Skip to content

MectaLock PEEK Suture Anchor 05.11.002Recalls & reported problems

Medacta International SA · MectaLock PEEK Ø2.9 – Long driver · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07630040727815), received between Jul 20, 2023 and Jul 20, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Fracture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 20, 2023Fracture; Fracture

    DURING ANCHOR INSERTION, THE SURGEON FELT LIKE A LOT OF FORCE WAS REQUIRED TO MALLET INTO BONE. ONCE THE ANCHOR WAS FULLY IMPLANTED, THE INTRODUCER CAME AWAY FROM THE ANCHOR IMMEDIATELY (WITHOUT NEED FOR COUNTERCLOCKWISE TURNS). THE BROKEN TIP OF THE INSERTER (REF. 66.01.010; MAT 33775) WAS VISIBLE IN THE INTRAOPERATIVE XRAY. IT IS FEASIBLE THAT, DURING THE MALLETING PHASE, THE ANCHOR AND THE INSERTER WERE NOT FULLY AXIALLY ALIGNED WITH THE PILOT HOLE AND THAT THE MALLETING WAS NOT FULLY AXIALLY ALIGNED WITH THE HANDLE OF THE INSERTER. SLOTTED AIMER WAS NOT USED. THE TIP WAS LEFT INSIDE THE PATIENT. SUTURE FIXATION WAS ADEQUATE AND SURGEON USED 2 MORE JLOCK ANCHORS SUCCESSFULLY FOLLOWING AND 1 PRIOR TO THE BREAKAGE. NO REVISION SURGERY IS PLANNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17361571 ↗

Other Fastener, Fixation, Nondegradable, Soft Tissue devices

All Fastener, Fixation, Nondegradable, Soft Tissue devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page