cobas e 801 Immunoassay analyzer 08459606001Recalls & reported problems

Roche Diagnostics GmbH · Class 1

All models of cobas e 801 Immunoassay analyzer (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613336158562), received between Jul 16, 2021 and Jul 16, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Non Reproducible Results” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Non Reproducible Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 16, 2021Non Reproducible Results; Non Reproducible Results

    THE INITIAL REPORTER COMPLAINED OF A DISCREPANT NEGATIVE RESULT FOR 1 PATIENT SAMPLE TESTED FOR ELECSYS TOXO IGG (TOXO IGG) ON A COBAS E801 MODULE. THE INITIAL RESULT WAS 0.51 UI/ML (NEGATIVE). ON (B)(6) 2021 A NEW SAMPLE WAS OBTAINED AND THE RESULT WAS 9.26 UI/ML (POSITIVE). ON (B)(6) 2021 THE ORIGINAL SAMPLE WAS REPEATED WITH A RESULT OF 8.9 UI/ML (POSITIVE). THE NEGATIVE RESULT WAS NOT REPORTED OUTSIDE OF THE LABORATORY. THE TOXO IGG REAGENT LOT NUMBER WAS 517479. THE EXPIRATION DATE WAS NOT PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12183238 ↗

Other ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE devices

All ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE devices →