Skip to content

FLOW-i C40 6677400Recalls & reported problems

Maquet Critical Care AB · Class 2

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
4

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07325710001318), received between Oct 31, 2023 and Apr 29, 2026 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Output Problem” (5 reports).

4 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 24, 2026Open, Classified

    Events observed where anesthetic agent was not delivered when the vaporizer was activated. Failure to deliver the anesthetic agent may result in the patient receiving insufficient anesthesia and may lead to intraoperative awareness.

    Z-2860-2026 · Root cause: Under Investigation by firm · FDA record ↗

  2. Dec 11, 2023Open, Classified

    Potential cybersecurity vulnerability that could lead to a Denial of Service (DoS) attack, tampering, or remote code execution via remote login to the Flow Anesthesia System, any of which could lead to a hazardous situation.

    Z-0989-2024 · Root cause: Software design · FDA record ↗

  3. Dec 22, 2022Open, Classified

    Due to a software bug, under certain conditions, pressure cannot be built up resulting in no ventilation. If gas delivery is stopped, a sustained decrease in delivered O2 concentration may lead to hypoventilation and hypoxia.

    Z-0959-2023 · Root cause: Software Design Change · FDA record ↗

  4. Jan 29, 2021Terminated

    Maquet Critical Care AB received complaints where the FiCO2 value was not zero or as close to zero as expected and/or experienced System Check Out failures when docking the absorber to the Flow i Anesthesia machine.

    Z-1181-2021 · Root cause: Device Design · FDA record ↗

Most reported problems

Device problems

  • Output Problem5
  • Failure of Device to Self-Test1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 29, 2026Output Problem

    IT WAS REPORTED THAT THE SYSTEM STOPPED TO DELIVER ANESTHETIC AGENT DURING SOME PART OF THE TREATMENT. THERE WAS NO PATIENT HARM. MANUFACTURER'S REFERENCE #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25024954 ↗
  • MalfunctionMay 15, 2025Failure of Device to Self-Test; Output Problem; Failure of Device to Self-Test; Output Problem

    IT WAS REPORTED THAT THE ANESTHESIA SYSTEM FAILED DURING THE SYSTEM CHECKOUT TEST. ACCORDING TO THE RECEIVED SYSTEM DEVICE LOGS, TWO TECHNICAL ERROR CODES INDICATING AIRWAY PRESSURE SENSOR ERROR HAD BEEN GENERATED. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REFERENCE #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22029138 ↗
  • MalfunctionJan 17, 2024Output Problem; Output Problem

    MANUFACTURER'S REF: #(B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18524183 ↗
  • MalfunctionNov 6, 2023Output Problem; Output Problem

    MANUFACTURER'S REFERENCE NUMBER: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18073008 ↗
  • MalfunctionOct 31, 2023Output Problem; Output Problem

    MANUFACTURER'S REFERENCE NUMBER: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18038269 ↗

Other Gas-Machine, Anesthesia devices

All Gas-Machine, Anesthesia devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.