Quantum Perfusion Centrifugal Blood Pump CP37 CP37V-V0Recalls & reported problems

Spectrum Medical Srl · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08051160300617), received between Mar 26, 2025 and Aug 29, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information4
  • Break2

Patient problems

  • Air Embolism3
  • Reduced Blood Flow1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 29, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    GOT AN URGENT CALL WHILE ON BYPASS ASKING IF THE MAX RPM'S FOR THE CP37 WAS 4000. COULD NOT FLOW ABOVE 2.5 LITERS WITH A NORMAL LINE PRESSURE. WENT ON FINE WITH NORMAL FLOWS BUT WHEN LOWERED THE FLOW TO APPLY THE CROSS CLAMP AND CAME BACK UP COULD NOT COME TO FULL FLOW. PROCEEDED TO HAND CRANK AND ACHIEVED NORMAL FLOW. SWITCHED MOTORS WITH SAME LOW FLOW HIGH RPM RESULTS. AFTER SEVERAL MINUTES THEY WERE ABLE TO COME UP TO A NORMAL FLOW. ACT WAS GOOD AND CLAMP WAS ALSO RE-ADJUSTED BUT DIDN'T HELP. I DISCUSSED WITH THEM THE POSSIBILITIES OF: 1. SEATING OF DISPOSABLE. 2. CLOT IMPEADING FLOW INTO THE PUMP HEAD. 3. PROBLEM WITH THEIR LINE PRESSURE TRANDSDUCER. 4. POSSIBLE PROBLEM WITH CIRCUIT BEFORE WHERE THEY MEASURE LINE PRESSURE.

    Patient problems: Reduced Blood Flow

    MAUDE report 22921738 ↗
  • MalfunctionAug 5, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    CP37 PUMPHEAD CAVITATING SIGNIFICANT AIR DURING PRIMING. IT WAS DEAIRED MULTIPLE TIMES BUT IT CONTINUED TO ENTRAIN AIR. NO NOISE OR CRACKS/VISIBLE DAMAGE TO PUMPHEAD. PRESSURIZED CIRCUIT TO TEST FOR LEAK, BUT NO LEAK NOTED. SWAPPED OUT FOR NEW PUMPHEAD AND THE ISSUE DISAPPEARED.

    Patient problems: Air Embolism

    MAUDE report 22715133 ↗
  • MalfunctionApr 10, 2025Break; Insufficient Device Problem Information; Break; Insufficient Device Problem Information

    CP37 PUMPHEAD CAVITATING SIGNIFICANT AIR DURING DE-AIRING OF PRIMED HLM. ATTEMPTS TO DE-AIR RESULTED IN TINY CHAMPAGNE BUBBLE FORMATION ORIGINATING FROM THE CENTRIFUGAL PUMP. NO NOISE OR CRACKS/VISIBLE DAMAGE TO PUMPHEAD. PRESSURIZED CIRCUIT TO TEST FOR LEAK, BUT NO LEAK NOTED. SWAPPED OUT FOR NEW PUMPHEAD AND THE ISSUE DISAPPEARED.

    Patient problems: Air Embolism

    MAUDE report 21811739 ↗
  • MalfunctionMar 26, 2025Break; Insufficient Device Problem Information; Break; Insufficient Device Problem Information

    CP37 PUMPHEAD CAVITATING SIGNIFICANT AIR DURING PRIMING. IT WAS DEAIRED 3 TIMES BUT IT CONTINUED TO ENTRAIN AIR. NO NOISE OR CRACKS/VISIBLE DAMAGE TO PUMPHEAD. PRESSURIZED CIRCUIT TO TEST FOR LEAK, BUT NO LEAK NOTED. SWAPPED OUT FOR NEW PUMPHEAD AND THE ISSUE DISAPPEARED.

    Patient problems: Air Embolism

    MAUDE report 21695243 ↗

Other Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type devices

All Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type devices →