Oscar 2.0/20-180/127Recalls & reported problems

Biotronik AG · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07640130455597), received between Aug 10, 2023 and Mar 12, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Fracture” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture4
  • Failure to Advance1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 12, 2024Fracture; Fracture

    AN OSCAR PERIPHERAL MULTIFUNCTIONAL CATHETER SYSTEM WAS SELECTED FOR REVASCULARIZATION OF A STAGE 4 POSTERIOR TIBIAL ARTERY. THE OSCAR DILATOR WAS MOVED TO THE TIBIAL FIBULAR TRUNK BUT WERE UNABLE TO BE MOVED ON THE WIRE. WHEN THE OSCAR BALLOON WAS INTRODUCED IN THE SUPPORT CATHETER, THE WHITE/BLUE HANDLE (LOCK GRIP) TOTALLY DISCONNECTED FROM THE CATHETER THEY THEN HAD TO RETRIEVE THE WHOLE SYSTEM AND CONTINUED THE PROCEDURE IN THE TRADITIONAL WAY WITH NO HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18890270 ↗
  • MalfunctionFeb 22, 2024Fracture; Fracture

    AN OSCAR MULTIFUNCTIONAL CATHETER WAS SELECTED FOR TREATMENT OF A SEVERELY CALCIFIED LESION (99 PERCENT STENOSIS DEGREE ) IN THE MILDLY TORTUOUS POSTERIOR TIBIAL ARTERY. THE OSCAR BALLOON WAS PREPARED AS PER IFU AND INSERTED INTO THE OSCAR SUPPORT CATHETER. THE BALLOON WAS ADVANCED TO LESION AND INFLATED TO RBP TWO TIMES. IT WAS ATTEMPTED TO ADVANCE THE BALLOON FURTHER DISTALLY, BUT THE STENOSIS WAS VERY TIGHT. THE CONSTANT FORWARD PRESSURE ON THE BALLOON CAUSED THE HUB OF THE SUPPORT CATHETER TO FRACTURE FROM THE HUB OF THE OSCAR DEVICE. THE OSCAR WAS REMOVED AND SUBSEQUENTLY THE SUPPORT CATHETER AND BALLOON SEPARATELY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18759270 ↗
  • MalfunctionFeb 8, 2024Fracture; Failure to Advance; Fracture; Failure to Advance

    THE OSCAR BALLOON SHAFT BROKE RIGHT AFTER THE METALLIC HUPOTUBE WHILE TRYING TO CROSS A SEVERELY CALCIFIED LESION AT THE DISTAL PTA. EXTREME FORCE WAS USED IN THE ATTEMPT TO CROSS THE LESION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18669183 ↗
  • MalfunctionAug 10, 2023Fracture; Fracture

    AN OSCAR PERIPHERAL MULTIFUNCTIONAL CATHETER SYSTEM WAS SELECTED FOR TREATMENT. EVENT OCCURRED WHILE USING OSCAR ATTEMPTING TO CROSS A LESION (100 PERCENT STENOSIS DEGREE) IN THE POSTERIOR TIBIAL ARTERY. A CONTRALATERAL APPROACH WAS USED, AND WHILE ADVANCING THE OSCAR SYSTEM THE PROXIMAL SHAFT FRACTURED. SYSTEM WAS THEN REMOVED FROM PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17498944 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →