Provox2 Voice Prosthesis 8mm 7218Recalls & reported problems
Atos Medical AB · Provox2 is an indwelling, low resistance voice prosthesis intended for voice restoration after total laryngectomy. Provox2 is intended for both antero- and retrograde insertion at time of surgery or as replacement device. · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07331791010026), received between Feb 12, 2025 and Feb 12, 2025 — classified as 5 injuries.
The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Improper or Incorrect Procedure or Method5
- Patient-Device Incompatibility5
Patient problems
- Aspiration/Inhalation5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryFeb 12, 2025Improper or Incorrect Procedure or Method; Patient-Device Incompatibility; Improper or Incorrect Procedure or Method; Patient-Device Incompatibility
THIS REPORT REFLECTS ASPIRATION OF PROVOX2 WITH MODIFIED FLANGES EVENT 5 OUT OF 5 WHERE TRACHEAL RETRIEVAL WAS NECESSARY. ADVERSE EVENTS WERE FOUND DURING LITERATURE REVIEW FOR THE PERIODIC UPDATE OF THE CLINICAL EVALUATION REPORT FOR VOICE PROSTHESES. THE FOLLOWING TEXT WAS FOUND IN ARTICLE BY EBERSOLE BM, ZAFEREO M, HUTCHESON KA. A CUSTOM DEVICE FOR MANAGING RECURRENT TRACHEOESOPHAGEAL VOICE PROSTHESIS DISLODGEMENT AND ASPIRATION DUE TO TREATMENT REFRACTORY, SEVERELY ENLARGED PUNCTURE: A VIDEO CASE REPORT. HEAD & NECK. 2024;1-7. DOI:10.1002/ HED.27766). "HE WAS EVENTUALLY REFRACTORY TO THESE TREATMENTS AND USING A PROVOX 2 WITH CUSTOMIZED ANTERIOR AND POSTERIOR FLANGES (UP TO 30 MM DIAMETER SILICONE COLLARS), ACHIEVING VARIABLE ASPIRATION CONTROL. SILICONE OF VARYING THICKNESSES (0.015¿0.030 MM) AND DIAMETERS (23¿30 MM) WERE TRIALED IN THE FABRICATION OF HIS CUSTOM FLANGES TO IMPROVE THEIR STABILITY/SUPPORT, BUT WITH LITTLE EFFECT. IN THE 2 YEARS PRECEDING THE USE OF THE NEW INSET-VP DEVICE, 24 VP DISLODGEMENTS WERE REPORTED WITH 5 REQUIRING ED ADMISSION FOR TRACHEAL RETRIEVAL. ROOT CAUSES OF VP DISLODGEMENT WERE JUDGED TO BE DUE TO DISRUPTION OF FLANGE SEATING DURING SELF-DILATI
Patient problems: Aspiration/Inhalation
MAUDE report 21362900 ↗InjuryFeb 12, 2025Improper or Incorrect Procedure or Method; Patient-Device Incompatibility; Improper or Incorrect Procedure or Method; Patient-Device Incompatibility
THIS REPORT REFLECTS ASPIRATION OF PROVOX2 WITH MODIFIED FLANGES EVENT 4 OUT OF 5 WHERE TRACHEAL RETRIEVAL WAS NECESSARY. ADVERSE EVENTS WERE FOUND DURING LITERATURE REVIEW FOR THE PERIODIC UPDATE OF THE CLINICAL EVALUATION REPORT FOR VOICE PROSTHESES. THE FOLLOWING TEXT WAS FOUND IN ARTICLE BY EBERSOLE BM, ZAFEREO M, HUTCHESON KA. A CUSTOM DEVICE FOR MANAGING RECURRENT TRACHEOESOPHAGEAL VOICE PROSTHESIS DISLODGEMENT AND ASPIRATION DUE TO TREATMENT REFRACTORY, SEVERELY ENLARGED PUNCTURE: A VIDEO CASE REPORT. HEAD & NECK. 2024;1-7. DOI:10.1002/ HED.27766). "HE WAS EVENTUALLY REFRACTORY TO THESE TREATMENTS AND USING A PROVOX 2 