Bilirubin Total Gen.3 10421436190Recalls & reported problems
Roche Diagnostics GmbH · Class 2
All models of Bilirubin Total Gen.3 (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07613336235522), received between Feb 9, 2026 and Sep 29, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K131544); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings2
- Low Test Results2
- Unable to Obtain Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Incorrect, Inadequate or Imprecise Result or Readings
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50937364 ↗MalfunctionSep 11, 2026Unable to Obtain Readings
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50734674 ↗MalfunctionFeb 17, 2026Incorrect, Inadequate or Imprecise Result or Readings; Low Test Results
THE INITIAL REPORTER COMPLAINED OF DISCREPANT RESULTS FOR 1 PATIENT SAMPLE TESTED FOR BILT3 BILIRUBIN TOTAL GEN.3 ON A COBAS C 503 ANALYTICAL UNIT. THE INITIAL RESULT WAS 40.3 UMOL/L. THE INITIAL RESULT WAS HELD DUE TO AUTO-VALIDATION RULES. THE REPEAT RESULT WAS 4.36 UMOL/L AND WAS AUTO-VERIFIED. THE INITIAL RESULT WAS BELIEVED TO BE CORRECT. THE REPEAT RESULT IS IN QUESTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24367589 ↗MalfunctionFeb 9, 2026Low Test Results
THERE WAS AN ALLEGATION OF A QUESTIONABLE BILT3 BILIRUBIN TOTAL GEN.3 ASSAY RESULT FOR A PATIENT SAMPLE TESTED ON A COBAS C 503 ANALYTICAL UNIT. THE INITIAL RESULT WAS 5.0 UMOL/L. THE SAMPLE WAS REPEATED TWO TIMES, RESULTING IN VALUES OF 214 UMOL/L AND 213 UMOL/L. THE SAMPLE WAS ALSO REPEATED ON THE COBAS C 303 INSTRUMENT AND THE RESULT WAS 222 UMOL/L.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24290479 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K131544. One submission often covers many models, so check the product and lot details in each record.
- Z-0153-2020 · Sep 11, 2019 · TerminatedCobas c Bilirubin Total Gen.3 Catalog Number: 05795419190 - Product Usage: In vitro test for the quantitative determination of the total bilirubin in serum and plasma of adults an
- Z-1145-2019 · Dec 17, 2018 · TerminatedBilirubin Total Gen.3, Cat. Nos. 05795397190, 05795419190, 05795320160, 05795338160, 05795346160 ***added as of 5/10/19***05795397190, 05795648190 Product Usage: In vitro te
Other DIAZO COLORIMETRY, BILIRUBIN devices
- Dimension® Flex® reagent cartridge TBI 10444957Siemens Healthcare Diagnostics Inc. · DIAZO COLORIMETRY, BILIRUBIN2 recalls4 reports
- Dimension Vista® Flex® reagent cartridge TBIL 10445146Siemens Healthcare Diagnostics Inc. · DIAZO COLORIMETRY, BILIRUBIN1 recall7 reports
- VITROS 8383051Ortho-Clinical Diagnostics, Inc. · DIAZO COLORIMETRY, BILIRUBIN1 recall5 reports
- VITROS 1612365Ortho-Clinical Diagnostics, Inc. · DIAZO COLORIMETRY, BILIRUBIN1 recall1 report
- ARCHITECT 6L45-21Abbott Laboratories · DIAZO COLORIMETRY, BILIRUBIN46 reports
- ARCHITECT 6L45-22Abbott GmbH · DIAZO COLORIMETRY, BILIRUBIN30 reports
- ARCHITECT 6L45-41Abbott Laboratories · DIAZO COLORIMETRY, BILIRUBIN18 reports
- Bilirubin Direct Gen.2 08056951190Roche Diagnostics GmbH · DIAZO COLORIMETRY, BILIRUBIN17 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page