Skip to content

TRUWAVE T001709ARecalls & reported problems

Edwards Lifesciences LLC · PRESSURE MONITORING SET · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07460691947610), received between Dec 29, 2025 and Dec 29, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K171996, K183413); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 4, 2023Open, Classified

    Their is the potential for flushing difficulties during setup of the pressure monitoring kit with pressure transducer.

    Z-0413-2024 · Root cause: Device Design · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • Hemorrhage/Blood Loss/Bleeding1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 29, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    AS REPORTED, DURING USE IN PATIENT, THIS DISPOSABLE PRESSURE TRANSDUCER TUBING BROKE, CAUSING MINIMAL BLEEDING FROM THE ARTERY. THIS CAUSED A DISRUPTION IN THE PROCEDURE; THE DEVICE WAS REPLACED WITH A NEW UNIT AND THE ISSUE WAS SOLVED. AS PER CUSTOMER OPINION, THE FAILURE APPEARED TO BE RELATED TO MATERIAL FATIGUE OR A MANUFACTURING DEFECT. THERE WAS NO ALLEGATION OF PATIENT INJURY. AS PER IMAGE EVALUATION, PRESSURE TUBING WAS DETACHED RATHER BROKEN. PATIENT DEMOGRAPHICS WERE NOT PROVIDED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 23905192 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K171996, K183413. One submission often covers many models, so check the product and lot details in each record.

  • Z-0414-2024 · Oct 4, 2023 · Open, Classified
    TruWave with Vamp System Pressure monitoring set, REF: PXAVMP(3cc)/VAMP 68 in (173cm) PXAVMP3: x3 (3cc)/VAMP PXVK0364:(3cc)/VAMP 96in (241cm); PXVK0853: x3 (3cc)/VAMP 76in (193cm)

Other Transducer, Pressure, Catheter Tip devices

All Transducer, Pressure, Catheter Tip devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page