Clarius Ultrasound Scanner Clarius Scanner C3 HD3Recalls & reported problems
Clarius Mobile Health Corp · Portable Ultrasound Scanner · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07540205001189), received between Oct 17, 2024 and Nov 25, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Application Program Freezes or Fails to Launch” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Application Program Freezes or Fails to Launch1
- Misconnection1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 25, 2024Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
A SCANNER MALFUNCTION DUE TO A CORRUPTED SD CARD DELAYED DIAGNOSTIC IMAGING FOR A PATIENT WITH LOW OXYGEN LEVELS AND ELEVATED WHITE CELL COUNT. LUNG AUSCULTATION SUGGESTED PNEUMONIA, WHILE A CHEST X-RAY RAISED CONCERNS ABOUT PULMONARY CONGESTION. THE PATIENT'S CONDITION WORSENED, REQUIRING INCREASED OXYGEN SUPPORT AND EVENTUAL TRANSFER TO A LARGER FACILITY. LIMITED DIAGNOSTICS PREVENTED TIMELY INTERVENTION WITH DIURETICS, BUT THE DELAY DID NOT RESULT IN DEATH OR SEVERE HEALTH DETERIORATION. THE MALFUNCTION WAS RESOLVED WITH CUSTOMER SUPPORT ASSISTANCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20774543 ↗MalfunctionOct 17, 2024Misconnection; Misconnection
DURING A DIAGNOSTIC EVALUATION IN AN EMERGENCY ROOM IN CANADA, A PHYSICIAN ENCOUNTERED A CONNECTIVITY ISSUE WITH CLARIUS ULTRASOUND SCANNER C3 HD3 WHILE ATTEMPTING TO ASSESS A PATIENT PRESENTING WITH ABDOMINAL PAIN AND SUSPECTED FLUID ACCUMULATION. THE MALFUNCTION DID NOT RESULT IN DEATH, SERIOUS INJURY, OR ANY SIGNIFICANT ADVERSE DEVICE-RELATED OUTCOMES. HOWEVER, THE MALFUNCTION CAUSED A DELAY IN DIAGNOSTIC THAT COULD HAVE EXACERBATE THE PATIENT'S CONDITION AND LEAD TO A DEATH, SERIOUS INJURY OR SIGNIFICANT DEVICE EXPERIENCE IF IT WERE TO REOCCUR IN AN EMERGENCY ENVIRONMENT. THE EVENT WAS REPORTED TO CLARIUS ON SEPTEMBER 11TH, 2024 AND THE CLARIUS SCANNER C3 HD3 HAS BEEN CLEARED BY THE FDA UNDER 510(K) #K213436. THE MALFUNCTION IDENTIFIED UNDER THIS REPORT OCCURRED DUE TO A CONNECTION ISSUE BETWEEN CLARIUS APP AND CLARIUS SCANNER BECAUSE OF A CLARIUS APP PILOT UPDATE ( CLARIUS APP 11.3.1 TO CLARIUS APP 11.3.2). THE ISSUE WAS SUCCESSFULLY RESOLVED BY UNINSTALLING AND THEN REINSTALLING THE APPLICATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20471171 ↗
Other System, Imaging, Pulsed Doppler, Ultrasonic devices
- Transducer C5-1 989605412042Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic11 recalls0 reports
- Transducer X5-1 989605400802Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic10 recalls0 reports
- EPIQ 7 Diagnostic Ultrasound System 989605386721Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic3 recalls2 deaths146 reports
- Transducer S7-3t 989605406772Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic5 recalls1 death18 reports
- EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE OLYMPUS GF TYPE UCT180Olympus Medical Systems Corp. · System, imaging, pulsed doppler, ultrasonic199 reports
- Transducer S7-3t 989605439232Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic3 recalls4 reports
- Transducer 3D9-3v 989605395622Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic3 recalls1 report
- ACUSON Redwood Diagnostic Ultrasound System 11503314Siemens Medical Solutions USA, Inc. · System, imaging, pulsed doppler, ultrasonic3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page