Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 07613327422382), received between Feb 10, 2021 and Oct 29, 2024 — classified as 4 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Overheating of Device” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Overheating of Device7
  • Adverse Event Without Identified Device or Use Problem3

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Burn(s)2
  • Partial thickness (Second Degree) Burn2
  • Rupture2
  • Hemorrhage/Blood Loss/Bleeding1
  • Superficial (First Degree) Burn1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 29, 2024Overheating of Device; Overheating of Device

    IT WAS ORIGINALLY REPORTED THAT DURING A PROCEDURE, A PATIENT WAS CUT FROM A DEVICE THAT WAS LEFT ON THE DRAPE, AND INACTIVE. UPON FURTHER INVESTIGATION, IT WAS REPORTED THAT THE PATIENT WAS NOT CUT, BUT DID RECEIVE A SMALL BURN TO THEIR PALM, THIS WAS TREATED WITH AN OINTMENT. IT WAS ALSO REPORTED THAT THE THERE WAS AN UNKNOWN DELAY, AND THE PROCEDURE WAS COMPETED SUCCESSFULLY.

    Patient problems: Burn(s)

    MAUDE report 20560244 ↗
  • MalfunctionJul 24, 2024Overheating of Device; Overheating of Device

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE REPORTEDLY OVERHEATED. - 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19818525 ↗
  • MalfunctionApr 26, 2024Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT DURING A PROCEDURE THE PATIENTS SKIN GOT BURNED. IT WAS FURTHER REPORTED, THAT THERE WAS NO MEDICAL INTERVENTION AND NO DELAYS AS A RESULT OF THIS EVENT, AND THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: Burn(s)

    MAUDE report 19191497 ↗
  • InjurySep 28, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING A SURGERY, THE SURGEON MISTAKENLY CONTACTED THE SONOPET TIP ON THE VESSEL AND THE BLOOD VESSEL WAS DAMAGED. AN ANGIORRHAPHY WAS PERFORMED TO REPAIR THE VESSEL. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AFTER A 60 MINUTE DELAY. IT WAS FURTHER REPORTED THAT 1 WEEK LATER, A SUBARACHNOID HEMORRHAGE HAPPENED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 17834861 ↗
  • MalfunctionMar 28, 2023Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT DURING A JAWBONE OSTEOTOMY, A PATIENT HAD A BURN INJURY ON THE MOUTH CORNER. AN ANTIBIOTIC OINTMENT WAS PRESCRIBED TO TREAT THE BURN BY KEEPING THE AREA MOIST. THE BURN WILL NOT NEED ADDITIONAL TREATMENT AND NO INFECTION WAS OBSERVED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: Partial thickness (Second Degree) Burn; Superficial (First Degree) Burn

    MAUDE report 16625938 ↗
  • InjuryJan 10, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE PATIENT EXPERIENCED A DURAL INJURY. IT WAS ALSO REPORTED THERE WERE NO DELAYS AND NO MEDICAL INTERVENTION AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: Rupture

    MAUDE report 16130744 ↗
  • InjuryDec 29, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING A CRANIAL PROCEDURE, THE TIP OF THE DEVICE DAMAGED THE PATIENTS DURA. IT WAS ALSO REPORTED THAT THERE WAS AN UNINTENTIONAL DURAL OPENING AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THE PATIENTS DURA REQUIRED TO BE SUTURED. IT WAS ALSO REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: Rupture

    MAUDE report 16070980 ↗
  • InjuryNov 4, 2022Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT DURING A RHINOPLASTY, THE SURGEON REQUESTED THE SUCTION TURNED DOWN AND THE TECH ACCIDENTLY CLAMPED THE SUCTION IRRIGATION. THE PATIENT WAS BURNED. THE PATIENT WAS IMMEDIATELY TREATED WITH COLD SALINE FLUSHES AND SOAKED GAUZE, FOLLOWED BY ANTIBIOTIC COVERAGE WITH ADAPTIC DRESSING. CONTINUAL OINTMENT COVERAGE, ADAPTIC DRESSING AND BANDAID. IT WAS FURTHER REPORTED THAT THERE WAS A SLIGHT SURGICAL DELAY AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: Partial thickness (Second Degree) Burn

    MAUDE report 15736434 ↗
  • MalfunctionFeb 10, 2021Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT DURING A POSTERIOR CERVICAL DECOMPRESSION, PEDIATRIC SURGICAL PROCEDURE, THE HANDPIECE BURNED THE RIGHT HAND OF THE SURGEON AND ALSO THE PATIENTS NECK. IT WAS ALSO REPORTED THAT A NEW TIP WAS FITTED AND THE HANDPIECE CONTINUED TO OVERHEAT. THIS RECORD ADDRESSES ONE OF THE TIPS THAT CONTRIBUTED TO THE OVERHEATING EVENT.IT WAS ALSO REPORTED THERE WERE NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT A 10 MINUTE DELAY OCCURRED AS A RESULT OF THIS EVENT. IT WAS ALSO REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING A BACK-UP DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11309675 ↗
  • MalfunctionFeb 10, 2021Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT DURING A POSTERIOR CERVICAL DECOMPRESSION, PEDIATRIC SURGICAL PROCEDURE, THE HANDPIECE BURNED THE RIGHT HAND OF THE SURGEON AND ALSO THE PATIENTS NECK. IT WAS ALSO REPORTED THAT A NEW TIP WAS FITTED AND THE HANDPIECE CONTINUED TO OVERHEAT. THIS RECORD ADDRESSES ONE OF THE TIPS THAT CONTRIBUTED TO THE OVERHEATING EVENT. IT WAS ALSO REPORTED THERE WERE NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT A 10 MINUTE DELAY OCCURRED AS A RESULT OF THIS EVENT. IT WAS ALSO REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING A BACK-UP DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions