Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08033390270694), received between May 29, 2025 and May 29, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Failure to Osseointegrate” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate1
  • Patient Device Interaction Problem1

Patient problems

  • Implant Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 29, 2025Failure to Osseointegrate; Patient Device Interaction Problem; Failure to Osseointegrate; Patient Device Interaction Problem

    SHOULDER REVISION SURGERY PERFORMED ON (B)(6) 2025, DUE TO LOOSENING OF THE SMR CEMENTLESS MINI STEM (PRODUCT CODE: 1304.15.110, LOT#: 2408106 - STER. 2400113). ACCORDING TO THE RECEIVED INFORMATION, THE STEM DID NOT HAVE GOOD FIXATION PROXIMALLY, MORE THAN LIKELY AS A RESULT OF THE PRIMARY IMPLANT THAT WAS DUE TO A FRACTURE. IT WAS ALSO REPORTED THAT THE PATIENT PRESENTED OF COMPLAINING OF PAIN, AND X-RAYS SHOWED WHAT APPEARED TO BE BONE RESORPTION AND IT WAS DETERMINED THAT THE STEM WAS LOOSE. THE STEM ALONG WITH ALL OTHER STEM-RELATED PARTS WERE EXTRACTED, IN DETAILS: SMR CEMENTLESS MINI STEM (PRODUCT CODE: 1304.15.110, LOT#: 2408106 - STER. 2400113). SMR HUMERAL EXTENSION + 9 MM (PRODUCT CODE: 1352.15.001, LOT#: 2416019 - STER. 2400149). SMR REVERSE FINNED HUMERAL BODY (PRODUCT CODE: 135215050, LOT#: 2009741 - STER. 2000208). SMR REVERSE LINER + 3 MM (PRODUCT CODE: 136050815, LOT#: 23AT248 - STER. 2300197). THE DOCTOR THEN IMPLANTED COMPETITOR'S HUMERAL DEVICES. PREVIOUS SURGERY ON (B)(6) 2024. PATIENT IS A FEMALE, 60 YEARS OLD. NO OTHER CLINICAL DATA ARE AVAILABLE. EVENT HAPPENED IN THE US.

    Patient problems: Implant Pain

    MAUDE report 22104980 ↗

Other Prosthesis, Shoulder, Hemi-, Humeral, Metallic Uncemented devices