Skip to content

AU OSR6296Recalls & reported problems

Beckman Coulter, Inc. · LDL-Cholesterol · Class 1

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Dec 8, 2025Open, Classified

    Beckman Coulter identified that LDL Cholesterol OSR6x96 is not meeting the performance claim listed in the Instructions for Use (IFU). LDL Cholesterol reagent showed clinically significant interference due to triglyceride concentration up to 1500 mg/dL may cause a maximum positive bias of up to 28.8% in low LDL Cholesterol samples (concentration 80 mg/dL) and up to 11.02% in high LDL Cholesterol samples (concentration 160mg/dL).

    Z-1135-2026 · Root cause: Device Design · FDA record ↗

Other Ldl & Vldl Precipitation, Cholesterol Via Esterase-Oxidase, Hdl devices

All Ldl & Vldl Precipitation, Cholesterol Via Esterase-Oxidase, Hdl devices →