HDL-Cholesterol Gen.4 08057877190Recalls & reported problems

Roche Diagnostics GmbH · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07613336121306), received between Jan 12, 2024 and Jul 2, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings2
  • Low Test Results2
  • Non Reproducible Results2
  • Output Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 2, 2025Low Test Results; Low Test Results

    THERE WAS AN ALLEGATION OF QUESTIONABLE HDL-CHOLESTEROL GEN.4 RESULTS FOR 1 PATIENT ON A COBAS C 303 ANALYTICAL UNIT. THE INITIAL RESULT WAS 15.5 MG/DL. THE RESULT WAS CONSIDERED TO BE TOO LOW AND THE SAMPLE WAS REPEATED. ON (B)(6) 2025, THE REPEATED RESULT WAS 14.8 MG/DL. BOTH VALUES WERE QUESTIONED AS THE PATIENT'S EXPECTED RESULT WAS 30 MG/DL. THE SAMPLE WAS SENT TO ANOTHER LABORATORY (METHOD UNKNOWN) TO BE TESTED, AND THE RESULT WAS 25 MG/DL. THE PATIENT'S CHOLESTEROL RESULT WAS 156 MG/DL, THE LDL RESULT WAS 94.0 MG/DL, AND THE TRIGLYCERIDE RESULT WAS 225 MG/DL. THE PHYSICIAN DID NOT AGREE WITH THE HDL RESULTS BASED ON THE PATIENT'S CLINICAL HISTORY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22383432 ↗
  • MalfunctionJun 2, 2025Output Problem; Non Reproducible Results; Output Problem; Non Reproducible Results

    WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 1 PATIENT SAMPLE TESTED FOR HDL-CHOLESTEROL GEN.4 (HDLC4) ASSAY ON A COBAS C503 ANALYTICAL UNIT. INITIAL RESULT: 3.35 MG/DL. THE CUSTOMER QUESTIONED THE INITIAL RESULT AND REPEATED THE SAMPLE. REPEAT RESULT: 92.80 MGDL. THE REPEAT RESULT OF 92.8 MGDL WAS DEEMED TO BE CORRECT. NO QUESTIONABLE RESULT WAS REPORTED TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22124149 ↗
  • MalfunctionFeb 19, 2025Non Reproducible Results; Non Reproducible Results

    THE INITIAL REPORTER RECEIVED QUESTIONABLE HDL- CHOLESTEROL GEN.4 (HDLC4) ASSAY RESULTS FROM TWO PATIENT SAMPLES TESTED ON THE COBAS PURE C 303 ANALYTICAL UNIT. SAMPLE 1: THE INITIAL RESULT WAS 218 MG/DL. THE REPEAT RESULT WAS 64 MG/DL. SAMPLE 2: THE INITIAL RESULT WAS 104 MG/DL. THE REPEAT RESULT WAS 42 MG/DL. THE INITIAL RESULTS WERE NOT REPORTED OUTSIDE THE LABORATORY BECAUSE THEY WERE DEEMED HIGH AND INCONSISTENT WITH PATIENTS' HISTORIES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21409818 ↗
  • MalfunctionJan 6, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THERE WAS AN ALLEGATION OF QUESTIONABLE HDL-CHOLESTEROL GEN.4 RESULTS FOR 1 PATIENT SAMPLE ON A COBAS C 303 ANALYTICAL UNIT. THE INITIAL RESULT WAS 3.18 MMOL/L. THE CUSTOMER CONSIDERED THE RESULT TO BE UNACCEPTABLY HIGH AND REPEATED THE SAMPLE. ON (B)(6) 2024 THE REPEATED RESULT WAS 1.64 MMOL/L.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21082972 ↗
  • MalfunctionJan 12, 2024Incorrect, Inadequate or Imprecise Result or Readings; Low Test Results; Incorrect, Inadequate or Imprecise Result or Readings; Low Test Results

    THERE WAS AN ALLEGATION OF QUESTIONABLE RESULTS FROM THE COBAS PURE C 303 ANALYTICAL UNIT. THE INITIAL HDLC4 RESULT WAS 38.2 MG/DL AND THE REPEAT RESULT WAS 76.9 MG/DL. THE INITIAL CHOLESTEROL RESULT WAS 283 MG/DL AND THE REPEAT RESULT WAS 161 MG/DL. THE INITIAL TRIGLYCERIDE RESULT WAS 249 MG/DL AND THE REPEAT RESULT WAS 58.1 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18508642 ↗