Passeo-35 Xeo 8/40/90Recalls & reported problems

Biotronik AG · Class 2

All models of Passeo-35 Xeo (17)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07640130451629), received between Jun 20, 2023 and Oct 1, 2025 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Material Rupture” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture3
  • Entrapment of Device1
  • Fracture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Hemorrhage/Blood Loss/Bleeding1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 1, 2025Fracture; Material Rupture; Fracture; Material Rupture

    A PASSEO-35 XEO WAS SELECTED FOR TREATMENT OF THE LEFT COMMON ILIAC ARTERY. THE BALLOON RUPTURED DURING DEFLATION TO REMOVE THE BALLOON FROM GUIDEWIRE. THE FIRST PART OF THE BALLOON WAS EXTRACTED THROUGH THE GUIDEWIRE, BUT THE OTHER PART REMAINED IN THE PATIENT. THIS REQUIRED AN EXTENSIVE SURGICAL APPROACH TO EXTRACT THE MISSING PART AND THE USE OF A SURGICAL PATCH.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 23189248 ↗
  • MalfunctionFeb 7, 2024Entrapment of Device; Material Rupture; Entrapment of Device; Material Rupture

    A PASSEO-35 XEO BALLOON CATHETER WAS SELECTED FOR TREATMENT OF A MODERATELY CALCIFIED LESION (50 PERCENT STENOSIS DEGREE) IN THE MODERATELY TORTUOUS PROXIMAL SFA. THE PASSEO-35 XEO BURST AT A PRESSURE OF 9 ATM. SINCE IT WAS NOT POSSIBLE TO PULL THE BALLOON OUT THROUGH THE SHEATH, IT WAS REMOVED ALL TOGETHER. NOTHING REMAINED IN PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18658437 ↗
  • InjuryJun 20, 2023Material Rupture; Material Rupture

    DURING A PTA SHUNT PROCEDURE, THE PASSEO-35 XEO BALLOON WAS INFLATED UP TO 22 ATM AND RUPTURED CIRCULAR. A PART OF THE BALLOON REMAINED IN THE VESSEL WHEN IT WAS REMOVED. THE REMAINING PART OF THE BALLOON WAS REMOVED BY SURGERY.