Provox ActiValve Strong 6 mm 7161Recalls & reported problems
Atos Medical AB · The Provox ActiValve voice prosthesis has a one-way valve and two retention flanges. A rigid blue ring sits inside the prosthesis, adding stability and providing an even sealing surface for the valve flap. The blue ring and valve flap can be seen on X-ray. Magnets in the ring and valve flap determine the force needed to open the valve (the magnets are not adjustable). Provox ActiValve comes in different opening forces. · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07331791000584), received between Sep 26, 2024 and Feb 18, 2026 — classified as 3 injuries.
The most frequently reported specific device problem is “Fracture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture2
- Adverse Event Without Identified Device or Use Problem1
- Insufficient Device Problem Information1
Patient problems
- Aspiration/Inhalation3
- Localized Skin Lesion1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryFeb 18, 2026Fracture
ACCORDING TO THE AVAILABLE INFORMATION, THIS COMPLAINT CONCERNS AN END USER OF A PROVOX ACTIVALVE STRONG VOICE PROSTHESIS (22.5FR/6MM) DURING USE THE DEVICE WAS PLACED IN (B)(6) 2025. APPROXIMATELY FOUR MONTHS LATER, THE PATIENT REPORTED TO THE SPEECH LANGUAGE PATHOLOGIST (SLP) THAT THE SMALL BLUE CENTER PIECE DETACHED WHILE DRINKING AND WAS SWALLOWED BY THE PATIENT. THE PATIENT REPORTED FEELING THE PIECE, SPITTING IT BACK UP, AND THEN EXPERIENCING CHOKING DURING THE NEXT SIP OF LIQUID. THE PATIENT REPORTED THAT WITHOUT THE SMALL BLUE CENTER PIECE, HE WAS CHOKING AND UNABLE TO VOICE/COMMUNICATE. THE SPEECH LANGUAGE PATHOLOGIST (SLP) EVALUATED THE PATIENT THE FOLLOWING MORNING IN CLINIC, REMOVED THE DEVICE, AND PLACED A NEW, DIFFERENT DEVICE. WITHOUT THE BENEFIT OF EXAMINATION AND TESTING, COLOPLAST IS PRECLUDED FROM COMMENTING ON THE CONDITION OF THE DEVICE OR THE CAUSE OF THE OCCURRENCE. SHOULD ADDITIONAL FACTS PROMPT US TO ALTER OR SUPPLEMENT ANY INFORMATION OR CONCLUSIONS CONTAINED IN THE ORIGINAL MDR OR IN ANY PRIOR SUPPLEMENTAL REPORTS, A FOLLOW-UP REPORT WILL BE SUBMITTED.
Patient problems: Aspiration/Inhalation
MAUDE report 24380279 ↗InjuryFeb 17, 2026Fracture
ACCORDING TO THE AVAILABLE INFORMATION, THIS COMPLAINT CONCERNS AN END USER OF A PROVOX ACTIVALVE STRONG VOICE PROSTHESIS (22.5FR/6MM) DURING USE THE DEVICE WAS PLACED IN (B)(6) 2025. APPROXIMATELY FOUR MONTHS LATER, THE PATIENT REPORTED TO THE SPEECH LANGUAGE PATHOLOGIST (SLP) THAT THE SMALL BLUE CENTER PIECE DETACHED WHILE DRINKING AND WAS SWALLOWED BY THE PATIENT. THE PATIENT REPORTED FEELING THE PIECE, SPITTING IT BACK UP, AND THEN EXPERIENCING CHOKING DURING THE NEXT SIP OF LIQUID. THE PATIENT REPORTED THAT WITHOUT THE SMALL BLUE CENTER PIECE, HE WAS CHOKING AND UNABLE TO VOICE/COMMUNICATE. THE SPEECH LANGUAGE PATHOLOGIST (SLP) EVALUATED THE PATIENT THE FOLLOWING MORNING IN CLINIC, REMOVED THE DEVICE, AND PLACED A NEW, DIFFERENT DEVICE. WITHOUT THE BENEFIT OF EXAMINATION AND TESTING, COLOPLAST IS PRECLUDED FROM COMMENTING ON THE CONDITION OF THE DEVICE OR THE CAUSE OF THE OCCURRENCE. SHOULD ADDITIONAL FACTS PROMPT US TO ALTER OR SUPPLEMENT ANY INFORMATION OR CONCLUSIONS CONTAINED IN THE ORIGINAL MDR OR IN ANY PRIOR SUPPLEMENTAL REPORTS, A FOLLOW-UP REPORT WILL BE SUBMITTED.
Patient problems: Aspiration/Inhalation; Localized Skin Lesion
MAUDE report 24365489 ↗InjurySep 26, 2024Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
THE PATIENT WAS COMMITTED TWICE TO THE EMERGENCY DEPARTMENT ON AN EMERGENCY BASIS DUE TO SUSPECTED ASPIRATION OF FOREIGN BODY (PROVOX ACTIVALVE STRONG) INTO THE RESPIRATORY TRACT. FIRST TIME 2024-08-17 AN ENDOSCOPY THROUGH THE PROTHESIS AND A CHEST X-RAY WERE PERFORMED, NO FOREIGN BODY WAS VISIBLE IN RESPIRATORY TRACT. SECOND TIME 2024-08-19 PATIENT COMPLAINED OF SHORTNESS OF BREATH AFTER CLEANING THE DENTURE, CLAIMING THAT HE PROBABLY ASPIRATED WHOLE PROVOX ACTIVALVE STRONG. DUE TO STRONG SUSPICIOUS OF FOREIGN BODY ASPIRATION INTO THE RESPIRATORY TRACT CHEST CT WAS PERFORMED AND PROVOX ACTIVALVE WAS REMOVED SAFETY FROM LEFT BRONCHUS MAIN. AFTER HOSPITALIZATION (2 DAYS) PATIENT WAS ADMITTED HOME.
Patient problems: Aspiration/Inhalation
MAUDE report 20307994 ↗
Other Prosthesis, Laryngeal (Taub) devices
- InHealth Technologies; Blom-Singer; Classic IN 1610-LFFreudenberg Medical, LLC · Prosthesis, Laryngeal (Taub)1 recall0 reports
- InHealth Technologies; Blom-Singer; Classic IN 1610-LEFFreudenberg Medical, LLC · Prosthesis, Laryngeal (Taub)1 recall0 reports
- Provox2 Voice Prosthesis 8mm 7218Atos Medical AB · Prosthesis, Laryngeal (Taub)5 reports
- Provox Vega 22.5Fr 6mm 8283Atos Medical AB · Prosthesis, Laryngeal (Taub)2 reports
- Provox Vega 17Fr 6mm 8271Atos Medical AB · Prosthesis, Laryngeal (Taub)2 reports
- Provox ActiValve Strong 4.5 mm 7160Atos Medical AB · Prosthesis, Laryngeal (Taub)2 reports
- Provox Vega 22.5Fr 10mm 8285Atos Medical AB · Prosthesis, Laryngeal (Taub)2 reports
- Provox Vega 20Fr 6mm 8277Atos Medical AB · Prosthesis, Laryngeal (Taub)1 death2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