NA 5450840000Recalls & reported problems
Stryker Medtech K.K. · Straight Universal Handpiece · Class U
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 07613327291933), received between Aug 21, 2018 and Jul 26, 2023 — classified as 12 malfunctions.
The most frequently reported specific device problem is “Overheating of Device” (9 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Overheating of Device9
- Inability to Irrigate4
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- No Patient Involvement3
- No Consequences Or Impact To Patient2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 26, 2023Inability to Irrigate; Inability to Irrigate
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE EXPERIENCED AN IRRIGATION ISSUE THAT COULD LEAD TO OVERHEATING. - 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17401357 ↗MalfunctionJan 29, 2023Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE REPORTEDLY OVERHEATED. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16262075 ↗MalfunctionJul 26, 2022Inability to Irrigate; Inability to Irrigate
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE EXPERIENCED AN IRRIGATION ISSUE THAT COULD LEAD TO OVERHEATING. - 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15108023 ↗MalfunctionJul 21, 2021Inability to Irrigate; Overheating of Device; Inability to Irrigate; Overheating of Device
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE REPORTEDLY EXPERIENCED AN IRRIGATION ISSUE THAT COULD LEAD TO OVERHEATING AND OVERHEATED. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12204185 ↗MalfunctionApr 26, 2021Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE REPORTEDLY OVERHEATED. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11729115 ↗MalfunctionJan 22, 2021Inability to Irrigate; Inability to Irrigate
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE EXPERIENCED AN IRRIGATION ISSUE THAT COULD LEAD TO OVERHEATING. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11218295 ↗MalfunctionJul 27, 2020Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE REPORTEDLY OVERHEATED. - 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement
MAUDE report 10329763 ↗MalfunctionOct 28, 2019Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES NOE 1 / NOE MALFUNCTION EVENT IN WHICH THE DEVICE REPORTEDLY OVERHEATED. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Patient Involvement
MAUDE report 9245239 ↗MalfunctionApr 24, 2019Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT. 1 EVENT HAD NO KNOWN PATIENT IMPACT OR CONSEQUENCES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8546070 ↗MalfunctionJan 29, 2019Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES <NOE> 1 </NOE> MALFUNCTION EVENT IN WHICH THE DEVICE REPORTEDLY OVERHEATED. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8285189 ↗MalfunctionSep 24, 2018Overheating of Device; Overheating of Device
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE AT THE USER FACILITY THE HANDPIECE WAS HEATING UP. THE PROCEDURE WAS COMPLETED SUCCESSFULLY, AND THERE WERE NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7903147 ↗MalfunctionAug 21, 2018Overheating of Device; Overheating of Device
IT WAS REPORTED THAT DURING TESTING CONDUCTED BY A MANUFACTURER SALES REPRESENTATIVE AT THE USER FACILITY, THERE WAS NO IRRIGATION. A LACK OF IRRIGATION IN THE DEVICE IS KNOWN TO CAUSE OVERHEATING. THE USER FACILITY REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Patient Involvement
MAUDE report 7800862 ↗
Other Instrument, ultrasonic surgical devices
- Harmonic ACE HARH36Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical4 deaths7,609 reports
- Harmonic HARHD36Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical1 recall2 deaths5,037 reports
- Harmonic ACE HAR36Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical2,726 reports
- Harmonic HAR1136Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical2 deaths1,538 reports
- THUNDERBEAT 5 mm,35 cm, Front-actuated Grip Type S TB-0535FCSOlympus Medical Systems Corp. · Instrument, ultrasonic surgical3 recalls1,373 reports
- THUNDERBEAT 5 mm,35 cm,Front-actuated Grip Type S TB-0535FCSOlympus Medical Systems Corp. · Instrument, ultrasonic surgical3 recalls2 deaths1,128 reports
- THUNDERBEAT 5 mm,35 cm,Front-actuated Grip Type S TB-0535FCSOlympus Medical Systems Corp. · Instrument, ultrasonic surgical2 recalls1 death970 reports
- Harmonic ACE HARH23Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical935 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.