WITH CUSTOMIZED ANTERIOR AND POSTERIOR FLANGES (UP TO 30 MM DIAMETER SILICONE COLLARS), ACHIEVING VARIABLE ASPIRATION CONTROL. SILICONE OF VARYING THICKNESSES (0.015¿0.030 MM) AND DIAMETERS (23¿30 MM) WERE TRIALED IN THE FABRICATION OF HIS CUSTOM FLANGES TO IMPROVE THEIR STABILITY/SUPPORT, BUT WITH LITTLE EFFECT. IN THE 2 YEARS PRECEDING THE USE OF THE NEW INSET-VP DEVICE, 24 VP DISLODGEMENTS WERE REPORTED WITH 5 REQUIRING ED ADMISSION FOR TRACHEAL RETRIEVAL. ROOT CAUSES OF VP DISLODGEMENT WERE JUDGED TO BE DUE TO DISRUPTION OF FLANGE SEATING DURING SELF-DILATI
Patient problems: Aspiration/Inhalation
MAUDE report 21362844 ↗InjuryFeb 12, 2025Improper or Incorrect Procedure or Method; Patient-Device Incompatibility; Improper or Incorrect Procedure or Method; Patient-Device Incompatibility
THIS REPORT REFLECTS ASPIRATION OF PROVOX2 WITH MODIFIED FLANGES EVENT 3 OUT OF 5 WHERE TRACHEAL RETRIEVAL WAS NECESSARY. ADVERSE EVENTS WERE FOUND DURING LITERATURE REVIEW FOR THE PERIODIC UPDATE OF THE CLINICAL EVALUATION REPORT FOR VOICE PROSTHESES. THE FOLLOWING TEXT WAS FOUND IN ARTICLE BY EBERSOLE BM, ZAFEREO M, HUTCHESON KA. A CUSTOM DEVICE FOR MANAGING RECURRENT TRACHEOESOPHAGEAL VOICE PROSTHESIS DISLODGEMENT AND ASPIRATION DUE TO TREATMENT REFRACTORY, SEVERELY ENLARGED PUNCTURE: A VIDEO CASE REPORT. HEAD & NECK. 2024;1-7. DOI:10.1002/ HED.27766). "HE WAS EVENTUALLY REFRACTORY TO THESE TREATMENTS AND USING A PROVOX 2 WITH CUSTOMIZED ANTERIOR AND POSTERIOR FLANGES (UP TO 30 MM DIAMETER SILICONE COLLARS), ACHIEVING VARIABLE ASPIRATION CONTROL. SILICONE OF VARYING THICKNESSES (0.015¿0.030 MM) AND DIAMETERS (23¿30 MM) WERE TRIALED IN THE FABRICATION OF HIS CUSTOM FLANGES TO IMPROVE THEIR STABILITY/SUPPORT, BUT WITH LITTLE EFFECT. IN THE 2 YEARS PRECEDING THE USE OF THE NEW INSET-VP DEVICE, 24 VP DISLODGEMENTS WERE REPORTED WITH 5 REQUIRING ED ADMISSION FOR TRACHEAL RETRIEVAL. ROOT CAUSES OF VP DISLODGEMENT WERE JUDGED TO BE DUE TO DISRUPTION OF FLANGE SEATING DURING SELF-DILATI
Patient problems: Aspiration/Inhalation
MAUDE report 21362823 ↗InjuryFeb 12, 2025Improper or Incorrect Procedure or Method; Patient-Device Incompatibility; Improper or Incorrect Procedure or Method; Patient-Device Incompatibility
THIS REPORT REFLECTS ASPIRATION OF PROVOX2 WITH MODIFIED FLANGES EVENT 2 OUT OF 5 WHERE TRACHEAL RETRIEVAL WAS NECESSARY. ADVERSE EVENTS WERE FOUND DURING LITERATURE REVIEW FOR THE PERIODIC UPDATE OF THE CLINICAL EVALUATION REPORT FOR VOICE PROSTHESES. THE FOLLOWING TEXT WAS FOUND IN ARTICLE BY EBERSOLE BM, ZAFEREO M, HUTCHESON KA. A CUSTOM DEVICE FOR MANAGING RECURRENT TRACHEOESOPHAGEAL VOICE PROSTHESIS DISLODGEMENT AND ASPIRATION DUE TO TREATMENT REFRACTORY, SEVERELY ENLARGED PUNCTURE: A VIDEO CASE REPORT. HEAD & NECK. 2024;1-7. DOI:10.1002/ HED.27766). "HE WAS EVENTUALLY REFRACTORY TO THESE TREATMENTS AND USING A PROVOX 2 WITH CUSTOMIZED ANTERIOR AND POSTERIOR