Phoenix CAT02504Recalls & reported problems

Stryker Corporation · PHOENIX DRILL BIT · Class 1

Quick summary

This Stryker Phoenix drill bit is used to drill into bone during surgery. Most of the 30 reports describe the bit tip breaking off during a procedure, sometimes leaving fragments in the patient that had to be retrieved, in one case with cartilage damage. There are 7 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
30
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 30 adverse event reports that list this exact device (DI 07613252633242), received between Sep 19, 2018 and Aug 28, 2024 — classified as 7 injuries, 23 malfunctions.

The most frequently reported specific device problem is “Break” (26 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break26
  • Material Fragmentation4
  • Device Fell1
  • Signal Artifact/Noise1

Patient problems

  • No Clinical Signs, Symptoms or Conditions18
  • No Known Impact Or Consequence To Patient6
  • Device Embedded In Tissue or Plaque5
  • Insufficient Information5
  • Foreign Body In Patient4
  • Cartilage damage1
  • Vessel Or Plaque, Device Embedded In1

Recent adverse event reports

The 25 most recent of 30 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 28, 2024Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE DRILL BIT BROKE WHEN ENTERED INTO THE DRILL GUIDE PRIOR TO DRILLING INTO BONE. SURGEON BEGAN TO DRILL AND REALIZED THE TIP OF THE BIT WAS BROKEN. HAD TO BE REMOVED FROM THE PATIENT'S JOINT SPACE. CARTILAGE WAS DAMAGED WHILE TRYING TO RETRIEVE THE BROKEN DRILL BIT TIP.

    Patient problems: Cartilage damage

    MAUDE report 20091729 ↗
  • MalfunctionMar 22, 2024Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE TIP BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18965196 ↗
  • MalfunctionDec 18, 2023Break; Material Fragmentation; Break; Material Fragmentation

    IT WAS REPORTED THAT TIP OF THE DEVICE BROKE DURING THE PROCEDURE. THE PIECE WAS RETRIEVED WITH USE OF X-RAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18351471 ↗
  • MalfunctionAug 2, 2023Break; Break

    IT WAS REPORTED THAT THE TIP OF THE DEVICE BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17446341 ↗
  • MalfunctionJul 28, 2023Break; Break

    IT WAS REPORTED THAT THE TIP OF THE DEVICE BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17416183 ↗
  • MalfunctionJun 21, 2023Break; Break

    IT WAS REPORTED THAT THE PHOENIX DRILL TIP IS IN THE BONE. THE PIECE OF THE TIP POTENTIALLY REMAINED IN THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17175422 ↗
  • MalfunctionFeb 15, 2023Break

    IT WAS REPORTED THAT THE TIP OF THE DEVICE BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16382286 ↗
  • MalfunctionFeb 15, 2023Break

    IT WAS REPORTED THAT THE TIP OF THE DEVICE BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16382282 ↗
  • MalfunctionFeb 15, 2023Break

    IT WAS REPORTED THAT THE TIP OF THE DEVICE BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16382280 ↗
  • MalfunctionFeb 15, 2023Break

    IT WAS REPORTED THAT THE TIP OF THE DEVICE BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16382279 ↗
  • MalfunctionFeb 15, 2023Break

    IT WAS REPORTED THAT THE TIP OF THE DEVICE BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16382270 ↗
  • MalfunctionFeb 15, 2023Break

    IT WAS REPORTED THAT THE TIP OF THE DEVICE BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16382170 ↗
  • MalfunctionJan 20, 2023Break; Break

    IT WAS REPORTED THAT THE TIP OF THE DEVICE BROKE DURING THE PROCEDURE WHICH REQUIRED AN ADDITIONAL SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16218715 ↗
  • MalfunctionDec 6, 2022Break

    IT WAS REPORTED THAT THE DISTAL END OF THE DEVICE BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15929007 ↗
  • MalfunctionDec 6, 2022Break

    IT WAS REPORTED THAT THE DISTAL END OF THE DEVICE BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15928941 ↗
  • MalfunctionOct 13, 2022Break

    IT WAS REPORTED THAT THE DEVICE BROKE DURING THE PROCEDURE. THE PIECE WAS RETRIEVED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 15595182 ↗
  • MalfunctionAug 26, 2021Break

    IT WAS REPORTED THAT THE TIP BROKE DURING THE PROCEDURE. THE PIECE WAS RETRIEVED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 12371981 ↗
  • MalfunctionApr 6, 2021Break

    IT WAS REPORTED THAT THE TIP OF THE DEVICE BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 11625005 ↗
  • InjurySep 14, 2020Break

    IT WAS REPORTED THAT THE TIP BROKE OFF DURING THE PROCEDURE AND A PIECE REMAINED EMBEDDED IN THE BONE.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Vessel Or Plaque, Device Embedded In

    MAUDE report 10530851 ↗
  • InjuryApr 8, 2020Break

    IT WAS REPORTED THAT THE TIP OF THE DEVICE BROKE AND REMAINS IN THE JOINT.

    Patient problems: Device Embedded In Tissue or Plaque; Foreign Body In Patient

    MAUDE report 9939162 ↗
  • InjuryMar 4, 2020Break; Break

    IT WAS REPORTED THAT THE DEVICE BROKE DURING THE PROCEDURE AND POTENTIALLY REMAINS IN THE PATIENT.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 9787443 ↗
  • InjuryJan 9, 2020Break

    IT WAS REPORTED THAT THE TIP BROKE AND A PIECE. REMAINS IN THE PATIENT.

    Patient problems: Device Embedded In Tissue or Plaque; Foreign Body In Patient

    MAUDE report 9570056 ↗
  • InjuryJan 9, 2020Break

    IT WAS REPORTED THAT THE TIP BROKE AND POTENTIALLY REMAINS IN THE PATIENT.

    Patient problems: Device Embedded In Tissue or Plaque; Foreign Body In Patient

    MAUDE report 9570053 ↗
  • MalfunctionDec 17, 2019Break

    IT WAS REPORTED THE DEVICE BROKE DURING THE PROCEDURE. THE PIECE WAS RETRIEVED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9484448 ↗
  • MalfunctionNov 18, 2019Break

    IT WAS REPORTED THAT THE DEVICE BROKE DURING THE PROCEDURE. THE PIECE WAS RETRIEVED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9338182 ↗

Other BIT, DRILL devices