ACL TOP 500 CTS 00000280040Recalls & reported problems

Instrumentation Laboratory Company · ACL TOP 500 CTS · Class 2

All models of ACL TOP 500 CTS (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
8
Deaths reported
0
Injuries reported
8
Recalls naming it
1

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 08426950453499), received between Sep 11, 2017 and Oct 13, 2023 — classified as 8 injuries.

The most frequently reported specific device problem is “Use of Device Problem” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 22, 2022Class IIOpen, Classified

    Updated and revised user instructions issued 7/14/22. Potential carryover issue may cause elevated quality control and sample results for heparin, apixaban or rivaroxaban when HemosIL Liquid Anti-Xa reagent is used in the same run with both HemosIL Liquid Antithrombin and HemosIL Q.F.A. Thrombin (Bovine), specific to instrument models that utilize a single reagent probe

    Z-1014-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem6
  • Use of Device Problem2
  • Insufficient Device Problem Information1

Patient problems

  • Therapeutic Response, Increased6
  • Abrasion1
  • Needle Stick/Puncture1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 13, 2023Use of Device Problem; Use of Device Problem

    CUSTOMER INJURED FINGER WHILE REMOVING SAMPLE TUBE FROM CTS SAMPLE PROBE WHILE TROUBLESHOOTING. CUSTOMER WAS WEARING GLOVES AT THE TIME. CUSTOMER WENT TO OCCUPATIONAL HEALTH TO BE EVALUATED.

    Patient problems: Abrasion

    MAUDE report 17934001 ↗
  • InjuryMay 17, 2023Use of Device Problem; Use of Device Problem

    THE CUSTOMER WAS INJURED DURING MAINTENANCE PROCEDURE "CLEAN DEEP WASH AND CLEAN CUP AREA". THE CUSTOMER CLEANED THE SAMPLE CUP AREA AND AFTER THIS OPERATION TURNED UP THE FOOTER AND WAS INJURED WITH THE PIERCER. THE CUSTOMER WAS WEARING GLOVES. THERE WAS NO KNOWN ADVERSE EFFECT FROM THE PUNCTURE.

    Patient problems: Needle Stick/Puncture

    MAUDE report 16948092 ↗
  • InjurySep 3, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A CUSTOMER REPORTED THAT THEIR ACL TOP 500 CTS GENERATED ERRONEOUS APTT RESULTS USING HEMOSIL SYNTHASIL. FOUR PATIENTS HAD RESULTS THAT WERE CLINICALLY SIGNIFICANT AND NEEDED TO BE CORRECTED. ONE PATIENT WAS ADMINISTERED AN INCREASED DOSE OF HEPARIN BASED ON THE ERRONEOUS APTT RESULT. THE APTT RESULT WAS LATER CORRECTED AND THE HEPARIN DOSE WAS ADJUSTED AS APPROPRIATE. THERE WERE NO KNOWN COMPLICATIONS TO THE PATIENT FROM THE ADDITIONAL HEPARIN ADMINISTERED.

    Patient problems: Therapeutic Response, Increased

    MAUDE report 8956671 ↗
  • InjuryJan 8, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A CUSTOMER REPORTED THAT THEIR ACL TOP 500 CTS GENERATED ERRONEOUS APTT RESULTS USING HEMOSIL SYNTHASIL. FOUR PATIENTS WERE ADMINISTERED AN INCREASED DOSE OF HEPARIN BASED ON THE ERRONEOUS APTT RESULTS. AFTER THE ADDITIONAL DOSE OF HEPARIN, THE PATIENTS WERE MONITORED WITH NO FURTHER REPORTS OF CLINICAL IMPLICATIONS.

    Patient problems: Therapeutic Response, Increased

    MAUDE report 8225762 ↗
  • InjuryOct 31, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A CUSTOMER REPORTED THAT THEIR ACL TOP 500 CTS GENERATED AN ERRONEOUS APTT RESULT (36.9 SECONDS) USING HEMOSIL SYNTHASIL. A PATIENT WAS ADMINISTERED AN INCREASED DOSE OF HEPARIN BASED ON THIS ERRONEOUS APTT RESULT. SUBSEQUENTLY, ANOTHER SAMPLE WAS DRAWN FROM THE SAME PATIENT (APPROXIMATELY 3 HOURS LATER) AND IT REPORTED AN APTT RESULT OF > 200. THE HEPARIN ADMINISTRATION WAS STOPPED AND THE CUSTOMER INDICATED THAT THE PATIENT'S APTT RESULT WAS BROUGHT BACK INTO THE THERAPEUTIC RANGE. THERE WERE NO FURTHER REPORTS OF CLINICAL IMPLICATIONS FOR THE PATIENT.

    Patient problems: Therapeutic Response, Increased

    MAUDE report 8024101 ↗
  • InjuryNov 16, 2017Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem

    CUSTOMER REPORTED THAT THEIR ACL TOP 500 CTS REPORTED ERRONEOUS APTT RESULTS FOR 6 PATIENTS USING HEMOSIL SYNTHASIL. ALL 6 PATIENT SAMPLE RESULTS WERE RE-RUN AND CORRECTED. ONE PATIENT (CH) WAS ADMINISTERED A BOLUS OF HEPARIN. CH DEVELOPED A SUPERFICIAL RIGHT GROIN HEMATOMA. THE CUSTOMER WAS UNABLE TO DETERMINE IF THIS WAS A RESULT OF THE BOLUS OF HEPARIN.

    Patient problems: Therapeutic Response, Increased

    MAUDE report 7038808 ↗
  • InjurySep 22, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    CUSTOMER REPORTED THAT THEIR ACL TOP 500 CTS REPORTED ERRONEOUS APTT PATIENT RESULTS USING HEMOSIL SYNTHASIL. SIX PATIENT SAMPLES GENERATED ERRONEOUS RESULTS. ONE PATIENT WAS ADMINISTERED AN UNNECESSARY BOLUS OF HEPARIN DUE TO THE APTT RESULT BEING LOWER THAN EXPECTED. ALL SIX SAMPLES WERE SUBSEQUENTLY CORRECTED AND THERE WERE NO REPORTED SERIOUS ADVERSE EVENTS.

    Patient problems: Therapeutic Response, Increased

    MAUDE report 6885953 ↗
  • InjurySep 11, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    CUSTOMER REPORTED THAT THEIR ACL TOP 500 CTS REPORTED ERRONEOUS APTT PATIENT RESULTS USING HEMOSIL SYNTHASIL. OF THE 41 AFFECTED PATIENT SAMPLES, FIFTEEN TO TWENTY PATIENTS WERE ADMINISTERED UNNECESSARY BOLUSES OF HEPARIN DUE TO RESULTS BEING LOWER THAN EXPECTED. ALL FORTY-ONE SAMPLES WERE SUBSEQUENTLY CORRECTED AND THERE WERE NO REPORTED SERIOUS ADVERSE EVENTS.

    Patient problems: Therapeutic Response, Increased

    MAUDE report 6856196 ↗

Other INSTRUMENT, COAGULATION, AUTOMATED devices

All INSTRUMENT, COAGULATION, AUTOMATED devices →