Skip to content

Pacific™ Plus PCP020120180Recalls & reported problems

Medtronic, Inc. · CATH PCP020120180 PCP D2.00 L 120 UL1800 · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08033477055602), received between Dec 13, 2023 and Dec 13, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 13, 2023Burst Container or Vessel; Burst Container or Vessel

    PHYSICIAN ATTEMPTED TO USE A PACIFIC PLUS PTA BALLOON DURING TREATMENT OF A PLAQUE LESION IN THE PATIENTS DISTAL POSTERIOR TIBIAL ARTERY. MODERATE VESSEL TORTUOSITY AND SEVERE VESSEL CALCIFICATION WERE REPORTED. LESION EXHIBITED CTO (CHRONIC TOTAL OCCLUSION-100%) STENOSIS. THERE WERE NO ABNORMALITIES REPORTED IN RELATION TO ANATOMY. EMBOLIC PROTECTION WAS NOT USED. A NON-MEDTRONIC INFLATION DEVICE WAS USED FOR BALLOON INFLATION. THERE WAS NO DAMAGE NOTED TO THE PRODUCT PACKAGING PRIOR TO USE. NO ISSUES WERE NOTED DURING REMOVAL OF THE DEVICE FROM THE PACKAGING. IFU WAS FOLLOWED AND THE DEVICE WAS PREPPED WITHOUT ISSUE. THE DEVICE WAS NOT PASSED THROUGH A PREVIOUSLY DEPLOYED STENT. NO RESISTANCE WAS NOTED DURING ADVANCEMENT OF THE DEVICE. IT WAS REPORTED THAT A BALLOON BURST OCCURRED DURING INFLATION AT 5ATMS. ALL FRAGMENTS OF THE BALLOON WERE RETRIEVED. A REPLACEMENT PACIFIC PLUS PTA BALLOON WAS USED TO COMPLETE PROCEDURE. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18317100 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page