppler, ultrasonic) under 21 CFR 892.1550 as a Class II device. 21 CFR 892.1550 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the SIUI Apogee 2300
Has the SIUI Apogee 2300 been recalled?
No FDA recall record in our data names this exact device identifier (DI 06938396423001), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the SIUI Apogee 2300?
The FDA's MAUDE database lists 71 adverse event reports for this device, received between Aug 4, 2021 and Sep 11, 2026. In those reports, the most common specific device problem is “Use of Device Problem” (8); the most common patient problems are “Perforation” (58), “Laceration(s)” (8) and “Sepsis” (4). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the SIUI Apogee 2300?
2 reports are classified as deaths, 68 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the SIUI Apogee 2300 FDA approved?
It was cleared through 510(k) K173000 on Dec 7, 2017. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the SIUI Apogee 2300 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the SIUI Apogee 2300, and is it still sold?
The FDA's device database lists Shantou Institute of Ultrasonic Instruments Co., Ltd. as the company responsible for it (brand SIUI, model Apogee 2300). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other System, Imaging, Pulsed Doppler, Ultrasonic devices
- Transducer C5-1 989605412042Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic11 recalls0 reports
- Transducer X5-1 989605400802Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic10 recalls0 reports
- EPIQ 7 Diagnostic Ultrasound System 989605386721Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic3 recalls2 deaths143 reports
- Transducer S7-3t 989605406772Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic5 recalls1 death18 reports
- EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE OLYMPUS GF TYPE UCT180Olympus Medical Systems Corp. · System, imaging, pulsed doppler, ultrasonic194 reports
- Transducer S7-3t 989605439232Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic3 recalls4 reports
- Transducer 3D9-3v 989605395622Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic3 recalls1 report
- ACUSON Redwood Diagnostic Ultrasound System 11503314Siemens Medical Solutions USA, Inc. · System, imaging, pulsed doppler, ultrasonic3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.