ICONIX 3910-500-512Recalls & reported problems

Stryker Corporation · 1.4MM ICONIX 1 ANCHOR W/INTELLIBRAID TECHNOLOGY 1 STRAND #2 FORCE FIBER · Class 2 · Implant

All models of ICONIX (18)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Stryker ICONIX soft tissue anchor is a small suture anchor used to fasten tissue to bone. The 30 reports mostly describe the anchor or its inserter tip breaking during placement, sometimes leaving a fragment or wire inside the patient, along with a few reports of expired product being used. Reports mention 15 injuries and no deaths, with 15 malfunctions. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
30
Deaths reported
0
Injuries reported
15
Recalls naming it
0

Summary

The FDA's MAUDE database contains 30 adverse event reports that list this exact device (DI 07613252450337), received between Feb 18, 2019 and Aug 27, 2026 — classified as 15 injuries, 15 malfunctions.

The most frequently reported specific device problem is “Break” (14 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break14
  • Fracture5
  • Adverse Event Without Identified Device or Use Problem3
  • Expiration Date Error3
  • Nonstandard Device3
  • Patient Device Interaction Problem3
  • Contamination /Decontamination Problem1
  • Material Fragmentation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions18
  • Foreign Body In Patient9
  • Insufficient Information3
  • No Known Impact Or Consequence To Patient3
  • Scar Tissue3
  • Synovitis3
  • Hypersensitivity/Allergic reaction2
  • Reaction to Medicinal Component of Device1

Recent adverse event reports

The 25 most recent of 30 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 27, 2026Fracture

    It was reported that the broken tip portion of the ICONIX remained in the patient's glenoid cavity.

    Patient problems: Foreign Body In Patient

    MAUDE report 50488472 ↗
  • MalfunctionSep 24, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE TIP OF THE INSERTER BROKE OFF, LEAVING FRAGMENTS IN THE PATIENT'S BONE.

    Patient problems: Foreign Body In Patient

    MAUDE report 23139111 ↗
  • InjuryApr 29, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE TIP OF THE INSTRUMENT HOLDING THE ANCHOR BROKE OFF DURING IMPLANT INSERTION. THE TIP OF THE ICONIX REMAINED IN THE PATIENT BONE.

    Patient problems: Foreign Body In Patient; Insufficient Information

    MAUDE report 21931158 ↗
  • InjuryJun 8, 2024Fracture; Fracture

    IT WAS REPORTED THAT AFTER INSERTING THE ANCHOR INTO THE BONE, THE TIP PART OF THE INSERTER WAS BROKEN AND WAS LEFT IN THE PATIENT'S BODY WITH THE ANCHOR.

    Patient problems: Foreign Body In Patient

    MAUDE report 19493688 ↗
  • MalfunctionMay 8, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING THE SURGICAL PROCEDURE THE ANCHOR UNIT WAS USED AND FAILED IN PERFORMANCE. THE WIRE REMAINED IN THE PATIENT AND IT WAS NOT POSSIBLE TO REMOVE IT, THE ANCHOR WAS NEGLECTED, AND CAUSED NO HARM TO THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 19274360 ↗
  • InjuryMay 7, 2024Fracture; Fracture

    IT WAS REPORTED THAT THERE WAS TIP BREAKAGE OF THE DEVICE AND THE FRAGMENT REMAINED IN THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 19265131 ↗
  • MalfunctionNov 9, 2023Expiration Date Error; Expiration Date Error

    IT WAS REPORTED THAT AN EXPIRED PRODUCT WAS USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18107019 ↗
  • MalfunctionJul 12, 2023Expiration Date Error; Expiration Date Error

    IT WAS REPORTED THAT AN EXPIRED PRODUCT WAS USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17312739 ↗
  • MalfunctionMay 25, 2023Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT THERE WAS THERE WAS INTENSE SYNOVITIS AND SCARRING AROUND THE ANCHOR PREVIOUSLY IMPLANTED. AN ADDITIONAL PROCEDURE WAS COMPLETED TO REMOVE.

