TruWave Pressure monitoring set, REF: PX212: cc/12 in… recallZ-0413-2024 · Edwards Lifesciences, LLC · initiated Oct 4, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Edwards Lifesciences, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Their is the potential for flushing difficulties during setup of the pressure monitoring kit with pressure transducer.
FDA root cause category: Device Design
What the company is doing
On 10/18/2023, Edwards Lifesciences issued a "Urgent: Medical device Correction Action via FedEx mail. Edwards ask consignees to take the following actions: 1. Follow the IFU for proper setup and flushing of your device. 2. Follow the instructions included in the enclosed customer acknowledgement form to complete the acknowledgement process. 3. Verify your inventory on the attached customer acknowledgement form. 3a.E-mail the completed form to Edwards Customer Service at [email protected], within 15 days from receipt of this notification. 4. Distribute this notice within your organization or to any organization where the potentially impacted product has been transferred. 5. Please post a copy of this notice near and/or with affected devices. 6. Contact Edwards Customer Service at 1-888-352-0904 if you would like to return any of your devices or have further questions. 7. Action Requested for Distributors: In addition to the actions above, please forward this customer communication to any of your customers who have purchased the impacted product.
Recall details
- Recall number
- Z-0413-2024
- Status
- Open, Classified
- Initiated by the company
- Oct 4, 2023
- Posted by the FDA
- Nov 29, 2023
- Recalling firm
- Edwards Lifesciences, LLC
- Quantity
- 118,570 units
- Product code
- DRS
- FDA submissions
- K142749, K183413, K171996
- Distribution
- US: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA WA WI WV WY OUS: Argentina Australia Austria Bahrain Belgium Brazil Canada CANARY ISLANDS Chile Colombia Costa Rica Czech Republic Denmark Ecuador Faroe Islands France Germany Greece Guatemala Hong Kong Indonesia Ireland Israel Italy Japan Malaysia Netherlands New Zealand Northern Ireland Norway Panama Poland Portugal Saudi Arabia Singapore Slovakia Slovenia South Korea Spain Sweden Switzerland United Arab Emirates (UAE) United Kingdom Vietnam
Product description
TruWave Pressure monitoring set, REF: PX212: cc/12 in (30cm), PX260: (3 cc/60in (150cm), PX272: 3 cc/72in (180cm), PX284: 3 cc/84 in (210 cm), PX2X2: x2 (3cc), PX3X3284C: X3 (3cc), PX4X4: x4 (3cc), PX600F (3cc), PXMK1876: x 3 (3 cc), PXMK1940: (3 cc), PXMK2024: (3cc)/77 in (196cm),PXMK2041: (3 cc), PXMK2321: (3cc)/48in (122cm), PXMK2331: x4 (3cc)/12 in (30cm), PXMK9146: x 2 (3cc), PXMK1876: x3(3cc), PXVK1099:(3cc)/62in (157cm), PXMK0102: x2 (3cc)/84in (213cm), PXMK2043:X3(3CC), T001657A: x2 (3cc)/210cm, T001658A: x2 (3cc)/180 cm, T001660A: X3 (3CC)/150CM, T001709A: X2 (3 cc)/210 cm,T001767A: (3cc)/60in (150cm), T321573A: (3cc)/66 in (165cm), T433803A (3cc)/200cm, T441T01C: (3cc)/225 cm, T494C00B: (3 cc)/180 cm,
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
PX212: UDI: 07460691950221/Lot # 65087613; PX260: UDI: 07460691950191/Lot #64817892, 64818036, 64818038, 64964776, 64964780; PX272: UDI: 07460691950207/Lot # 65067588 PX284: UDI: 07460691950214/Lot # 64980120, 64980121, 65008760, 65019888, 65019889, 65035079; PX2X2: UDI: 07460691950061/Lot # 64955114, 65012789, 65027868, 65170585; PX3X3284C:UDI: 07460691950269/Lot # 64906878, 65007202, 65019845; PX4X4: UDI: 07460691950665/Lot # 64807960, 64835551, 64858629; PX600F: UDI: 07460691945555/Lot # 64918553, 64951413, 65016989; PXMK1876: UDI: 07460691953826/Lot # 64817828, 64817963; PXMK1940: UDI: 07460691955981/Lot # 64858039; PXMK2024: UDI: 07460691958180/Lot # 64964690; PXMK2041: UDI: 07460691951464/Lot # 64933443, 64933444; PXMK2321: UDI: 07460691958784/Lot # 64912231; PXMK2331: UDI: 07460691959224/Lot # 64933498; PXMK9146: UDI: 07460691954427/Lot # 64992075; PXMK1876: UDI: 07460691953826/Lot # 64817963; PXVK1099: UDI: 07460691958753 /Lot # 65067568 PXMK0102: UDI: 00690103199932/ Lot # 65008741 PXMK2043: UDI: 07460691956230/ Lot # 64980031 T001657A: UDI: 07460691902046/Lot# 64686945, 64807966, 64918558, 64948620; T001658A :UDI: 07460691902053/Lot #65007203, 65036124; T001660A: UDI: 07460691902077/Lot # 64964732, 65019857; T001709A: UDI: 07460691947610/Lot# 64918560, 64964727, 65079483; T001767A: UDI: 07460691947306/Lot # 64891376, 64906871, 64980035, 64983290, 65035029; T321573A: UDI: 07460691910270/Lot# 64807975; T433803A: UDI: 07460691926714/Lot # 64722635; T441T01C: UDI: 07460691928848/Lot # 64835565; T494C00B:UDI: 07460691950313/Lot #64807969, 64918562;
Devices named in this recall
24 devices in the FDA device registry whose identifier appears in the recall's code information.
