Skip to content

Provox Vega 22.5Fr 10mm 8285Recalls & reported problems

Atos Medical AB · Provox Vega voice prosthesis is an indwelling, low resistance voice prosthesis intended for voice restoration after surgical removal of the larynx. The Provox Insertion System is a single-use replacement device intended for anterograde insertion only. · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07331791010897), received between Dec 14, 2023 and Dec 12, 2024 — classified as 2 injuries.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Insufficient Device Problem Information1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Foreign Body In Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 12, 2024Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    THE PATIENT STATED THAT HE/SHE HAS STARTED TO HAVE AN ISSUE OF VP DISLODGEMENT INTO THE FOOD PATH SIDE 3 TIMES IN 1 MONTH WHILE EATING FOOD. DISLODGEMENT OF VOICE PROSTHESIS OCCURRED WHEN PATIENT WAS EATING MEALS AND AT THE TIME OF COUGHING. THIS REPORT IS FOR THE FIRST DISLODGEMENT. THE DETAILS ARE AS FOLLOWS: FIRST DISLODGEMENT: (B)(6) 2024. SECOND DISLODGEMENT: ABOUT 1 WEEK AFTER (B)(6) 2024. THIRD DISLODGEMENT: (B)(6) 2024. THE DOCTOR MENTIONED THAT THE PATIENT HAS LOST SOME WEIGHT RECENTLY. USERS SUFFERED NO HARM IN CONNECTION WITH THIS COMPLAINT, HOWEVER, MEDICAL INTERVENTION WAS REQUIRED FOR CHECKING WHERE THE VP WAS BY MEANS OF X-RAY. IT WAS CONFIRMED THAT VP WAS IN THE STOMACH EACH TIME. LOT NUMBER IS NOT KNOWN. IT IS CONFIRMED THAT THE BOX HAS BEEN DISCARDED. ADDITIONAL INFORMATION 26 NOVEMBER: THE DISLODGEMENT OCCURRED DURING EATING MEALS AT THE TIME OF COUGHING. THE 1ST TIME REINSERTED 10 MM VEGA VP AND CHECKED IF THE VP IS INSIDE THE STOMACH BY MEANS OF X-RAY. THE 2ND TIME REINSERTED 12.5MM VEGA VP. THE 3RD TIME REINSERTED 10MM VEGA XTRASEAL VP. THE DOCTOR STATED THAT THE PATIENT CLAIMED INCREASED PRODUCTION OF MUCUS DURING EATING.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 20910953 ↗
  • InjuryDec 14, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PROVOX VEGA 22.5 FR 10 MM WAS DISLODGED INTO THE ESOPHAGUS SIDE WHILE THE PATIENT WAS HOSPITALIZED. THE INCIDENT REQUIRED MEDICAL INTERVENTION. (B)(6) 2023: WHILE THE PATIENT WAS EATING AT THE HOSPITAL (UNKNOWN BREAKFAST, LUNCH OR DINNER), THE INDWELLED VOICE PROSTHESIS WAS DISLODGED INTO THE PATIENT'S ESOPHAGUS SIDE. THE PATIENT CALLED A NURSE FOR HELP BY HIM/HERSELF. THE DOCTOR CAME AND CONFIRMED THAT THE VOICE PROSTHESIS WAS STUCK IN THE ESOPHAGUS AND WAS TAKEN OUT BY MEANS OF AN ENDOSCOPE. THE ACTUAL VOICE PROSTHESIS (REF 8285) WAS DISPOSED BY THE HOSPITAL BECAUSE THE HOSPITAL DETERMINED NOT TO REPORT THIS AS AN INCIDENT INTERNALLY. 12/01/2023: A SALES REPRESENTATIVE FROM ATOS MEDICAL VISITED THE HOSPITAL TO HEAR ABOUT THE INCIDENT. THE INCIDENT WAS DISCUSSED WITH DOCTOR AND HEAD NURSE WHICH CONFIRMED THAT THE VOICE PROSTHESIS HAD NOT LEAKED AT ALL AND WAS NOT BROKEN - LOOKED IN A GOOD CONDITION. THEREFORE, IT WAS DETERMINED THAT IT WAS NOT CAUSED BY THE PRODUCT AND THE VOICE PROSTHESIS WAS DISPOSED. THE VOICE PROSTHESIS WAS INDWELLED CLOSER TO, AND JUST BESIDE THE STOMA, BECAUSE THE PATIENT HAD A KUCYO SAIKEN PROCEDURE AND THERE WAS NOT AN OPTION OF A GOOD POSITION FOR MAKING

    Patient problems: Foreign Body In Patient

    MAUDE report 18324824 ↗

Other Prosthesis, Laryngeal (Taub) devices

All Prosthesis, Laryngeal (Taub) devices →

Sources and updates