Skip to content

SMR System 1304.15.190Recalls & reported problems

Limacorporate SPA · SMR System - Finned Stem Dia. 19 mm L. 80 mm · Class 2 · Implant

All models of SMR System (47)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08033390270823), received between Sep 24, 2025 and Apr 22, 2026 — classified as 2 injuries.

The most frequently reported specific device problem is “Patient Device Interaction Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Patient Device Interaction Problem1

Patient problems

  • Joint Laxity1
  • Medical device site infection1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 22, 2026Patient Device Interaction Problem

    REVISION SURGERY OCCURRED ON (B)(6) 2026, DUE TO GLENOID BONE LOSS AND STEM LOOSENING. PREVIOUS SURGERY WAS PERFORMED ON (B)(6) 2025. DURING ORIGINAL SURGERY FOLLOWING HUMERAL COMPONENTS WERE IMPLANTED: FINNED STEM DIA. 19 MM L. 80MM (PN: 1304.15.190, LOT: 2513645, STER. 2500096). 140° HUMERAL BODY REVERSE (PN: 1352.15.011, LOT: 2513035, STER. 2500105). EXTENSION FOR HUMERAL REVERSE BODY (PN: 1352.15.001, LOT: 2430343, STER. 2400276). SMR REVERSE LINER +3MM (PN: 1360.50.815, LOT: 21AT277, STER. 2200049). IN COMBINATION WITH GLENOID COMPONENTS FROM ANOTHER MANUFACTURER. DURING REVISION THE SURGEON EXPLANTED THE WHOLE ASSEMBLY, BOTH GLENOID AND HUMERAL COMPONENTS, THAT WERE REPLACED WITH GLENOID FROM OTHER MANUFACTURED AND SAME HUMERAL COMPONENTS WITH THE EXCEPTION OF A THICKER FINNED STEM AND A THICKER LINER, BELOW DETAILED: FINNED STEM DIA. 20 MM L. 80MM (PN: 1304.15.200). SMR REVERSE LINER +6MM (PN: 1360.50.820). SURGERY WAS COMPLETED AS INTENDED. PATIENT IS MALE, DATE OF BIRTH ON (B)(6) 1961. EVENT OCCURRED IN THE UNITED STATES.

    Patient problems: Joint Laxity

    MAUDE report 24950301 ↗
  • InjurySep 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE COMPLAINT SOURCE REPORTED A REVISION SURGERY PERFORMED ON (B)(6) 2025, DUE TO LOW GRAD INFECTION. THE HUMERAL STEM (PRODUCT CODE:1304.15.190 - LOT: 2400630 - STER. (B)(4) WAS EXPLANTED. THE PREVIOUS SURGERY WAS PERFORMED ON (B)(6) 2024. PATIENT IS MALE, 85 YEARS OLD. THIS EVENT OCCURRED IN GERMANY.

    Patient problems: Medical device site infection

    MAUDE report 23130495 ↗

Other Prosthesis, Shoulder, Hemi-, Humeral, Metallic Uncemented devices

All Prosthesis, Shoulder, Hemi-, Humeral, Metallic Uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page