Skip to content

PACIFIC XTREME PCU070200130 (UDI 08033477058320)Recalls & reported problems

Medtronic, Inc. · CATH PCU070200130 PCU US 07 00L200UL1300 · Class 2

All models of PACIFIC XTREME (25)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08033477058320), received between Dec 18, 2017 and Dec 18, 2017 — classified as 1 injury.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

1 other recall cover devices cleared under the same FDA submission (K103464); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Burst Container or Vessel1

Patient problems

  • Patient Problem/Medical Problem1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 18, 2017Burst Container or Vessel; Burst Container or Vessel

    PHYSICIAN EMPLOYED THE USE OF A PACIFIC XTREME PTA BALLOON CATHETER WITH A NON-MDT INFLATION DEVICE FOR THE TREATMENT OF A 100% STENOSED FIBROUS/PLAQUE LESION IN THE MID-PROXIMAL SUPERFICIAL FEMORAL ARTERY (SFA). LESION WAS SLIGHTLY CALCIFIED. VESSEL SLIGHTLY TORTUOUS. IFU WAS FOLLOWED. DEVICE PREPPED WITHOUT ISSUE. IT WAS REPORTED THAT A CIRCUMFERENTIAL/RADIAL BALLOON BURST OCCURRED AT AN INFLATION PRESSURE OF 8-9ATMS. ONE INFLATION WAS APPLIED TO THE BALLOON PREVIOUS TO BURST. THE DEVICE DID PASS THROUGH A PREVIOUSLY DEPLOYED STENT, BUT NO RESISTANCE WAS ENCOUNTERED / NO EXCESSIVE FORCE REQUIRED. A PART OF THE BALLOON RUPTURED OFF THE CATHETER AND BECAME LODGED WITHIN THE MID-SFA. A 10MM SNARE DEVICE WAS USED TO RETRIEVE THE RUPTURED BALLOON (ALL FRAGMENTS RETRIEVED). NO PATIENT INJURY WAS REPORTED.

    Patient problems: Patient Problem/Medical Problem

    MAUDE report 7127035 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K103464. One submission often covers many models, so check the product and lot details in each record.

  • Z-3055-2020 · Aug 26, 2020 · Terminated
    Pacific Xtreme (PTCA Balloon Dilation Catheter)

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates