Skip to content

FASSIER-DUVAL TELESCOPIC IM SYSTEM MR320Recalls & reported problems

Orthopediatrics Canada ULC · HANDLE ASSEMBLY · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07540194013095), received between Mar 4, 2026 and Jun 17, 2026 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Device Handling Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Handling Problem1
  • Difficult to Open or Close1

Patient problems

  • Bone Fracture(s)1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 17, 2026Difficult to Open or Close

    THE DISTRIBUTOR FROM (B)(6) REPORTED THAT THE MALE RETRIEVER (MR300) COULD NOT BE DISASSEMBLED AFTER SURGERY. THE HANDLE ASSEMBLY, TUBE, AND COIL-PULLER COULD NOT BE REMOVED, THE RELEASE KEY COULD NOT BE INSERTED, AND THE MECHANICAL ASSEMBLY WAS LOOSE. INSERTION OF THE TORQUE HANDLE DID NOT ALLOW ROTATION. THE EVENT RESULTED IN A DELAY OF MORE THAN 30 MINUTES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25533489 ↗
  • InjuryMar 4, 2026Device Handling Problem

    A HOSPITAL REPRESENTATIVE FROM THE UNITED KINGDOM REPORTED AN ISSUE INVOLVING A MALE RETRIEVER HANDLE (MR300). SPECIFICALLY, HE STATED THAT TWO MECHANICAL SCREWS WERE MISSING AND THE RATCHET MECHANISM WAS NOT FUNCTIONING.

    Patient problems: Bone Fracture(s)

    MAUDE report 24506262 ↗

Other EXTRACTOR devices

All EXTRACTOR devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page