    Patient problems: Scar Tissue; Synovitis

    MAUDE report 17007934 ↗
  • InjuryMay 25, 2023Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT THERE WAS THERE WAS INTENSE SYNOVITIS AND SCARRING AROUND THE ANCHOR PREVIOUSLY IMPLANTED. AN ADDITIONAL PROCEDURE WAS COMPLETED WHERE IMMEDIATE RELIEF WAS FELT UPON REMOVAL.

    Patient problems: Scar Tissue; Synovitis

    MAUDE report 17007933 ↗
  • InjuryMay 25, 2023Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT THERE WAS THERE WAS INTENSE SYNOVITIS AND SCARRING AROUND THE ANCHOR PREVIOUSLY IMPLANTED. AN ADDITIONAL PROCEDURE WAS COMPLETED TO REMOVE.

    Patient problems: Scar Tissue; Synovitis

    MAUDE report 17007060 ↗
  • MalfunctionMar 15, 2023Expiration Date Error

    IT WAS REPORTED THAT AN EXPIRED DEVICE WAS USED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16549750 ↗
  • MalfunctionMar 1, 2023Break

    IT WAS REPORTED THAT THE TIP OF THE INSERTER BROKE DURING PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16463973 ↗
  • MalfunctionDec 1, 2022Break

    IT WAS REPORTED THAT THE INSERTER TIP BROKE OFF AND REMAINED IN THE PATIENT.

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 15898727 ↗
  • MalfunctionJan 19, 2022Break

    IT WAS REPORTED THAT THE TIP OF THE INSERTER BROKE DURING THE PROCEDURE. THE BROKEN PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13292065 ↗
  • MalfunctionNov 9, 2021Nonstandard Device; Adverse Event Without Identified Device or Use Problem; Nonstandard Device; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAD A REACTION TO THE DEVICES. NOTE, THERE ARE NO PLANS TO PERFORM AN ADDITIONAL PROCEDURE.

    Patient problems: Hypersensitivity/Allergic reaction; No Clinical Signs, Symptoms or Conditions

    MAUDE report 12779202 ↗
  • MalfunctionNov 9, 2021Nonstandard Device; Adverse Event Without Identified Device or Use Problem; Nonstandard Device; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAD A REACTION TO THE DEVICES. NOTE, THERE ARE NO PLANS TO PERFORM AN ADDITIONAL PROCEDURE.

    Patient problems: Hypersensitivity/Allergic reaction; No Clinical Signs, Symptoms or Conditions

    MAUDE report 12779201 ↗
  • MalfunctionNov 9, 2021Nonstandard Device; Adverse Event Without Identified Device or Use Problem; Nonstandard Device; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAD A REACTION TO THE DEVICES. NOTE, THERE ARE NO PLANS TO PERFORM AN ADDITIONAL PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Reaction to Medicinal Component of Device

    MAUDE report 12779198 ↗
  • MalfunctionOct 11, 2021Break

    IT WAS REPORTED THAT THE TIP OF THE INSERTER BROKE DURING THE PROCEDURE. THE PIECE WAS RETRIEVED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 12606264 ↗
  • InjuryAug 2, 2021Break; Break

    IT WAS REPORTED THAT METAL REMAINED IN THE JOINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12264169 ↗
  • InjuryAug 2, 2021Break; Break

    IT WAS REPORTED THAT METAL REMAINED IN THE JOINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12264167 ↗
  • InjuryAug 2, 2021Break; Break

    IT WAS REPORTED THAT METAL REMAINED IN THE JOINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12264154 ↗
  • InjuryAug 2, 2021Break; Break

    IT WAS REPORTED THAT METAL REMAINED IN THE JOINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12264098 ↗
  • InjuryAug 2, 2021Break; Break

    IT WAS REPORTED THAT METAL REMAINED IN THE JOINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12264076 ↗
  • InjuryAug 2, 2021Break; Break

    IT WAS REPORTED THAT METAL REMAINED IN THE JOINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12264075 ↗