FLANGES (UP TO 30 MM DIAMETER SILICONE COLLARS), ACHIEVING VARIABLE ASPIRATION CONTROL. SILICONE OF VARYING THICKNESSES (0.015¿0.030 MM) AND DIAMETERS (23¿30 MM) WERE TRIALED IN THE FABRICATION OF HIS CUSTOM FLANGES TO IMPROVE THEIR STABILITY/SUPPORT, BUT WITH LITTLE EFFECT. IN THE 2 YEARS PRECEDING THE USE OF THE NEW INSET-VP DEVICE, 24 VP DISLODGEMENTS WERE REPORTED WITH 5 REQUIRING ED ADMISSION FOR TRACHEAL RETRIEVAL. ROOT CAUSES OF VP DISLODGEMENT WERE JUDGED TO BE DUE TO DISRUPTION OF FLANGE SEATING DURING SELF-DILATI
Patient problems: Aspiration/Inhalation
MAUDE report 21362807 ↗InjuryFeb 12, 2025Improper or Incorrect Procedure or Method; Patient-Device Incompatibility; Improper or Incorrect Procedure or Method; Patient-Device Incompatibility
THIS REPORT REFLECTS ASPIRATION OF PROVOX2 WITH MODIFIED FLANGES EVENT 1 OUT OF 5 WHERE TRACHEAL RETRIEVAL WAS NECESSARY. ADVERSE EVENTS WERE FOUND DURING LITERATURE REVIEW FOR THE PERIODIC UPDATE OF THE CLINICAL EVALUATION REPORT FOR VOICE PROSTHESES. THE FOLLOWING TEXT WAS FOUND IN ARTICLE BY EBERSOLE BM, ZAFEREO M, HUTCHESON KA. A CUSTOM DEVICE FOR MANAGING RECURRENT TRACHEOESOPHAGEAL VOICE PROSTHESIS DISLODGEMENT AND ASPIRATION DUE TO TREATMENT REFRACTORY, SEVERELY ENLARGED PUNCTURE: A VIDEO CASE REPORT. HEAD & NECK. 2024;1-7. DOI:10.1002/ HED.27766). "HE WAS EVENTUALLY REFRACTORY TO THESE TREATMENTS AND USING A PROVOX 2 WITH CUSTOMIZED ANTERIOR AND POSTERIOR FLANGES (UP TO 30 MM DIAMETER SILICONE COLLARS), ACHIEVING VARIABLE ASPIRATION CONTROL. SILICONE OF VARYING THICKNESSES (0.015¿0.030 MM) AND DIAMETERS (23¿30 MM) WERE TRIALED IN THE FABRICATION OF HIS CUSTOM FLANGES TO IMPROVE THEIR STABILITY/SUPPORT, BUT WITH LITTLE EFFECT. IN THE 2 YEARS PRECEDING THE USE OF THE NEW INSET-VP DEVICE, 24 VP DISLODGEMENTS WERE REPORTED WITH 5 REQUIRING ED ADMISSION FOR TRACHEAL RETRIEVAL. ROOT CAUSES OF VP DISLODGEMENT WERE JUDGED TO BE DUE TO DISRUPTION OF FLANGE SEATING DURING SELF-DILATI
Patient problems: Aspiration/Inhalation
MAUDE report 21362784 ↗
Other Prosthesis, Laryngeal (Taub) devices
- InHealth Technologies; Blom-Singer; Classic IN 1610-LFFreudenberg Medical, LLC · Prosthesis, Laryngeal (Taub)1 recall0 reports
- InHealth Technologies; Blom-Singer; Classic IN 1610-LEFFreudenberg Medical, LLC · Prosthesis, Laryngeal (Taub)1 recall0 reports
- Provox ActiValve Strong 6 mm 7161Atos Medical AB · Prosthesis, Laryngeal (Taub)3 reports
- Provox Vega 22.5Fr 6mm 8283Atos Medical AB · Prosthesis, Laryngeal (Taub)2 reports
- Provox Vega 17Fr 6mm 8271Atos Medical AB · Prosthesis, Laryngeal (Taub)2 reports
- Provox ActiValve Strong 4.5 mm 7160Atos Medical AB · Prosthesis, Laryngeal (Taub)2 reports
- Provox Vega 22.5Fr 10mm 8285Atos Medical AB · Prosthesis, Laryngeal (Taub)2 reports
- Provox Vega 20Fr 6mm 8277Atos Medical AB · Prosthesis, Laryngeal (Taub)1 death2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page