The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07630040728119), received between Oct 14, 2021 and Mar 10, 2026 — classified as 4 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated3
  • Adverse Event Without Identified Device or Use Problem1
  • Migration1

Patient problems

  • Joint Dislocation2
  • Joint Laxity1
  • Pain1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 10, 2026Device Dislodged or Dislocated

    AT 1 YEAR AND 4 MONTHS FROM THE PATIENT CAME IN PRESENTING PAIN DUE TO A DISLOCATION OF THE GLENOSPHERE AND LINER. THERE WAS NO REPORTED TRAUMA AS A CAUSE. THE SURGEON REVISED THE METAPHYSIS, GLENOSPHERE AND LINER TO A CONSTRAINED LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Joint Dislocation

    MAUDE report 24558206 ↗
  • InjurySep 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AT 11 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN DUE TO INSTABILITY WITH THE CAUSE UNKNOWN. THE SURGEON REVISED THE GLENOSPHERE, METAPHYSIS, AND POLY. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Joint Laxity

    MAUDE report 22931756 ↗
  • InjuryJan 28, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ABOUT 6 WEEKS AFTER THE PRIMARY SURGERY, THE X-RAYS SHOWED THAT THE GLENOSPHERE WAS PARTIALLY DISSOCIATED FROM THE BASEPLATE DUE TO GLENOSPHERE SCREW UNSCREWING. THE PATIENT HAD NO SYMPTOMS. REVISION SURGERY HAS BEEN PERFORMED, GLENOSPHERE AND GLENOSPHERE SCREW REVISED. GLENOSPHERE GUIDE WAS NOT USED DURING THE PRIMARY SURGERY.

    Patient problems: Joint Dislocation

    MAUDE report 21251764 ↗
  • InjuryOct 14, 2021Device Dislodged or Dislocated; Migration; Device Dislodged or Dislocated; Migration

    AT 4 MONTHS AFTER THE PRIMARY, THE PATIENT CAME IN REPORTING PAIN AND A GRINDING FEELING IN THE SHOULDER. THE SURGEON TOOK X-RAYS AND IT WAS DETERMINED THAT GLENOSPHERE SCREW AND IMPLANT WERE LOOSE. THE SURGEON REVISED THE LATERALIZED GLENOSPHERE AND THE POLY. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Pain

    MAUDE report 12631996 ↗