AU OSR6196Recalls & reported problems
Beckman Coulter, Inc. · LDL-Cholesterol · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Dec 8, 2025Class IIOpen, Classified
Beckman Coulter identified that LDL Cholesterol OSR6x96 is not meeting the performance claim listed in the Instructions for Use (IFU). LDL Cholesterol reagent showed clinically significant interference due to triglyceride concentration up to 1500 mg/dL may cause a maximum positive bias of up to 28.8% in low LDL Cholesterol samples (concentration 80 mg/dL) and up to 11.02% in high LDL Cholesterol samples (concentration 160mg/dL).
Z-1135-2026 · Root cause: Device Design · Recall details · FDA record ↗
FDA clearance and classification
Cleared through Special 510(k) K040471 on Mar 10, 2004 (decision: Substantially Equivalent). Submitted by Olympus America, Inc. as “Olympus Ldl Cholesterol Reagent, Osr6196/Osr6296; Olympus Ldl Cholesterol Calibrator, Odc0024”. The FDA review took 15 days. FDA record ↗
The FDA regulates product code LBS (ldl & vldl precipitation, cholesterol via esterase-oxidase, hdl) under 21 CFR 862.1475 as a Class I device. 21 CFR 862.1475 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the AU OSR6196
Has the AU OSR6196 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1135-2026 (Class II), began on Dec 8, 2025: Beckman Coulter identified that LDL Cholesterol OSR6x96 is not meeting the performance claim listed in the Instructions for Use (IFU). LDL Cholesterol reagent showed clinically significant interference due to… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
Is the AU OSR6196 FDA approved?
It was cleared through Special 510(k) K040471 on Mar 10, 2004. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the AU OSR6196, and is it still sold?
The FDA's device database lists Beckman Coulter, Inc. as the company responsible for it (brand AU, model OSR6196). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Ldl & Vldl Precipitation, Cholesterol Via Esterase-Oxidase, Hdl devices
- Atellica CH HDLC 11537213Siemens Healthcare Diagnostics Inc. · LDL & VLDL PRECIPITATION, CHOLESTEROL VIA ESTERASE-OXIDASE, HDL1 recall0 reports
- AU OSR6296Beckman Coulter, Inc. · Ldl & Vldl Precipitation, Cholesterol Via Esterase-Oxidase, Hdl1 recall0 reports
- HDL-Cholesterol Gen.4 07528582190Roche Diagnostics GmbH · LDL & VLDL PRECIPITATION, CHOLESTEROL VIA ESTERASE-OXIDASE, HDL6 reports
- HDL-Cholesterol Gen.4 08057877190Roche Diagnostics GmbH · LDL & VLDL PRECIPITATION, CHOLESTEROL VIA ESTERASE-OXIDASE, HDL5 reports
- HDL-Cholesterol Gen.4 07528566190Roche Diagnostics GmbH · LDL & VLDL PRECIPITATION, CHOLESTEROL VIA ESTERASE-OXIDASE, HDL5 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.