PLATINIUM CRT-D 1711Recalls & reported problems

Microport Crm Srl · Class 3 · Implant

All models of PLATINIUM (4)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
4
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08031527015538), received between Nov 14, 2017 and Jan 18, 2024 — classified as 4 injuries, 1 malfunction.

The most frequently reported specific device problem is “Inappropriate/Inadequate Shock/Stimulation” (2 reports).

1 FDA recall record name this device's identifier.

3 other recalls cover devices cleared under the same FDA submission (P060027); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 9, 2018Class IITerminated

    Hardware configuration may lead to overconsumption, followed by loss of pacing and sensing capabilities. The device then cannot identify an arrhythmia that would require defibrillation shock therapy.

    Z-2824-2018 · Root cause: Component design/selection · Recall details · FDA record ↗

Most reported problems

Device problems

  • Inappropriate/Inadequate Shock/Stimulation2
  • Over-Sensing2
  • Adverse Event Without Identified Device or Use Problem1
  • Defibrillation/Stimulation Problem1
  • High impedance1
  • Inappropriate or Unexpected Reset1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Information1
  • No Known Impact Or Consequence To Patient1
  • Shock from Patient Lead(s)1
  • Unspecified Infection1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 18, 2024Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset

    REPORTEDLY, DURING THE INTERROGATION OF THE SUBJECT DEVICE, THE FOLLOWING WARNING MESSAGES WERE DISPLAYED - [27], [A26] AND [A3] INDICATING THE DEVICE HAD BEEN RESET 1 TIME SINCE BEGINNING OF LIFE ON DEC 6, 2023 STATING THE DEFIBRILLATION SYSTEM POTENTIALLY INEFFECTIVE. THE PATIENT WAS SENT HOME. THE CHARGE TIME WAS 13 SECONDS TODAY WITH A BATTERY VOLTAGE OF 2.9V. NO OTHER INFORMATION WAS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18535530 ↗
  • InjuryMar 5, 2020Inappropriate/Inadequate Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation

    REPORTEDLY, ON (B)(6) 2020, THE PATIENT WENT TO THE HOSPITAL FOLLOWING SHOCKS. A MAGNET HAS BEEN APPLIED ON THE DEVICE FOR A WEEK; HOWEVER, WHEN REMOVING THE MAGNET, THE PATIENT RECEIVED SHOCKS AGAIN. A LEAD FRACTURE IS SUSPECTED. PRELIMINARY ANALYSIS CONFIRMED THE PRESENCE OF VENTRICULAR NOISE OVERSENSING EPISODES LEADING TO INAPPROPRIATE SHOCKS DELIVERIES. THIS NOISE OVERSENSING MOST PROBABLY RESULTED FROM A VENTRICULAR LEAD ISSUE AND/OR CONNECTION ISSUE; HOWEVER, NONE OF THEM COULD BE CONFIRMED WITH CERTAINTY.

    Patient problems: No Information

    MAUDE report 9793477 ↗
  • InjuryDec 27, 2019High impedance; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; High impedance; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation

    REPORTEDLY, SEVERAL WARNINGS WERE DISPLAYED ON THE OVERVIEW SCREEN DURING A FOLLOW-UP PERFORMED ON 03 DECEMBER 2019. MOREOVER, NOISE OVERSENSING WAS OBSERVED ON THE ATRIAL AND VENTRICULAR CHANNELS IN THE RECORDED EPISODES. THE ASSOCIATED RV LEAD WAS IMPLANTED FOLLOWING THE RE-INTERVENTION PERFORMED ON (B)(6) 2018 TO REPLACE THE PREVIOUSLY IMPLANTED RV LEAD. PRELIMINARY ANALYSIS RESULTS SHOWED NON-PHYSIOLOGICAL SIGNALS IN THE ATRIAL AND VENTRICULAR CHANNELS, LEADING TO THE REPORTED NOISE OVERSENSING IN BOTH CHANNELS AND THE DELIVERY OF TWO INAPPROPRIATE SHOCKS. THE OBSERVED NOISE MOST PROBABLY RESULTED FROM LEADS ISSUES AND/OR LEADS/CRT-D CONNECTION ISSUES.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 9523167 ↗
  • InjuryJul 13, 2018Over-Sensing; Defibrillation/Stimulation Problem; Over-Sensing; Defibrillation/Stimulation Problem

    REPORTEDLY, EPISODES SHOWING VENTRICULAR OVERSENSING WERE OBSERVED IN THE DEVICE MEMORIES DURING A ROUTINE FOLLOW-UP PERFORMED ON (B)(6) 2018. IN ONE EPISODE, A SHOCK WAS TRIGGERED BUT COULD NOT BE DELIVERED DUE TO OVERLOAD DETECTION. PRELIMINARY ANALYSIS CONFIRMED THE REPORTED ISSUE AND SHOWED THAT IT WAS MOST PROBABLY DUE TO A DEFIBRILLATION LEAD ISSUE. BASED ON AVAILABLE DATA, NO ISSUE IS SUSPECTED ON THE ICD. RECOMMENDATIONS TO REPLACE THE DEFIBRILLATION LEAD WERE SENT ON (B)(6) 2018.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7685877 ↗
  • InjuryNov 14, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE ENTIRE SYSTEM WAS EXPLANTED DUE TO INFECTION ON (B)(6) 2017. AN INVESTIGATION IS REQUIRED.

    Patient problems: Unspecified Infection

    MAUDE report 7027692 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P060027. One submission often covers many models, so check the product and lot details in each record.

  • Z-3233-2017 · Jul 24, 2017 · Terminated
    Platinium CRT-D 1741, Model Number TDF037U, UDI/GTIN 08031527015484, Implantable cardioverter defibrillator, Biventricular or ventricular antitachycardia pacing, Dual or single c
  • Z-3230-2017 · Jul 24, 2017 · Terminated
    Platinium CRT DF1 1711, Model Number TDF033U, UDI/GTIN 08031527015538, Implantable cardioverter defibrillator, Biventricular or ventricular antitachycardia pacing, Dual or single
  • Z-2579-2010 · Jun 7, 2010 · Terminated
    Sorin Paradym CRT-D Model 8750 and Sorin Group Paradym sonR CRT-D Model 8770. Made in Italy by Sorin Biomedical CRM S.r.l. Via Crescentino s.n. , 13040 Saluggia (VC) Italy

Other Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) devices

All Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) devices →