NA 5450815114Recalls & reported problems

Stryker Corporation · Sonic Control Serrated Aggressive Knife · Class U

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Stryker Sonic Control surgical knife is an ultrasonic instrument for cutting tissue during surgery. The 26 reports that name its identifier all describe malfunctions, mostly the tip or blade breaking or fracturing during setup or surgery, with no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
26
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 26 adverse event reports that list this exact device (DI 07613252518402), received between Jul 2, 2018 and Apr 22, 2026 — classified as 26 malfunctions.

The most frequently reported specific device problem is “Fracture” (19 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture19
  • Break6
  • Packaging Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions20
  • No Consequences Or Impact To Patient5
  • No Patient Involvement1

Recent adverse event reports

The 25 most recent of 26 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 22, 2026Fracture

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1 ¿ MARCH 31 2026. THIS REPORT SUMMARIZES 2 EVENTS FOR THE FAILURE: FRACTURE. - 2 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24954355 ↗
  • MalfunctionJan 23, 2026Fracture

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1 ¿ DECEMBER 31 2025. THIS REPORT SUMMARIZES 2 EVENTS FOR THE FAILURE: FRACTURE. - 2 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24163598 ↗
  • MalfunctionJul 29, 2025Fracture; Fracture

    THIS REPORT SUMMARIZES 3 EVENTS FOR THE FAILURE: FRACTURE. 3 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22647441 ↗
  • MalfunctionJan 22, 2024Fracture; Fracture

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18555901 ↗
  • MalfunctionOct 23, 2023Fracture; Fracture

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17985874 ↗
  • MalfunctionJul 26, 2023Fracture; Fracture

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. 2 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17397361 ↗
  • MalfunctionFeb 22, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING SET-UP FOR A PROCEDURE, WHEN PRIMING, THE TIP WAS BROKEN. IT WAS FURTHER REPORTED THAT THERE WAS NO DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16416229 ↗
  • MalfunctionDec 2, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE TIP OF THE DEVICE BROKE. IT WAS ALSO REPORTED THERE WERE NO DELAYS, NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15904101 ↗
  • MalfunctionJul 12, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING A PROCEDURE, THE TIP OF THE SONOPET KNIFE BROKE. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT OR USER AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14993551 ↗
  • MalfunctionJul 11, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, IT WAS NOTED THAT THE TIP BROKE ON THREE DEVICES WITHIN THE SAME PROCEDURE. IT WAS ALSO REPORTED THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THIS RECORD ADDRESSES ONE OF THE REPORTED INSTANCES OF THE TIP BREAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14982010 ↗
  • MalfunctionJul 11, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, IT WAS NOTED THAT THE TIP BROKE ON THREE DEVICES WITHIN THE SAME PROCEDURE. IT WAS ALSO REPORTED THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THIS RECORD ADDRESSES ONE OF THE REPORTED INSTANCES OF THE TIP BREAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14981948 ↗
  • MalfunctionJul 11, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, IT WAS NOTED THAT THE TIP BROKE ON THREE DEVICES WITHIN THE SAME PROCEDURE. IT WAS ALSO REPORTED THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THIS RECORD ADDRESSES ONE OF THE REPORTED INSTANCES OF THE TIP BREAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14981902 ↗
  • MalfunctionJul 6, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, IT WAS NOTED THAT THE TIP OF THE DEVICE BROKE. IT WAS ALSO REPORTED THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14933923 ↗
  • MalfunctionAug 24, 2021Fracture; Fracture

    IT WAS REPORTED THAT THE TIP BROKE DURING A LOWER JAW SURGICAL PROCEDURE. IT WAS ALSO REPORTED THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12359670 ↗
  • MalfunctionApr 23, 2021Fracture; Fracture

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE TIP BROKE OFF AND WAS REMOVED. IT WAS FURTHER REPORTED THAT THERE WAS A TWO MINUTE DELAY. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11716021 ↗
  • MalfunctionApr 12, 2021Fracture; Fracture

    IT WAS REPORTED THAT AFTER A SSRO (SAGITAL SPLIT RAMUS OSTEOTOMY) SURGICAL PROCEDURE, IT WAS OBSERVED THE TIP OF THE DEVICE BROKE. IT WAS ALSO REPORTED THAT THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11648023 ↗
  • MalfunctionMar 19, 2021Break; Break

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE TIP OF THE DEVICE BROKE. IT WAS ALSO REPORTED THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11527316 ↗
  • MalfunctionJan 13, 2021Break; Break

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE TIP BROKE. IT WAS ALSO REPORTED THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11161946 ↗
  • MalfunctionOct 20, 2020Fracture; Fracture

    IT WAS REPORTED THAT PRIOR TO A SURGICAL PROCEDURE, THE TIP OF THE DEVICE BROKE DURING DEVICE ASSEMBLY. IT WAS ALSO REPORTED THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10707892 ↗
  • MalfunctionSep 17, 2020Packaging Problem; Packaging Problem

    IT WAS REPORTED THAT THERE WAS A POTENTIAL STERILITY BREACH ON THE PACKAGING OF THE DEVICE, IT WAS NOTED THAT THERE WAS A HAIR IN THE PACK DURING VISUAL INSPECTION. IT WAS ALSO REPORTED THAT THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE, AND THERE WAS NO DELAY OR ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10550596 ↗
  • MalfunctionDec 9, 2019Fracture; Fracture

    IT WAS REPORTED VIA MEDWATCH REPORT # 0501120000-2019-8008 THAT THE TIP BROKE INTO 2 PIECES DURING A LEFORT 1, BILATERAL SAGITTAL SPLIT OSTEOTOMY PROCEDURE. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A REPLACEMENT DEVICE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9437993 ↗
  • MalfunctionNov 13, 2019Fracture; Fracture

    IT WAS REPORTED THAT DURING AN ORAL PLATE FIXATION SURGERY THAT THE TIP BROKE AND WAS REMOVED. IT WAS ALSO REPORTED THAT THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9318124 ↗
  • MalfunctionJul 23, 2019Break; Break

    THIS REPORT SUMMARIZES <NOE> 1 </NOE> MALFUNCTION EVENT IN WHICH THE DEVICE HAD A BROKEN CUTTING ACCESSORY STILL ATTACHED. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Patient Involvement

    MAUDE report 8816022 ↗
  • MalfunctionApr 24, 2019Break; Break

    THIS REPORT SUMMARIZES <NOE> 3 </NOE> MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A BROKEN CUTTING ACCESSORY STILL ATTACHED. 3 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8546058 ↗
  • MalfunctionJan 29, 2019Break; Break

    THIS REPORT SUMMARIZES <NOE> 1 </NOE> MALFUNCTION EVENT IN WHICH THE DEVICE HAD A BROKEN CUTTING ACCESSORY STILL ATTACHED. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8285187 ↗

Other Instrument, ultrasonic surgical devices

All Instrument, ultrasonic surgical devices →