Skip to content

INSTRUMENT 6541-4-518Recalls & reported problems

Howmedica Osteonics Corp. · PEG DRILL TRIATHLON INSTRUMENTS · Class 1

All models of INSTRUMENT (131)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 07613327241440), received between Jan 5, 2017 and Nov 5, 2024 — classified as 1 injury, 9 malfunctions.

The most frequently reported specific device problem is “Fracture” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture8
  • Break3
  • Insufficient Device Problem Information1
  • Material Fragmentation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • No Known Impact Or Consequence To Patient3
  • Foreign Body In Patient1
  • Insufficient Information1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 5, 2024Fracture; Fracture

    DURING THE SURGERY, THE SURGEON USED THE DRILL, AND WHILE PERFORATING THE MEASUREMENT GUIDE, IT BROKE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20609623 ↗
  • MalfunctionOct 1, 2024Fracture; Fracture

    BREAKAGE WHEN USING DRILL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20342921 ↗
  • MalfunctionJul 30, 2024Fracture; Fracture

    ''DRILL BROKE OFF WHILE PRE-DRILLING A PIN HOLE. THE PROCEDURE WAS COMPLETED USING REPLACEMENT. ALL BROKEN PARTS HAVE BEEN COMPLETELY REMOVED, FOR DATA PROTECTION REASONS, NO INFORMATION OF THE PATIENT IS POSSIBLE.''.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19861440 ↗
  • MalfunctionJul 1, 2024Fracture; Fracture

    1/8¿ DRILL SNAPPED INTRAOP AS IT WAS BEING REMOVED FROM PATIENT. SNAPPED PORTION REMOVED AND OPERATION CONTINUED. DRILLING WAS COMPLETE SO NO NEED TO ORGANISE REPLACEMENT. NOTICED BY SURGEON INTRAOP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19643242 ↗
  • MalfunctionJun 29, 2024Fracture; Fracture

    ONE EIGHTH PEG DRILL BROKE DURING DRILLING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19640543 ↗
  • InjuryMar 11, 2024Fracture; Fracture

    DURING DRILLING OF TRIATHLON AP SIZER, THE DRILL BROKE OFF AND STRAYED INTO THE BONE, WHICH WAS RADIOGRAPHED AND REMOVED. IN ADDITION, BONE GRAFTING WAS PERFORMED BECAUSE OF REMOVAL BONE TO LOCATE THE BROKEN DRILL.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 18873006 ↗
  • MalfunctionDec 26, 2018Fracture; Fracture

    TRIATHLON1 1/8TH DRILL BIT SNAPPED WHILE PRE DRILLING HOLE FOR TIBIAL HEADED PIN. SEEN BY SURGEON AND REP DURING SURGERY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8198075 ↗
  • MalfunctionDec 17, 2018Break; Material Fragmentation

    SURGEON WAS USING A DRILL DURING THIS PROCEDURE AND THE DRILL BIT BROKE WHILE IN USE. ALL PIECES OF DRILL BIT WERE FOUND AND REMOVED FROM PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 8168756 ↗
  • MalfunctionNov 12, 2018Break; Fracture; Break; Fracture

    THE CUSTOMER REPORTED THAT 1/8" PEG DRILL BELONGING TO PTG TRIATHLON IS BROKEN.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8061871 ↗
  • MalfunctionJan 5, 2017Break; Insufficient Device Problem Information; Insufficient Device Problem Information

    WHILE PRE-DRILLING TIBIAL BASEPLATE, 1/8 DRILL WITH THE STOP BROKE. LOOSE PIECE WAS FOUND. NO ADVERSE CONSEQUENCES.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 6226790 ↗

Other BIT, DRILL devices

All BIT, DRILL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page