Paradym RF CRT-D 9750Recalls & reported problems
Microport Crm Srl · Class 3 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08031527012278), received between Jun 13, 2018 and Jul 13, 2021 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Device Operates Differently Than Expected” (1 report).
3 other recalls cover devices cleared under the same FDA submission (P060027); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Device Operates Differently Than Expected1
- Failure to Capture1
- Inappropriate or Unexpected Reset1
- Over-Sensing1
- Premature Discharge of Battery1
- Premature Elective Replacement Indicator1
- Reset Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Chest Pain1
- Dizziness1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryJul 13, 2021Failure to Capture; Over-Sensing; Failure to Capture; Over-Sensing
REPORTEDLY, THE PATIENT HAD DIZZINESS FOR A COUPLE OF WEEKS. UPON INTERROGATION OF THE CRT-D, NOISE OVERSENSING AND LOSS OF CAPTURE WERE OBSERVED ON THE RIGHT VENTRICULAR CHANNEL. PRELIMINARY ANALYSIS REVEALED THAT A VENTRICULAR LEAD ISSUE IS SUSPECTED.
Patient problems: Dizziness
MAUDE report 12163802 ↗MalfunctionJun 1, 2021Reset Problem; Reset Problem
REPORTEDLY, UPON DEVICE INTERROGATION ON (B)(6) 2021, A WARNING STATING THAT A RESET OCCURRED ON (B)(6) 2021 AND RE-INITIALIZATION WAS PERFORMED WAS DISPLAYED. THE LAST BATTERY REFORMING WAS PERFORMED ON (B)(6) 2021. ON (B)(6) 2021, THE BATTERY VOLTAGE WAS AT 2.88V. A CHARGE TIME OF 15S WAS MEASURED. THE DEVICE WAS SUCCESSFULLY REPROGRAMMED AND THE STATISTICS WERE RESET. PROPER DEVICE OPERATION WAS OBSERVED. THE PATIENT HAD UNDERGONE SEVERAL SURGERIES SINCE THE BEGINNING OF THE YEAR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11913198 ↗InjuryDec 31, 2020Premature Discharge of Battery; Premature Elective Replacement Indicator; Premature Discharge of Battery; Premature Elective Replacement Indicator
REPORTEDLY, THE SUBJECT DEVICE REACHED THE RECOMMENDED REPLACEMENT TIME (RRT) EARLY.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 11100333 ↗InjuryJun 13, 2018Device Operates Differently Than Expected; Inappropriate or Unexpected Reset; Device Operates Differently Than Expected; Inappropriate or Unexpected Reset
IT WAS REPORTED THAT THE SUBJECT DEVICE PRESENTED A RESET WARNING MESSAGE DATED (B)(6) 2018 AND PARAMETERS HAD BEEN REPROGRAMMED TO VVI MODE AT 60 BPM WITH A VF ZONE AT 190 BPM. REPORTEDLY, FOLLOW-UP TESTING WAS PERFORMED WITH NO ABNORMALITIES OBSERVED AND THE DEVICE WAS RE-PROGRAMMED SUCCESSFULLY TO THE PRE-RESET PARAMETERS. A CHARGE TIME TEST DID NOT REVEAL ANY IRREGULARITY. THE PATIENT HAS RECENTLY COMPLAINED OF CHEST PAIN AND HAS COME TO THE EMERGENCY ROOM SEVERAL TIMES COMPLAINING OF A TINGLING IN THEIR CHEST.
Patient problems: Chest Pain
MAUDE report 7595229 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P060027. One submission often covers many models, so check the product and lot details in each record.
- Z-3230-2017 · Jul 24, 2017 · TerminatedPlatinium CRT DF1 1711, Model Number TDF033U, UDI/GTIN 08031527015538, Implantable cardioverter defibrillator, Biventricular or ventricular antitachycardia pacing, Dual or single
- Z-3233-2017 · Jul 24, 2017 · TerminatedPlatinium CRT-D 1741, Model Number TDF037U, UDI/GTIN 08031527015484, Implantable cardioverter defibrillator, Biventricular or ventricular antitachycardia pacing, Dual or single c
- Z-2579-2010 · Jun 7, 2010 · TerminatedSorin Paradym CRT-D Model 8750 and Sorin Group Paradym sonR CRT-D Model 8770. Made in Italy by Sorin Biomedical CRM S.r.l. Via Crescentino s.n. , 13040 Saluggia (VC) Italy
Other Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) devices
- Gallant™ CDHFA500QSt. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)2 recalls58 deaths8,401 reports
- Quartet™ 1458Q-86St. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)207 deaths6,631 reports
- QUADRA ASSURA MP™ CD3369-40QSt. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)103 deaths6,084 reports
- VIGILANT™ X4 CRT-D G247Boston Scientific Corporation · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)17 deaths2,066 reports
- Quadra Assura™ CD3365-40QSt. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)53 deaths2,056 reports
- Cobalt™ XT HF Quad CRT-D MRI SureScan™ DTPA2QQMedtronic, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)4 recalls40 deaths1,639 reports
- Unify Assura™ CD3357-40QSt. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)54 deaths1,784 reports
- Unify Assura™ CD3357-40CSt. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)56 deaths1,588 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.