Skip to content

CinchLock Flex CAT02643Recalls & reported problems

Stryker Corporation · FG, CINCHLOCK FLEX KNOTLESS ANCHOR WITH INSERTER · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Stryker CinchLock Flex knotless anchor is used with an inserter to fasten soft tissue to bone. Most of the 58 reports describe the anchor fracturing or breaking during placement, sometimes leaving wire or anchor fragments in the bone that could not be removed; 17 involved injuries and no deaths were described. One open recall from January 2025 covers complaints of the pullwire breaking so a piece stays inside the anchor.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
58
Deaths reported
0
Injuries reported
17
Recalls naming it
1

Summary

The FDA's MAUDE database contains 58 adverse event reports that list this exact device (DI 07613327123661), received between Aug 28, 2018 and Mar 10, 2026 — classified as 17 injuries, 41 malfunctions.

The most frequently reported specific device problem is “Fracture” (26 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 3 reportsSep 2024: 0 reportsOct 2024: 7 reportsNov 2024: 8 reportsDec 2024: 3 reportsJan 2025: 0 reportsFeb 2025: 3 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 29, 2025Open, Classified

    Increase in complaints concerning "pullwire" breakage that may result in the broken pullwire remaining in the anchor implant body after deployment.

    Z-1328-2025 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Fracture26
  • Break23
  • Detachment of Device or Device Component4
  • Difficult to Remove4
  • Material Fragmentation4
  • Activation, Positioning or Separation Problem3
  • Entrapment of Device3
  • Physical Resistance/Sticking2

Patient problems

  • No Clinical Signs, Symptoms or Conditions42
  • Foreign Body In Patient10
  • Insufficient Information10
  • No Known Impact Or Consequence To Patient8
  • Device Embedded In Tissue or Plaque2
  • No Information1

Recent adverse event reports

The 25 most recent of 58 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 10, 2026Material Fragmentation

    IT WAS REPORTED THAT THE ANCHOR DID NOT RELEASE PROPERLY AND LEAD TO THE SUTURE BEING CUT AND HALF THE ANCHOR REMAINED ON THE DEPLOYMENT HANDLE. THIS LEAD TO NO SUTURE, HALF THE ANCHOR BEING LEFT IN THE BONE.

    Patient problems: Foreign Body In Patient

    MAUDE report 24560115 ↗
  • MalfunctionAug 26, 2025Fracture; Fracture

    IT WAS REPORTED THAT A PULL WIRE BREAKAGE OCCURRED IN A CASE AND IT WAS REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22898100 ↗
  • MalfunctionApr 15, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE CINCHLOCK FLEX IMPLANT FAILED TO RELEASE THE NITINOL WIRE FROM THE IMPLANT UPON DEPLOYMENT. THERE WAS A SMALL AMOUNT OF THE WIRE LEFT IN THE IMPLANT WHICH MAY OR MAY NOT BECOME AN ISSUE GOING FORWARD. FOR THIS DEVICE, THE PULL WIRE WAS NOT LEFT IN PATIENT.

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21844597 ↗
  • InjuryFeb 27, 2025Fracture; Fracture

    IT WAS REPORTED THAT WHEN THE ANCHOR WAS RELEASED, THE WIRE BROKE AND REMAINED IN THE ANCHOR. THE ANCHOR OR WIRE COULD BE REMOVED FROM THE PATIENT, ALTHOUGH IT BROKE INTO SEVERAL PIECES. THE SURGEON THINKS THAT THE WIRE BROKE WHEN HE RETURNED THE RELEASE LEVER TO ITS ORIGINAL POSITION. A RESIDUAL PIECE WAS LEFT IN THE BONE. IT WAS IMPOSSIBLE TO REMOVE FOREIGN BODY.

    Patient problems: Insufficient Information

    MAUDE report 21485090 ↗
  • MalfunctionFeb 27, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE PULL WIRE REMAINED IN THE ANCHOR. THE PULL WIRE WAS ABLE TO BE REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21484619 ↗
  • MalfunctionFeb 21, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE PULL WIRE REMAINED IN THE ANCHOR. THE PULL WIRE WAS ABLE TO BE REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21435234 ↗
  • MalfunctionDec 17, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE PULL WIRE REMAINED IN THE ANCHOR. THE PULL WIRE WAS ABLE TO BE CUT OUT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20951695 ↗
  • MalfunctionDec 12, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE PULL WIRE REMAINED IN THE ANCHOR. THE PULL WIRE WAS ABLE TO BE CUT OUT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20919572 ↗
  • MalfunctionDec 11, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE PULL WIRE REMAINED IN THE ANCHOR. THE PULL WIRE WAS ABLE TO BE REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20908412 ↗
  • MalfunctionNov 27, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE CINCHLOCK FLEX IMPLANT FAILED TO RELEASE THE NITINOL WIRE FROM THE IMPLANT UPON DEPLOYMENT. THERE WAS A SMALL AMOUNT OF THE WIRE LEFT IN THE IMPLANT WHICH MAY OR MAY NOT BECOME AN ISSUE GOING FORWARD.

