Passeo-18 5/200/90Recalls & reported problems
Biotronik AG · Class 2
All models of Passeo-18 (54)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 07640130429888), received between Sep 28, 2021 and Jan 10, 2023 — classified as 1 injury, 6 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fluid/Blood Leak4
- Fracture2
- Material Rupture1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Insufficient Information1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryJan 10, 2023Fracture; Fracture
A PASSEO-18 BALLOON CATHETER WAS CHOSEN FOR AN ANTEGRADE TREATMENT OF THE AFS TOGETHER WITH A COMMAND .018 WIRE. DURING DILATATION OF THE BALLOON UP TO NP PRESSURE, THE BALLOON RUPTURED AND WAS WITHDRAWN BUT KEPT HOOKING AT THE END OF THE 4F SHEATH. WHEN THE BALLOON CATHETER WAS OUT OF THE PATIENT, IT WAS SEEN THAT THE BALLOON WAS NOT COMPLETE, SOME PARTS DID REMAIN IN THE AFS. REMOVAL OF THESE PARTS OUT OF THE VESSEL WAS NOT SUCCESSFUL, IT WAS DECIDED TO STOP THE PROCEDURE. A BYPASS SURGERY IS PLANNED SOON.
Patient problems: Insufficient Information
MAUDE report 16134396 ↗MalfunctionSep 23, 2022Fracture; Fracture
A PASSEO-18 BALLOON CATHETER WAS SELECTED FOR TREATMENT OF A MILDLY CALCIFIED LESION (80 PERCENT STENOSIS DEGREE) IN THE POPLITEAL ARTERY. WHILE INFLATING THE BALLOON INSIDE THE PATIENT, THE PATIENT MOVED HIS KNEE AND THE BALLOON RUPTURED. UPON REMOVING, THE BALLOON FRACTURED IN TWO PARTS. ALL PARTS WERE REMOVED FROM PATIENTS BODY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15469722 ↗MalfunctionDec 14, 2021Material Rupture; Material Rupture
A PASSEO-18 BALLOON CATHETER WAS CHOSEN FOR TREATMENT. THE BALLOON RUPTURED DURING ITS FIRST INFLATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13003204 ↗MalfunctionSep 28, 2021Fluid/Blood Leak; Fluid/Blood Leak
A PASSEO-18 BALLOON CATHETER WAS CHOSEN FOR TREATMENT. THE BALLOON WAS PLACED PROPERLY INSIDE THE LESION. AS THE BALLOON WAS INFLATED UP TO 6 BAR THE PRESSURE OF THE BALLOON WOULD SLOWLY TURN DOWN. TO KEEP THE BALLOON INFLATED AT 6 BAR IT WAS NECESSARY TO KEEP ON GIVING MORE PRESSURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12540817 ↗MalfunctionSep 28, 2021Fluid/Blood Leak; Fluid/Blood Leak
A PASSEO-18 BALLOON CATHETER WAS CHOSEN FOR TREATMENT. THE BALLOON WAS PLACED PROPERLY INSIDE THE LESION. AS THE BALLOON WAS INFLATED UP TO 6 BAR THE PRESSURE OF THE BALLOON WOULD SLOWLY TURN DOWN. TO KEEP THE BALLOON INFLATED AT 6 BAR IT WAS NECESSARY TO KEEP ON GIVING MORE PRESSURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12540808 ↗MalfunctionSep 28, 2021Fluid/Blood Leak; Fluid/Blood Leak
A PASSEO-18 BALLOON CATHETER WAS CHOSEN FOR TREATMENT. THE BALLOON WAS PLACED PROPERLY INSIDE THE LESION. AS THE BALLOON WAS INFLATED UP TO 6 BAR THE PRESSURE OF THE BALLOON WOULD SLOWLY TURN DOWN. TO KEEP THE BALLOON INFLATED AT 6 BAR IT WAS NECESSARY TO KEEP ON GIVING MORE PRESSURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12540775 ↗MalfunctionSep 28, 2021Fluid/Blood Leak; Fluid/Blood Leak
A PASSEO-18 BALLOON CATHETER WAS CHOSEN FOR TREATMENT. THE BALLOON WAS PLACED PROPERLY INSIDE THE LESION. AS THE BALLOON WAS INFLATED UP TO 6 BAR THE PRESSURE OF THE BALLOON WOULD SLOWLY TURN DOWN. TO KEEP THE BALLOON INFLATED AT 6 BAR IT WAS NECESSARY TO KEEP ON GIVING MORE PRESSURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12540773 ↗
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page