SMR Shoulder Stem ImpactorRecalls & reported problems
Limacorporate SPA · Stem Impactor · Class 2
All models of SMR Shoulder (67)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08033390098342), received between Oct 15, 2024 and Oct 15, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K100858); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Material Integrity Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 15, 2024Break; Material Integrity Problem; Break; Material Integrity Problem
THE COMPLAINT SOURCE REPORTED THAT DURING NORMAL INSERTION OF THE TRIAL STEMS IN SEQUENCE, WHEN TRAIL STEM 16 SHORT WAS INSERTED AND IMPACTED, THE BLACK TIP AT THE END OF THE SMR - STEM IMPACTOR (PRODUCT CODE: 9013.02.302, LOT NR. 15AA096B) BROKE AND WAS STUCK IN THE TRIAL STEM. THE SURGEON WAS ABLE TO REMOVE THE STUCK TRIAL STEM FROM HUMERAL CANAL BY HIMSELF. IT WAS NOT POSSIBLE TO USE ADAPTOR 9095.11.M51 BECAUSE OF THE STUCK BLACK TIP. THE BROKEN TIP WAS REMOVED WITH A BIT OF FORCE. THE SURGERY WAS PROLONGED FOR 4 MINUTES. THE COMPLAINT SOURCE TESTED THE TRIAL 16 STEM WITH ANOTHER IMPACTOR, AND IT FUNCTIONED FINE. THE EVENT OCCURRED IN GERMANY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20448951 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K100858. One submission often covers many models, so check the product and lot details in each record.
- Z-2540-2015 · Aug 14, 2015 · TerminatedSMR Shoulder/Spalla Trial Heads ECC .4 mm Adaptor, REF 9013.30.021; SMR Shoulder/Spalla Trial Heads ECC .8 mm Adaptor, REF 9013.30.031; for use in shoulder implant surgery.
- Z-2539-2015 · Aug 14, 2015 · TerminatedSMR Shoulder/Spalla Trial Humeral Head diameter .42 mm, REF 9013.22.425; SMR Shoulder/Spalla Trial Humeral Head diameter .44 mm, REF 9013.22.445; SMR Shoulder/Spalla Trial Humeral
Other Prosthesis, Shoulder, Hemi-, Humeral, Metallic Uncemented devices
- DJO SURGICAL 509-02-032Encore Medical, L.P. · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED364 reports
- DELTA XTEND 130790036Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED171 reports
- DELTA XTEND 130770018Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED159 reports
- TRIGEN 71118200Smith & Nephew, Inc. · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED123 reports
- DJO SURGICAL 509-02-432Encore Medical, L.P. · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED119 reports
- DELTA XTEND 130738206Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED116 reports
- COMPREHENSIVE SHOULDER SYSTEM 113633Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED2 recalls4 reports
- DELTA XTEND 130742206Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED99 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.