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20799885 ↗
  • MalfunctionNov 27, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE CINCHLOCK FLEX IMPLANT FAILED TO RELEASE THE NITINOL WIRE FROM THE IMPLANT UPON DEPLOYMENT. THERE WAS A SMALL AMOUNT OF THE WIRE LEFT IN THE IMPLANT WHICH MAY OR MAY NOT BECOME AN ISSUE GOING FORWARD.

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20799594 ↗
  • InjuryNov 27, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE CINCHLOCK FLEX IMPLANT FAILED TO RELEASE THE NITINOL WIRE FROM THE IMPLANT UPON DEPLOYMENT. THERE WAS A SMALL AMOUNT OF THE WIRE LEFT IN THE IMPLANT WHICH MAY OR MAY NOT BECOME AN ISSUE GOING FORWARD.

    Patient problems: Foreign Body In Patient

    MAUDE report 20799352 ↗
  • MalfunctionNov 27, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE CINCHLOCK FLEX IMPLANT FAILED TO RELEASE THE NITINOL WIRE FROM THE IMPLANT UPON DEPLOYMENT. THERE WAS A SMALL AMOUNT OF THE WIRE LEFT IN THE IMPLANT WHICH MAY OR MAY NOT BECOME AN ISSUE GOING FORWARD.

    Patient problems: Foreign Body In Patient

    MAUDE report 20799330 ↗
  • MalfunctionNov 7, 2024Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT THE PULL WIRE REMAINED IN THE ANCHOR. THE PULL WIRE WAS ABLE TO BE REMOVED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20630615 ↗
  • MalfunctionNov 7, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE PULL WIRE REMAINED IN THE ANCHOR. THE PULL WIRE WAS ABLE TO BE REMOVED.

    Patient problems: Insufficient Information

    MAUDE report 20626467 ↗
  • MalfunctionNov 6, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE PULL WIRE REMAINED IN THE ANCHOR. THE PULL WIRE WAS ABLE TO BE REMOVED.

    Patient problems: Insufficient Information

    MAUDE report 20624269 ↗
  • MalfunctionNov 5, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE PULL WIRE REMAINED IN THE ANCHOR. THE PULL WIRE WAS ABLE TO BE REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20615952 ↗
  • MalfunctionOct 30, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE PULL WIRE REMAINED IN THE ANCHOR. THE PULL WIRE WAS ABLE TO BE REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20570717 ↗
  • MalfunctionOct 30, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE PULL WIRE REMAINED IN THE ANCHOR. THE PULL WIRE WAS ABLE TO BE REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20570456 ↗
  • MalfunctionOct 30, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE PULL WIRE REMAINED IN THE ANCHOR. THE PULL WIRE WAS ABLE TO BE REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20570450 ↗
  • MalfunctionOct 30, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE PULL WIRE REMAINED IN THE ANCHOR. THE PULL WIRE WAS ABLE TO BE REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20570385 ↗
  • MalfunctionOct 30, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE PULL WIRE REMAINED IN THE ANCHOR. THE PULL WIRE WAS ABLE TO BE REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20570310 ↗
  • MalfunctionOct 24, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE PULL WIRE REMAINED IN THE ANCHOR. THE PULL WIRE WAS ABLE TO BE REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20528056 ↗
  • MalfunctionOct 24, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE PULL WIRE REMAINED IN THE ANCHOR. THE PULL WIRE WAS ABLE TO BE REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20526838 ↗
  • MalfunctionAug 28, 2024Fracture; Material Fragmentation; Fracture; Material Fragmentation

    IT WAS REPORTED THAT THE PULL WIRE REMAINED IN THE ANCHOR AND THE EXPANSION CYLINDER BROKE. THE PULL WIRE WAS ABLE TO BE REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20091805 ↗

Other FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE devices

All FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page