Skip to content

Passeo-35 Xeo 9/40/90Recalls & reported problems

Biotronik AG · Class 2

All models of Passeo-35 Xeo (17)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07640130451636), received between Apr 16, 2026 and Apr 16, 2026 — classified as 1 death.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture1
  • Material Rupture1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • DeathApr 16, 2026Fracture; Material Rupture

    DURING A FISTULAGRAM, AN IN-STENT RESTENOSIS WAS FOUND. THE PASSEO-35 XEO PERIPHERAL DILATATION CATHETER WAS USED FOR THE TREATMENT BUT WHILE INFLATING, THE BALLOON RUPTURED. MULTIPLE ATTEMPTS WERE MADE TO REMOVE THE BALLOON, BUT THEY WERE UNSUCCESSFUL, AND THE BALLOON SEPARATED INTO 2 PIECES, WITH A PORTION CONTINUING TO MIGRATE FURTHER THE PATIENT. JUST ABOUT 1/4 OF THE BALLOON WAS ABLE TO BE REMOVED. THE PATIENT WAS TRANSPORTED FROM THE OUTPATIENT CENTER TO THE HOSPITAL FOR FURTHER EVALUATION AND INTERVENTION. THE NEXT DAY, WHEN THE HOSPITAL DECIDED TO PERFORM ANOTHER PROCEDURE TO REMOVE THE REMAINING PARTS OF THE DEVICE, THE PATIENT PASSED AWAY.

    Patient problems: Insufficient Information

    MAUDE report 24893426 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page