ACL TOP 500 CTS 00000280040Recalls & reported problems
Instrumentation Laboratory Company · ACL TOP 500 CTS · Class 2
All models of ACL TOP 500 CTS (2)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 08426950453499), received between Sep 11, 2017 and Oct 13, 2023 — classified as 8 injuries.
The most frequently reported specific device problem is “Use of Device Problem” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Feb 22, 2022Class IIOpen, Classified
Updated and revised user instructions issued 7/14/22. Potential carryover issue may cause elevated quality control and sample results for heparin, apixaban or rivaroxaban when HemosIL Liquid Anti-Xa reagent is used in the same run with both HemosIL Liquid Antithrombin and HemosIL Q.F.A. Thrombin (Bovine), specific to instrument models that utilize a single reagent probe
Z-1014-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem6
- Use of Device Problem2
- Insufficient Device Problem Information1
Patient problems
- Therapeutic Response, Increased6
- Abrasion1
- Needle Stick/Puncture1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryOct 13, 2023Use of Device Problem; Use of Device Problem
CUSTOMER INJURED FINGER WHILE REMOVING SAMPLE TUBE FROM CTS SAMPLE PROBE WHILE TROUBLESHOOTING. CUSTOMER WAS WEARING GLOVES AT THE TIME. CUSTOMER WENT TO OCCUPATIONAL HEALTH TO BE EVALUATED.
Patient problems: Abrasion
MAUDE report 17934001 ↗InjuryMay 17, 2023Use of Device Problem; Use of Device Problem
THE CUSTOMER WAS INJURED DURING MAINTENANCE PROCEDURE "CLEAN DEEP WASH AND CLEAN CUP AREA". THE CUSTOMER CLEANED THE SAMPLE CUP AREA AND AFTER THIS OPERATION TURNED UP THE FOOTER AND WAS INJURED WITH THE PIERCER. THE CUSTOMER WAS WEARING GLOVES. THERE WAS NO KNOWN ADVERSE EFFECT FROM THE PUNCTURE.
Patient problems: Needle Stick/Puncture
MAUDE report 16948092 ↗InjurySep 3, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A CUSTOMER REPORTED THAT THEIR ACL TOP 500 CTS GENERATED ERRONEOUS APTT RESULTS USING HEMOSIL SYNTHASIL. FOUR PATIENTS HAD RESULTS THAT WERE CLINICALLY SIGNIFICANT AND NEEDED TO BE CORRECTED. ONE PATIENT WAS ADMINISTERED AN INCREASED DOSE OF HEPARIN BASED ON THE ERRONEOUS APTT RESULT. THE APTT RESULT WAS LATER CORRECTED AND THE HEPARIN DOSE WAS ADJUSTED AS APPROPRIATE. THERE WERE NO KNOWN COMPLICATIONS TO THE PATIENT FROM THE ADDITIONAL HEPARIN ADMINISTERED.
Patient problems: Therapeutic Response, Increased
MAUDE report 8956671 ↗InjuryJan 8, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A CUSTOMER REPORTED THAT THEIR ACL TOP 500 CTS GENERATED ERRONEOUS APTT RESULTS USING HEMOSIL SYNTHASIL. FOUR PATIENTS WERE ADMINISTERED AN INCREASED DOSE OF HEPARIN BASED ON THE ERRONEOUS APTT RESULTS. AFTER THE ADDITIONAL DOSE OF HEPARIN, THE PATIENTS WERE MONITORED WITH NO FURTHER REPORTS OF CLINICAL IMPLICATIONS.
Patient problems: Therapeutic Response, Increased
MAUDE report 8225762 ↗InjuryOct 31, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A CUSTOMER REPORTED THAT THEIR ACL TOP 500 CTS GENERATED AN ERRONEOUS APTT RESULT (36.9 SECONDS) USING HEMOSIL SYNTHASIL. A PATIENT WAS ADMINISTERED AN INCREASED DOSE OF HEPARIN BASED ON THIS ERRONEOUS APTT RESULT. SUBSEQUENTLY, ANOTHER SAMPLE WAS DRAWN FROM THE SAME PATIENT (APPROXIMATELY 3 HOURS LATER) AND IT REPORTED AN APTT RESULT OF > 200. THE HEPARIN ADMINISTRATION WAS STOPPED AND THE CUSTOMER INDICATED THAT THE PATIENT'S APTT RESULT WAS BROUGHT BACK INTO THE THERAPEUTIC RANGE. THERE WERE NO FURTHER REPORTS OF CLINICAL IMPLICATIONS FOR THE PATIENT.
