Skip to content

Passeo-18 5/80/90Recalls & reported problems

Biotronik AG · Class 2

All models of Passeo-18 (54)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07640119558844), received between Oct 5, 2020 and Aug 7, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture3
  • Decrease in Pressure1
  • Material Puncture/Hole1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 7, 2025Material Puncture/Hole; Material Puncture/Hole

    PASSEO-18 BALLOON CATHETERS WERE SELECTED FOR TREATMENT OF A MODERATELY CALCIFIED LESION (70% STENOSIS DEGREE) IN A MODERATELY TORTUOUS PART OF THE MID SFA. TWO BALLOONS RUPTURED. THEY COULD NOT EVEN BE INFLATED UP TO MAXIMUM PRESSURE BEFORE THE BALLOONS RUPTURED AT 8 ATM. THIS REFERS TO 2ND BALLOON, THE OTHER BALLOON IS REPORTED SEPARATELY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22735467 ↗
  • MalfunctionJan 24, 2024Material Rupture; Material Rupture

    A PASSEO-18 PERIPHERAL BALLOON CATHETER WAS SELECTED FOR TREATMENT OF A HIGHLY RESISTANT STENOTIC LESION IN THE CEPHALIC VEIN (AV FISTULA). THE BALLOON INFLATED OUTSIDE OF THE SHEATH (CA. 10 CM FROM SHEATH). BALLOON BURST OCCURRED IMMEDIATELY WHEN BURST PRESSURE REACHED. THE BALLOON WAS REMOVED COMPLETELY WITHOUT ISSUES. ANOTHER PASSEO-18 WAS USED, AND THE TREATMENT COULD BE COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18574498 ↗
  • MalfunctionJul 30, 2021Decrease in Pressure; Decrease in Pressure

    A PASSEO-18 BALLOON CATHETER WAS CHOSEN FOR TREATMENT. IT WAS ATTEMPTED TO INFLATE THE BALLOON UP TO 5 BAR, BUT TO KEEP IT AT 5 BAR IT WAS NECESSARY TO GIVE MORE PRESSURE. EVENTUALLY THE PASSEO-18 HAD TO BE REPLACED BY ANOTHER BALLOON AS IT COULD NOT BE USED ANYMORE. LOT NUMBER AND EVENT DATE ARE UNAVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12252857 ↗
  • MalfunctionNov 16, 2020Material Rupture; Material Rupture

    A PASSEO-18 BALLOON CATHETER WAS CHOSEN FOR THE TREATMENT OF A SEVERELY CALCIFIED LESION IN THE SUPERFICIAL FERMORAL ARTERY. THE PASSEO-18 BALLOON RUPTURED AT 6 ATM WITHOUT CAUSING ANY HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10848377 ↗
  • MalfunctionOct 5, 2020Material Rupture; Material Rupture

    A PASSEO-18 BALLOON CATHETER WAS CHOSEN FOR THE TREATMENT OF A SEVERELY CALCIFIED LESION IN THE SUPERFICIAL FERMORAL ARTERY. DURING THE ATTEMPT OF POSTDILATATION OF A STENT, THE PASSEO-18 BALLOON RUPTURED AT 6 ATM WITHOUT CAUSING ANY HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10627964 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page