Biotronik AG · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07640130455757), received between Aug 8, 2023 and Feb 27, 2024 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Fracture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture2
- Difficult to Remove1
- Entrapment of Device1
- Material Rupture1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 27, 2024Entrapment of Device; Difficult to Remove; Material Rupture; Entrapment of Device; Difficult to Remove; Material Rupture
AN OSCAR PERIPHERAL MULTIFUNCTIONAL CATHETER SYSTEM WAS SELECTED FOR TREATMENT OF A MODERATELY CALCIFIED LESION (70 PERCENT STENOSIS DEGREE) IN THE MODERATELY TORTUOUS SFA. DURING A CROSSOVER PROCEDURE, THE OSCAR SUPPORT CATHETER WITH THE DILATATOR WAS USED BUT COULDNT CROSS THE LESION WITH THE DILATATOR. SO, IT WAS DECIDED TO CHANGE TO THE 6MM OSCAR PTA BALLOON. IT WAS MANAGED TO CROSS THE STENOSIS AND A DILATATION UP TO 6BAR WAS DONE. THE PTA BALLOON IMMEDIATELY RUPTURED. AFTERWARDS IT WAS NOT POSSIBLE TO PULL THE PTA BALLOON OUT OF THE SUPPORT CATHETER. IT GOT STUCK. SO, THE WHOLE OSCAR SYSTEM WAS REMOVED. NOTHING REMAINED IN PATIENT, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH OTHER PRODUCTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18789502 ↗MalfunctionAug 10, 2023Fracture; Fracture
AN OSCAR PERIPHERAL MULTIFUNCTIONAL CATHETER SYSTEM WAS SELECTED FOR TREATMENT. EVENT OCCURRED WHILE USING OSCAR ATTEMPTING TO CROSS A LESION (100 PERCENT STENOSIS DEGREE) IN THE SFA. A CONTRALATERAL APPROACH WAS USED, AND WHILE ADVANCING THE OSCAR SYSTEM THE PROXIMAL SHAFT OF DILATOR FRACTURED. SYSTEM WAS THEN REMOVED FROM PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17498939 ↗MalfunctionAug 8, 2023Fracture; Fracture
AN OSCAR PERIPHERAL MULTIFUNCTIONAL CATHETER SYSTEM WAS SELECTED FOR TREATMENT OF A SEVERELY CALCIFIED LESION (80 PERCENT STENOSIS DEGREE) IN THE MODERATELY TORTUOUS DISTAL RCA. THE OSCAR WAS USED CROSS OVER FROM THE LEFT TO THE RIGHT DISTAL SFA AND INSERTED WITH THE DILATATOR IN POSITION 0. ONLY THE DILATOR COULD BE PUSHED FROM THE MIDDLE SFA FURTHER DOWN BUT NOT THE SUPPORT CATHETER, SO IT WAS DECIDED TO EXCHANGE THE DILATATOR WITH THE 6MM PTA BALLOON. DURING PULLING THE DILATATOR OUT, MASSIVE FRICTION WAS NOTICED. EVERYTHING WAS STRAIGHT, NO KINK OF THE WIRE WAS NOTICED. DURING PULLING THE DILATATOR OUT, THE SHAFT BROKE WHERE THE METAL SHAFT ENDS AND OVERLAPS TO THE PLASTIC SHAFT. IT WAS POSSIBLE TO TAKE THE DILATATOR FULLY OUT. ANOTHER BALLOON WAS USED. THE INTERVENTION WAS SUCCESSFULLY COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17483370 ↗
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page