Patient problems: Therapeutic Response, Increased
MAUDE report 8024101 ↗InjuryNov 16, 2017Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem
CUSTOMER REPORTED THAT THEIR ACL TOP 500 CTS REPORTED ERRONEOUS APTT RESULTS FOR 6 PATIENTS USING HEMOSIL SYNTHASIL. ALL 6 PATIENT SAMPLE RESULTS WERE RE-RUN AND CORRECTED. ONE PATIENT (CH) WAS ADMINISTERED A BOLUS OF HEPARIN. CH DEVELOPED A SUPERFICIAL RIGHT GROIN HEMATOMA. THE CUSTOMER WAS UNABLE TO DETERMINE IF THIS WAS A RESULT OF THE BOLUS OF HEPARIN.
Patient problems: Therapeutic Response, Increased
MAUDE report 7038808 ↗InjurySep 22, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
CUSTOMER REPORTED THAT THEIR ACL TOP 500 CTS REPORTED ERRONEOUS APTT PATIENT RESULTS USING HEMOSIL SYNTHASIL. SIX PATIENT SAMPLES GENERATED ERRONEOUS RESULTS. ONE PATIENT WAS ADMINISTERED AN UNNECESSARY BOLUS OF HEPARIN DUE TO THE APTT RESULT BEING LOWER THAN EXPECTED. ALL SIX SAMPLES WERE SUBSEQUENTLY CORRECTED AND THERE WERE NO REPORTED SERIOUS ADVERSE EVENTS.
Patient problems: Therapeutic Response, Increased
MAUDE report 6885953 ↗InjurySep 11, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
CUSTOMER REPORTED THAT THEIR ACL TOP 500 CTS REPORTED ERRONEOUS APTT PATIENT RESULTS USING HEMOSIL SYNTHASIL. OF THE 41 AFFECTED PATIENT SAMPLES, FIFTEEN TO TWENTY PATIENTS WERE ADMINISTERED UNNECESSARY BOLUSES OF HEPARIN DUE TO RESULTS BEING LOWER THAN EXPECTED. ALL FORTY-ONE SAMPLES WERE SUBSEQUENTLY CORRECTED AND THERE WERE NO REPORTED SERIOUS ADVERSE EVENTS.
Patient problems: Therapeutic Response, Increased
MAUDE report 6856196 ↗
Other INSTRUMENT, COAGULATION, AUTOMATED devices
- BCS® XP System 10459330Siemens Healthcare Diagnostics Products GmbH · INSTRUMENT, COAGULATION, AUTOMATED1 recall63 reports
- ACL TOP 550 CTS 00000280045Instrumentation Laboratory Company · INSTRUMENT, COAGULATION, AUTOMATED2 recalls3 reports
- ACL TOP 350 CTS 00000280065Instrumentation Laboratory Company · INSTRUMENT, COAGULATION, AUTOMATED2 recalls1 report
- ACL TOP 750 LAS 00000280035Instrumentation Laboratory Company · INSTRUMENT, COAGULATION, AUTOMATED1 recall4 reports
- ACL TOP 750 CTS 00000280055Instrumentation Laboratory Company · INSTRUMENT, COAGULATION, AUTOMATED1 recall3 reports
- ACL TOP 300 CTS 00000280060Instrumentation Laboratory Company · INSTRUMENT, COAGULATION, AUTOMATED1 recall1 report
- ACL TOP 500 CTS 00000280040RInstrumentation Laboratory Company · INSTRUMENT, COAGULATION, AUTOMATED1 recall0 reports
- ACL TOP 300 CTS 00000280060RInstrumentation Laboratory Company · INSTRUMENT, COAGULATION, AUTOMATED1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page