The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07331791010811), received between Feb 18, 2026 and Feb 18, 2026 — classified as 1 death, 1 injury.
The most frequently reported specific device problem is “Biofilm coating in Device” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Biofilm coating in Device2
Patient problems
- Obstruction/Occlusion2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryFeb 18, 2026Biofilm coating in Device
ACCORDING TO THE AVAILABLE INFORMATION, THIS COMPLAINT CONCERNS AN END USER WITH A VOICE PROSTHESIS (VP) PLACED ON (B)(6) 2025 BY A SPEECH LANGUAGE PATHOLOGIST (SLP) FOLLOWING REMOVAL OF A PREVIOUSLY WORN PROSTHESIS. LEAKAGE THROUGH THE DEVICE WAS REPORTEDLY OBSERVED IMMEDIATELY AFTER PLACEMENT. THE SLP ADJUSTED THE TRACHEAL FLANGE BY PULLING IT FORWARD AND ROTATING THE DEVICE, AFTER WHICH LEAKAGE RESOLVED. ON (B)(6) 2025, WHILE THE PATIENT WAS HOSPITALIZED FOR AN UNRELATED BOWEL CONDITION, LEAKAGE THROUGH THE PROSTHESIS WAS AGAIN OBSERVED DURING SUCTIONING AND THE SLP WAS NOTIFIED. ON (B)(6) 2025, THE PROSTHESIS WAS NO LONGER PRESENT AT THE PUNCTURE SITE AND WAS PRESUMED SWALLOWED. IMAGING, INCLUDING X-RAY AND BRONCHOSCOPY, DID NOT IDENTIFY ANY PORTION OF THE DEVICE IN THE LUNGS. A RED RUBBER CATHETER WAS PLACED AT THE PUNCTURE SITE. ON 01/21/2026, THE MANUFACTURER WAS NOTIFIED THAT THE PATIENT PASSED AWAY ON (B)(6) 2026. AT THIS TIME, NO INFORMATION HAS BEEN RECEIVED INDICATING THAT THE DEATH WAS CAUSED BY OR CONTRIBUTED TO BY THE DEVICE. WITHOUT THE BENEFIT OF EXAMINATION AND TESTING, COLOPLAST IS PRECLUDED FROM COMMENTING ON THE CONDITION OF THE DEVICE OR THE CAUSE OF THE OCCUR
Patient problems: Obstruction/Occlusion
MAUDE report 24380296 ↗DeathFeb 18, 2026Biofilm coating in Device
ACCORDING TO THE AVAILABLE INFORMATION, THIS COMPLAINT CONCERNS AN END USER WITH A VOICE PROSTHESIS (VP) PLACED ON (B)(6) 2025 BY A SPEECH LANGUAGE PATHOLOGIST (SLP) FOLLOWING REMOVAL OF A PREVIOUSLY WORN PROSTHESIS. LEAKAGE THROUGH THE DEVICE WAS REPORTEDLY OBSERVED IMMEDIATELY AFTER PLACEMENT. THE SLP ADJUSTED THE TRACHEAL FLANGE BY PULLING IT FORWARD AND ROTATING THE DEVICE, AFTER WHICH LEAKAGE RESOLVED. ON (B)(6) 2025, WHILE THE PATIENT WAS HOSPITALIZED FOR AN UNRELATED BOWEL CONDITION, LEAKAGE THROUGH THE PROSTHESIS WAS AGAIN OBSERVED DURING SUCTIONING AND THE SLP WAS NOTIFIED. ON (B)(6) 2025, THE PROSTHESIS WAS NO LONGER PRESENT AT THE PUNCTURE SITE AND WAS PRESUMED SWALLOWED. IMAGING, INCLUDING X-RAY AND BRONCHOSCOPY, DID NOT IDENTIFY ANY PORTION OF THE DEVICE IN THE LUNGS. A RED RUBBER CATHETER WAS PLACED AT THE PUNCTURE SITE. ON (B)(6) 2025, THE MANUFACTURER WAS NOTIFIED THAT THE PATIENT PASSED AWAY ON (B)(6) 2026. AT THIS TIME, NO INFORMATION HAS BEEN RECEIVED INDICATING THAT THE DEATH WAS CAUSED BY OR CONTRIBUTED TO BY THE DEVICE. WITHOUT THE BENEFIT OF EXAMINATION AND TESTING, COLOPLAST IS PRECLUDED FROM COMMENTING ON THE CONDITION OF THE DEVICE OR THE CAUSE OF THE OCCU
Patient problems: Obstruction/Occlusion
MAUDE report 24378933 ↗
Other Prosthesis, Laryngeal (Taub) devices
- InHealth Technologies; Blom-Singer; Classic IN 1610-LFFreudenberg Medical, LLC · Prosthesis, Laryngeal (Taub)1 recall0 reports
- InHealth Technologies; Blom-Singer; Classic IN 1610-LEFFreudenberg Medical, LLC · Prosthesis, Laryngeal (Taub)1 recall0 reports
- Provox2 Voice Prosthesis 8mm 7218Atos Medical AB · Prosthesis, Laryngeal (Taub)5 reports
- Provox ActiValve Strong 6 mm 7161Atos Medical AB · Prosthesis, Laryngeal (Taub)3 reports
- Provox Vega 22.5Fr 6mm 8283Atos Medical AB · Prosthesis, Laryngeal (Taub)2 reports
- Provox Vega 17Fr 6mm 8271Atos Medical AB · Prosthesis, Laryngeal (Taub)2 reports
- Provox ActiValve Strong 4.5 mm 7160Atos Medical AB · Prosthesis, Laryngeal (Taub)2 reports
- Provox Vega 22.5Fr 10mm 8285Atos Medical AB · Prosthesis, Laryngeal (Taub)2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page