Skip to content

SMR Shoulder system 9013.02.301Recalls & reported problems

Limacorporate SPA · Stem Extractor · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08033390097987), received between Sep 19, 2025 and Sep 11, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 11, 2026Break

    During an SMR revision performed on (b)(6) 2026, an intraoperative issue occurred. While the humeral components were being removed using the stem extractor, the extractor was struck with a mallet to remove the stem. During this maneuver, the threaded portion of the extractor broke and remained lodged inside the stem. The event resulted in an approximately 20-minute delay to the surgical procedure.The surgeon followed the standard surgical workflow for shoulder revision. The involved instrument was:SMR - STEM EXTRACTOR (Product code 9013.02.301, lot 14AA150A).Event happened in Australia.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50729553 ↗
  • MalfunctionSep 19, 2025Break; Break

    DURING A REVISION SURGERY PERFORMED ON (B)(6) 2025, DUE TO LOW GRAD INFECTION, AN INTRAOPERATIVE ISSUE OCCURRED. WHEN EXPLANTING THE STEM, THE THREAD OF SMR - STEM EXTRACTOR (PRODUCT CODE: 9013.02.301 - LOT: 15AA269A) BROKE. THE STEM REMAINED IN THE HUMERAL BONE. THE OPERATOR HAD TO SAW THE BONE TO EXPLANT THE HUMERAL STEM (PRODUCT CODE: 1304.15.190 - LOT: 2400630). THEN THE BONE WAS FIXED WITH CERCLAGES. THE SURGERY WAS PROLONGED ABOUT 25 MINUTES. FOR THE REVISION SURGERY PROCEDURE COMPLAITN NUMBER (B)(4) (MDR REPORT NUMBER 3008021110-2025-00128) WAS CREATED AND WILL BE SUBMITTED SEPARATELY. EVENT OCCURRED IN GERMANY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23096027 ↗

Other Extractor devices

All Extractor devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page