Skip to content

Bilirubin Total Gen.3 10421410190Recalls & reported problems

Roche Diagnostics GmbH · Class 2

All models of Bilirubin Total Gen.3 (4)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613336235508), received between Jul 15, 2025 and Jul 15, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “High Test Results” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K131544); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 15, 2025High Test Results; High Test Results

    THERE WAS AN ALLEGATION OF QUESTIONABLE BILT3 BILIRUBIN TOTAL GEN.3 RESULTS FOR FOUR PATIENT SAMPLES ON A COBAS 6000 C (501) MODULE. PATIENT 1: THE INITIAL RESULT WAS 24.4 MG/DL, AND THE REPEATED RESULT WAS 1.4 MG/DL. PATIENT 2: ON (B)(6) 2025, THE INITIAL RESULT WAS 0.9 MG/DL, AND THE REPEATED RESULT WAS 2.4 MG/DL. PATIENT 3: ON (B)(6) 2025, THE INITIAL RESULT WAS 14.20 MG/DL, AND THE REPEATED RESULT WAS 0.7 MG/DL. PATIENT 4: ON (B)(6) 2025, THE INITIAL RESULT WAS 6.20 MG/DL, AND THE REPEATED RESULT WAS 0.3 MG/DL. THE SAMPLES WERE REPEATED BECAUSE THE PHYSICIAN QUESTIONED THE INITIAL RESULTS DUE TO THE RESULTS NOT MATCHING THE PATIENT'S CLINICAL HISTORY. THE REPEATED RESULTS WERE BELIEVED TO BE CORRECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22511353 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K131544. One submission often covers many models, so check the product and lot details in each record.

  • Z-0153-2020 · Sep 11, 2019 · Terminated
    Cobas c Bilirubin Total Gen.3 Catalog Number: 05795419190 - Product Usage: In vitro test for the quantitative determination of the total bilirubin in serum and plasma of adults an
  • Z-1145-2019 · Dec 17, 2018 · Terminated
    Bilirubin Total Gen.3, Cat. Nos. 05795397190, 05795419190, 05795320160, 05795338160, 05795346160 ***added as of 5/10/19***05795397190, 05795648190 Product Usage: In vitro te