ThermaClear 502500001Recalls & reported problems
Stryker Corporation · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07613327626391), received between Aug 18, 2026 and Sep 24, 2026 — classified as 2 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Complete Loss of Power” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Complete Loss of Power3
- Decrease in Pressure3
- Inability to Irrigate3
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Unspecified Tissue Injury2
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 24, 2026Inability to Irrigate
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50892157 ↗MalfunctionSep 24, 2026Inability to Irrigate
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50889814 ↗InjurySep 9, 2026Inability to Irrigate; Decrease in Pressure
This event was initially reported as a serious injury. Upon additional evaluation, this event is no longer considered reportable. \ \Stryker performed a medical risk assessment on the reported event details. Based on the assessment, the event details confirm that a temporary injury/impairment took place that did not require professional medical intervention and does not meet the criteria for a serious injury. Additionally, the reportable event does not meet the requirements of a malfunction MDR. Therefore, the event has been determined to be non-reportable.\ \The review of historical complaint data, risk documentation, and the reported information regarding this event do not suggest a recurrence of this event would be likely to result in a serious injury.
Patient problems: Unspecified Tissue Injury
MAUDE report 50695570 ↗InjurySep 4, 2026Decrease in Pressure; Complete Loss of Power
It was reported that an email received from the doctor on (b)(6) 2026. The email that describes a patient injury that occurred on (b)(6) 2025 (exact date was not given) stated, "during the device trial conducted before our purchase decision: the ThermaClear lost power during morcellation and the patient sustained a bladder injury"*Update from the doctor:Device failure resulted in loss of fluid irrigation necessary to maintain bladder distention during morcellation. The bladder decompressed suddenly and there was a morcellator injury to the bladder. We assessed the injury and determined that it was safe to proceed with the rest of the surgery. The Thermaclear pump had to be replaced with gravity drainage fluid setup, resulting in unnecessarily prolonged surgery and additional steps to recover hemostasis after the bladder decompression event.Surgical delay is estimated at 15-30 minutes due to troubleshooting equipment failure and replacing with backup equipment. Surgical steps (scope change, return to hemostasis, another scope change to proceed with morcellation) had to be repeated due to the loss of irrigation fluid.
Patient problems: Unspecified Tissue Injury
MAUDE report 50620217 ↗MalfunctionSep 4, 2026Decrease in Pressure; Complete Loss of Power
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50618511 ↗MalfunctionAug 18, 2026Complete Loss of Power
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50309166 ↗
FDA clearance and classification
Cleared through 510(k) K210628 on Nov 4, 2021 (decision: Substantially Equivalent). Submitted by Thermedx, LLC as “X-FLO Fluid Management System”. The FDA review took 247 days. FDA record ↗
The FDA regulates product code HIG (insufflator, hysteroscopic) under 21 CFR 884.1700 as a Class II device. 21 CFR 884.1700 ↗
Common questions about the ThermaClear 502500001
Has the ThermaClear 502500001 been recalled?
No FDA recall record in our data names this exact device identifier (DI 07613327626391), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the ThermaClear 502500001?
The FDA's MAUDE database lists 6 adverse event reports for this device, received between Aug 18, 2026 and Sep 24, 2026. In those reports, the most common specific device problems are “Complete Loss of Power” (3), “Decrease in Pressure” (3) and “Inability to Irrigate” (3); the most common patient problem is “Unspecified Tissue Injury” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ThermaClear 502500001?
None of the reports are classified as deaths, 2 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the ThermaClear 502500001 FDA approved?
It was cleared through 510(k) K210628 on Nov 4, 2021. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the ThermaClear 502500001 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the ThermaClear 502500001, and is it still sold?
The FDA's device database lists Stryker Corporation as the company responsible for it (brand ThermaClear, model 502500001). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Insufflator, hysteroscopic devices
- Fluent Fluid Management System FLT-112SHologic, Inc. · INSUFFLATOR, HYSTEROSCOPIC140 reports
- Aquilex® Fluid Control System AQL-100CBSW.O.M. World of Medicine GmbH · Insufflator, Hysteroscopic1 recall18 reports
- Aquilex Fluid Control System AQL-100PBSW.O.M. World of Medicine GmbH · Insufflator, Hysteroscopic1 recall8 reports
- ENDOMAT UP210Karl Storz GmbH & Co. KG · Insufflator, hysteroscopic4 deaths17 reports
- Aquilex® Fluid Control System AQL-100CSW.O.M. World of Medicine GmbH · Insufflator, Hysteroscopic15 reports
- Fluent Console FLT-100Hologic, Inc. · INSUFFLATOR, HYSTEROSCOPIC1 death13 reports
- Hysteroscopic Fluid Management System Stand 7210165Smith & Nephew, Inc. · INSUFFLATOR, HYSTEROSCOPIC9 reports
- Fluent Pro Fluid Management System FLT-200Hologic, Inc. · Insufflator, hysteroscopic1 death6 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page