Neuroform Atlas M003UZAS40210Recalls & reported problems
Stryker Corporation · WITHOUT TIP Stent System · Class 3 · Implant
All models of Neuroform Atlas (18)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 07613327382174), received between Feb 18, 2020 and Jul 16, 2026 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Premature Activation” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Premature Activation5
- Difficult or Delayed Activation3
- Entrapment of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 16, 2026Entrapment of Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25841258 ↗MalfunctionApr 20, 2026Premature Activation
IT WAS REPORTED THAT THE SUBJECT STENT WOULD NOT GO THROUGH THE MICROCATHETER. THE SUBJECT STENT WAS RETURNED FOR ANALYSIS AND WAS EXAMINED. DURING DEVICE INVESTIGATION AND ANALYSIS, IT WAS DISCOVERED THAT THE SUBJECT STENT HAD DEPLOYED PREMATURELY DURING USE. THERE IS NO ADDITIONAL INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24929832 ↗MalfunctionJan 28, 2026Difficult or Delayed Activation
THE SUBJECT STENT WAS RETURNED FOR ANALYSIS AND THE DEVICE INVESTIGATION REVEALED STENT PARTIAL DEPLOYMENT .THE SUBJECT DEVICE WAS REPLACED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24200460 ↗MalfunctionMar 11, 2024Premature Activation; Premature Activation
IT WAS REPORTED THAT DURING ANEURYSM STENT COIL PROCEDURE THE OPERATOR WAS ADVANCING THE SUBJECT STENT THROUGH THE MICROCATHETER. FRICTION WAS NOTED WHEN ADVANCING THE SUBJECT STENT TO THE TARGET LESION AND FLUOROSCOPY SHOWED THAT NO SUBJECT STENT WAS PRESENT ON THE DELIVERY WIRE. IT WAS CONCLUDED THAT THE SUBJECT STENT DEPLOYED PREMATURELY INSIDE THE MICROCATHETER. THE SUBJECT DEVICE WAS REPLACED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18881908 ↗MalfunctionDec 5, 2023Difficult or Delayed Activation; Difficult or Delayed Activation
ANALYSIS OF THE RETURNED DEVICE FOUND THAT THE SUBJECT STENT WAS PARTIALLY DEPLOYED. THERE WERE NO CLINICAL CONSEQUENCES TO THE PATIENT REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18267082 ↗MalfunctionMay 26, 2022Premature Activation; Premature Activation
IT WAS REPORTED THAT DURING THE PROCEDURE WHEN THE STENT (SUBJECT DEVICE) WAS DELIVERED THROUGH A MICROCATHETER, IT PREMATURELY DETACHED INSIDE THE CATHETER. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14520047 ↗MalfunctionMay 26, 2022Premature Activation; Premature Activation
IT WAS REPORTED THAT DURING THE PROCEDURE WHEN THE STENT (SUBJECT DEVICE) WAS DELIVERED THROUGH A MICROCATHETER, IT PREMATURELY DETACHED INSIDE THE CATHETER. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14519974 ↗MalfunctionFeb 25, 2021Difficult or Delayed Activation; Difficult or Delayed Activation
IT WAS REPORTED THAT DURING THE PROCEDURE THE SUBJECT STENT WAS PARTIALLY DEPLOYED. THE PHYSICIAN ATTEMPTED TO RESHEATH AND REPOSITION THE SUBJECT DEVICE BUT WAS UNABLE TO RESHEATH. WHILE THE SUBJECT STENT WAS BEING WITHDRAWN, IT GOT DEPLOYED PREMATURELY INSIDE THE RHV OF THE MICROCATHETER. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11382336 ↗MalfunctionFeb 18, 2020Premature Activation; Premature Activation
IT WAS REPORTED THAT THE PHYSICIAN DEPLOYED THE STENT (SUBJECT DEVICE) IN THE MICROCATHETER. THE SALES REPRESENTATIVE HAD THE PHYSICIAN PULL THE STENT OUT AND DISCARDED IT. THERE WERE NO PATIENT CLINICAL CONSEQUENCES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9723131 ↗
Other Intracranial coil-assist stent devices
- Neuroform Atlas M003UZAS30240Stryker Corporation · Intracranial coil-assist stent13 reports
- Neuroform Atlas M003UZAS30210Stryker Corporation · Intracranial coil-assist stent12 reports
- Neuroform Atlas M003UZAS45210Stryker Corporation · Intracranial coil-assist stent11 reports
- Neuroform Atlas M003UZAS45300Stryker Corporation · Intracranial coil-assist stent10 reports
- Stent LEV3018-PMAMicrovention, Inc. · Intracranial coil-assist stent7 reports
- Neuroform Atlas M003UZAS30150Stryker Corporation · Intracranial coil-assist stent6 reports
- Neuroform Atlas M003UZAS40240Stryker Corporation · Intracranial coil-assist stent1 death6 reports
- Stent LEV2517-PMAMicrovention, Inc. · Intracranial coil-assist stent5 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page