Skip to content

Passeo-35 4/150/130Recalls & reported problems

Biotronik AG · Class 2

All models of Passeo-35 (24)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07640130434639), received between Mar 9, 2021 and Dec 11, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak1
  • Inflation Problem1
  • Material Puncture/Hole1
  • Material Rupture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 11, 2024Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole

    THE PASSEO-35 BALLOON SYSTEM WAS SELECTED FOR PRE-DILATATION OF A MODERATELY CALCIFIED LESION IN A MODERATELY TORTUOUS PROXIMAL SFA, BUT THE BALLOON WAS LEAKING DURING INFLATION. ANOTHER BALLOON WAS USED TO FINISH THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20906050 ↗
  • MalfunctionMar 23, 2021Inflation Problem; Inflation Problem

    THE PASSEO-35 BALLOON CATHETER WAS SELECTED FOR THE TREATMENT OF A CALCIFIED LESION IN THE SFA. ALTHOUGH THE DOCTOR TRIED FOR SEVERAL TIMES, THE BALLOON COULD NOT BE INFLATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11554042 ↗
  • MalfunctionMar 9, 2021Material Rupture; Material Rupture

    THE PASSEO-35 BALLOON WAS SELECTED FOR TREATMENT OF A SEVERELY CALCIFIED LESION (80 PERCENT STENOSIS DEGREE, LENGTH 200MM) IN THE SFA. THE BALLOON WAS INFLATED BUT BURST AT 18 BAR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11448500 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page