Skip to content

Passeo-18 2/120/150Recalls & reported problems

Biotronik AG · Class 2

All models of Passeo-18 (54)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07640130414686), received between Dec 13, 2019 and Oct 25, 2023 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak2
  • Difficult to Remove1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 25, 2023Fluid/Blood Leak; Fluid/Blood Leak

    A PASSEO-18 BALLOON CATHETER WAS SELECTED FOR TREATMENT OF AN ARTERIOVENOUS FISTULA. AFTER INSERTION, DURING INFLATION OF THE DEVICE, IT WAS NOTICED THAT THE BALLOON WAS LEAKING. ANOTHER BALLOON WAS USED TO FINISH THE PROCEDURE. THE PATIENT WAS STABLE AND DISCHARGED HOME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18004694 ↗
  • MalfunctionNov 30, 2022Difficult to Remove; Difficult to Remove

    A PASSEO-18 BALLOON CATHETER WAS SELECTED FOR TREATMENT OF A MODERATELY CALCIFIED LESION (70 PERCENT STENOSIS DEGREE) IN THE MILDLY TORTUOUS OSTIAL RENAL ARTERY. AFTER DILATATION OF THE PASSEO-18 BALLOON CATHETER, THE BALLOON WAS BLOCKED AND GOT STUCK TOGETHER WITH THE GUIDEWIRE. THUS, BOTH DEVICES HAD TO BE WITHDRAWN AT THE SAME TIME, BUT THE ACCESS GOT LOST AND THEREFORE, THE OPERATION TIME WAS EXTENDED FOR 60 MINUTES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15887651 ↗
  • MalfunctionJan 27, 2020Leak/Splash; Leak/Splash

    THE PASSEO-18 BALLOON CATHETER WAS SELECTED FOR USE. DURING THE PROCEDURE, THE BALLOON COULD NOT BE INFLATED MORE THAN 2 ATM. AFTER REMOVAL, A HOLE WAS DETECTED AT THE END OF THE BALLOON.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9632163 ↗
  • MalfunctionDec 13, 2019Fluid/Blood Leak; Fluid/Blood Leak

    A PASSEO-18 BALLOON CATHETER WAS CHOSEN TO DILATE A CALCIFIED LESION (90 PERCENT STENOSIS DEGREE) IN A MILDLY TORTUOUS PROXIMAL ATA. DURING INFLATION A LEAKAGE OF THE CONTRAST MEDIUM WAS NOTICED AND THE BALLOON WAS CONFIRMED TO BE BROKEN.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9469576 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page