- TRUWAVE PX260Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall29 reports
- TRUWAVE PXMK2041Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall4 reports
- TRUWAVE PX284Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall3 reports
- TRUWAVE T001709AEdwards Lifesciences LLC · Transducer, Pressure, Catheter Tip1 recall1 report
- TRUWAVE PX2X2Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall1 report
- TRUWAVE PX272Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall1 report
- TRUWAVE PX212Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall1 report
- TRUWAVE PX3X3284CEdwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall1 report
- TRUWAVE T001657AEdwards Lifesciences LLC · Transducer, Pressure, Catheter Tip1 recall0 reports
- TRUWAVE T001658AEdwards Lifesciences LLC · Transducer, Pressure, Catheter Tip1 recall0 reports
- TRUWAVE T001660AEdwards Lifesciences LLC · Transducer, Pressure, Catheter Tip1 recall0 reports
- TRUWAVE T321573AEdwards Lifesciences LLC · Transducer, Pressure, Catheter Tip1 recall0 reports
- TRUWAVE T433803AEdwards Lifesciences LLC · Transducer, Pressure, Catheter Tip1 recall0 reports
- TRUWAVE T441T01CEdwards Lifesciences LLC · Transducer, Pressure, Catheter Tip1 recall0 reports
- TRUWAVE PX600FEdwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall0 reports
- TRUWAVE T494C00BEdwards Lifesciences LLC · Transducer, Pressure, Catheter Tip1 recall0 reports
- TRUWAVE PX4X4Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall0 reports
- TRUWAVE PXMK1876Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall0 reports
- TRUWAVE PXMK9146Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall0 reports
- TRUWAVE PXMK1940Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall0 reports
- TRUWAVE PXMK2024Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall0 reports
- TRUWAVE, VAMP PXVK1099Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall0 reports
- TRUWAVE PXMK2321Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall0 reports
- TRUWAVE PXMK2331Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall0 reports
Product lines
Other devices under the same FDA submission
50 devices cleared under K142749, K183413, K171996 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- TRUWAVE, VAMP PLUS PXVP2284AT3Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall9 reports
- TRUWAVE, VAMP PXAVMPEdwards Lifesciences LLC · Catheter, continuous flush1 recall7 reports
- TRUWAVE, VAMP PLUS PXVP2284Edwards Lifesciences LLC · Catheter, continuous flush1 recall6 reports
- TRUWAVE, VAMP PLUS PXVP0550Edwards Lifesciences LLC · Catheter, continuous flush1 recall5 reports
- TRUWAVE, VAMP PXAVMP3Edwards Lifesciences LLC · Catheter, continuous flush1 recall2 reports
- TRUWAVE, VAMP PLUS PXVP23X3Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall2 reports
- TRUWAVE, VAMP PLUS PXVP2260Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall2 reports
- TRUWAVE, VAMP PLUS PXVP23X3AT3Edwards Lifesciences LLC · Catheter, continuous flush1 recall2 reports
- TRUWAVE, VAMP PLUS PXVP2272Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall1 report
- TRUWAVE, VAMP PLUS PXVP2X2AT3Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall1 report
- TRUWAVE, VAMP PXVMP2X22Edwards Lifesciences LLC · Catheter, continuous flush1 recall0 reports
- TRUWAVE, VAMP PXVMP3X32Edwards Lifesciences LLC · Catheter, continuous flush1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93279).
- Z-0409-2024 · Open, ClassifiedAcumen IQ Sensor, REFs: AIQS6, AIQS65, AIQS6R, AIQS6R5, AIQS8, AIQS85, AIQS8R5, Rx only, Sterile EO. For use in intravascular pressure monitoring.
- Z-0410-2024 · Open, ClassifiedAcumen IQ Sensor with VAMP System, REFs: AIQS6AZ & AIQS6AZ5, STERILEEO, Rx only. For use in intravascular pressure monitoring.
- Z-0411-2024 · Open, ClassifiedFloTrac Sensor, REF: MHD8 & MHD85,MHD6R5, MHD8R, MHD8R5, MHD8S, STERILEEO, Rx only. For use in intravascular pressure monitoring.
- Z-0412-2024 · Open, ClassifiedFloTrac Sensor with Vamp System, REF: MHD6AZ, MHD6AZ5 , MHD6C502 STIRLEEO, Rx only. For use in intravascular pressure monitoring.
- Z-0414-2024 · Open, ClassifiedTruWave with Vamp System Pressure monitoring set, REF: PXAVMP(3cc)/VAMP 68 in (173cm) PXAVMP3: x3 (3cc)/VAMP PXVK0364:(3cc)/VAMP 96in (241cm); PXVK0853: x3 (3cc)/VAMP 76in (193cm)…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page