Skip to content

SMR Shoulder Stem ImpactorRecalls & reported problems

Limacorporate SPA · Stem Impactor · Class 2

All models of SMR Shoulder (67)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08033390098342), received between Oct 15, 2024 and Oct 15, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K100858); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 15, 2024Break; Material Integrity Problem; Break; Material Integrity Problem

    THE COMPLAINT SOURCE REPORTED THAT DURING NORMAL INSERTION OF THE TRIAL STEMS IN SEQUENCE, WHEN TRAIL STEM 16 SHORT WAS INSERTED AND IMPACTED, THE BLACK TIP AT THE END OF THE SMR - STEM IMPACTOR (PRODUCT CODE: 9013.02.302, LOT NR. 15AA096B) BROKE AND WAS STUCK IN THE TRIAL STEM. THE SURGEON WAS ABLE TO REMOVE THE STUCK TRIAL STEM FROM HUMERAL CANAL BY HIMSELF. IT WAS NOT POSSIBLE TO USE ADAPTOR 9095.11.M51 BECAUSE OF THE STUCK BLACK TIP. THE BROKEN TIP WAS REMOVED WITH A BIT OF FORCE. THE SURGERY WAS PROLONGED FOR 4 MINUTES. THE COMPLAINT SOURCE TESTED THE TRIAL 16 STEM WITH ANOTHER IMPACTOR, AND IT FUNCTIONED FINE. THE EVENT OCCURRED IN GERMANY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20448951 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K100858. One submission often covers many models, so check the product and lot details in each record.

  • Z-2540-2015 · Aug 14, 2015 · Terminated
    SMR Shoulder/Spalla Trial Heads ECC .4 mm Adaptor, REF 9013.30.021; SMR Shoulder/Spalla Trial Heads ECC .8 mm Adaptor, REF 9013.30.031; for use in shoulder implant surgery.
  • Z-2539-2015 · Aug 14, 2015 · Terminated
    SMR Shoulder/Spalla Trial Humeral Head diameter .42 mm, REF 9013.22.425; SMR Shoulder/Spalla Trial Humeral Head diameter .44 mm, REF 9013.22.445; SMR Shoulder/Spalla Trial Humeral

Other Prosthesis, Shoulder, Hemi-, Humeral, Metallic Uncemented devices

All Prosthesis, Shoulder, Hemi-, Humeral, Metallic Uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page